Parameters of Implant Stability Measurements Based on Resonance Frequency and Damping Capacity: A Comparative Finite Element Analysis

被引:0
|
作者
Winter, Werner [2 ]
Moehrle, Stefan [2 ]
Holst, Stefan [1 ]
Karl, Matthias [1 ]
机构
[1] Univ Erlangen Nurnberg, Dept Prosthodont, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Inst Appl Mech, D-91054 Erlangen, Germany
关键词
damping capacity; finite element analysis; implant stability; Osstell Mentor; Periotest; resonance frequency; DENTAL IMPLANT; IN-VITRO; NUMERICAL APPROACH; BONE LOSS; OSSEOINTEGRATION; PERIOTEST; BEHAVIOR;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Contradictory results have been reported on the comparability of implant stability measurements performed with the Periotest and the Osstell Mentor devices. The purpose of the present finite element analysis was to simulate the influence of the parameters implant length, bone quality (cortical thickness and damping factor), bone loss, and quality of transducer fixation on resonance frequency analysis (RFA) and damping capacity measurements. Materials and Methods: Three-dimensional finite element models of implants placed in human mandibular bone were designed for the simulation of Periotest (Periotest value; PTV) and RFA (implant stability quotient) measurements. Three values for each of the parameters implant length, damping capacity of cortical and trabecular bone, thickness of cortical bone, bone loss, and quality of transducer fixation were obtained. Measurements were simulated at four stages of osseointegration. Results: For all parameters, an increase in implant stability was found with increasing levels of osseointegration. Implant stability was positively correlated with implant length and thickness of cortical bone, with slightly converging values at increased levels of osseointegration. Varying the damping factor of bone had no significant effect. Implant stability was negatively correlated with bone loss, with slightly converging values at increased levels of osseointegration. Linear changes in implant length and bone loss caused nonlinear effects in implant stability values. Stiffness of transducer fixation had an impact on RFA measurements when values below 10 GPa were applied. Conclusion: Although both measuring devices reacted similarly when different parameters of implant stability were changed, good correlation between Periotest values and implant stability quotients was observed only when measurement values of implants without bone loss were considered. INT J ORAL MAXILLOFAC IMPLANTS 2010;25:532-539
引用
收藏
页码:532 / 539
页数:8
相关论文
共 50 条
  • [31] Resonance frequency measurements of implant stability in the dog mandible: experimental comparison with histomorphometric data
    Schliephake, H.
    Sewing, A.
    Aref, A.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 35 (10) : 941 - 946
  • [32] Influence of Simulated Bone Quality and Cortical Bone Thickness on Implant Stability Detection Using Resonance Frequency and Damping Factor Analysis
    Feng, Sheng-Wei
    Lin, Che-Tong
    Chang, Wei-Jen
    Lee, Sheng-Yang
    Cheng, Chiang-Hui
    Huang, Haw-Ming
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 (01) : 105 - 112
  • [33] Finite element analysis of vibration frequency and damping of cylindrical shell partially covered viscoelastic damping layer
    Chen Z.
    Sun W.
    Hangkong Dongli Xuebao/Journal of Aerospace Power, 2020, 35 (06): : 1176 - 1185
  • [34] Finite element analysis of mono- and bicortical mini-implant stability
    Holberg, Christof
    Winterhalder, Philipp
    Rudzki-Janson, Ingrid
    Wichelhaus, Andrea
    EUROPEAN JOURNAL OF ORTHODONTICS, 2014, 36 (05) : 550 - 556
  • [35] A nonlinear homogenized finite element analysis of the primary stability of the bone–implant interface
    Marzieh Ovesy
    Benjamin Voumard
    Philippe Zysset
    Biomechanics and Modeling in Mechanobiology, 2018, 17 : 1471 - 1480
  • [36] Subject specific finite element analysis of implant stability for a cementless femoral stem
    Pettersen, Sune H.
    Wik, Tina S.
    Skallerud, Bjorn
    CLINICAL BIOMECHANICS, 2009, 24 (06) : 480 - 487
  • [37] A simple mathematical model for the resonance frequency analysis of dental implant stability: Implant clamping quotient
    Argatov, Ivan
    Iantchenko, Alexei
    MECHANICS RESEARCH COMMUNICATIONS, 2019, 95 : 67 - 70
  • [38] Determine Corneal Biomechanical Parameters by Finite Element Simulation and Parametric Analysis Based on ORA Measurements
    Qin, Xiao
    Tian, Lei
    Zhang, Hui
    Zhang, Di
    Jie, Ying
    Zhang, Hai-Xia
    Li, Lin
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [39] Finite element based bone remodeling and resonance frequency analysis for osseointegration assessment of dental implants
    Li, Wei
    Lin, Daniel
    Rungsiyakull, Chaiy
    Zhou, Shiwei
    Swain, Michael
    Li, Qing
    FINITE ELEMENTS IN ANALYSIS AND DESIGN, 2011, 47 (08) : 898 - 905
  • [40] Resonance frequency measurements of implant stability in vivo - A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla
    Meredith, N
    Book, K
    Friberg, B
    Jemt, T
    Sennerby, L
    CLINICAL ORAL IMPLANTS RESEARCH, 1997, 8 (03) : 226 - 233