摘要
Purpose: Contradictory results have been reported on the comparability of implant stability measurements performed with the Periotest and the Osstell Mentor devices. The purpose of the present finite element analysis was to simulate the influence of the parameters implant length, bone quality (cortical thickness and damping factor), bone loss, and quality of transducer fixation on resonance frequency analysis (RFA) and damping capacity measurements. Materials and Methods: Three-dimensional finite element models of implants placed in human mandibular bone were designed for the simulation of Periotest (Periotest value; PTV) and RFA (implant stability quotient) measurements. Three values for each of the parameters implant length, damping capacity of cortical and trabecular bone, thickness of cortical bone, bone loss, and quality of transducer fixation were obtained. Measurements were simulated at four stages of osseointegration. Results: For all parameters, an increase in implant stability was found with increasing levels of osseointegration. Implant stability was positively correlated with implant length and thickness of cortical bone, with slightly converging values at increased levels of osseointegration. Varying the damping factor of bone had no significant effect. Implant stability was negatively correlated with bone loss, with slightly converging values at increased levels of osseointegration. Linear changes in implant length and bone loss caused nonlinear effects in implant stability values. Stiffness of transducer fixation had an impact on RFA measurements when values below 10 GPa were applied. Conclusion: Although both measuring devices reacted similarly when different parameters of implant stability were changed, good correlation between Periotest values and implant stability quotients was observed only when measurement values of implants without bone loss were considered. INT J ORAL MAXILLOFAC IMPLANTS 2010;25:532-539
引用
收藏
页码:532 / 539
页数:8
相关论文
共 50 条
  • [21] Effect of Model Parameters on Finite Element Analysis of Micromotions in Implant Dentistry
    Winter, Werner
    Klein, Daniel
    Karl, Matthias
    JOURNAL OF ORAL IMPLANTOLOGY, 2013, 39 (01) : 23 - 29
  • [22] A State Space Viscoelastic Shaft Finite Element for Stability and Response Analysis of Rotors with Structural and Frequency Dependent Damping
    Smitadhi Ganguly
    A. Nandi
    S. Neogy
    Journal of Vibration Engineering & Technologies, 2018, 6 : 1 - 18
  • [23] A State Space Viscoelastic Shaft Finite Element for Stability and Response Analysis of Rotors with Structural and Frequency Dependent Damping
    Ganguly, Smitadhi
    Nandi, A.
    Neogy, S.
    JOURNAL OF VIBRATION ENGINEERING & TECHNOLOGIES, 2018, 6 (01) : 1 - 18
  • [24] Two Implant/Abutment Joint Designs: A Comparative Finite Element Analysis
    Khraisat, Ameen
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (02) : E83 - E87
  • [25] Sacroiliac joint stability: Finite element analysis of implant number, orientation, and superior implant length
    Lindsey, Derek P.
    Kiapour, Ali
    Yerby, Scott A.
    Goel, Vijay K.
    WORLD JOURNAL OF ORTHOPEDICS, 2018, 9 (03): : 14 - 23
  • [26] Reliable method for the in vitro assessment of dental implant stability.: A comparison study of damping capacity measurement with the periotest instrument and resonance frequency analysis with the osstell instrument.
    Lachmann, S.
    Jaeger, B.
    Gomez-Roman, G.
    Axmann, D.
    Weber, H.
    JOURNAL OF DENTAL RESEARCH, 2003, 82 : B174 - B174
  • [27] Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects:: an in vitro study
    Tozum, T. F.
    Turkyilmaz, I.
    McGlumphy, E. A.
    JOURNAL OF ORAL REHABILITATION, 2008, 35 (10) : 739 - 744
  • [28] Measurement of dental implant stability by resonance frequency analysis: A review of the literature
    Pilar Quesada-Garcia, Ma
    Prados-Sanchez, Estrella
    Victoria Olmedo-Gaya, Ma
    Munoz-Soto, Esther
    Paloma Gonzalez-Rodriguez, Ma
    Vallecillo-Capilla, Manuel
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (10): : E538 - E546
  • [29] Resonance frequency analysis of dental implant stability during the healing period
    Boronat Lopez, Araceli
    Balaguer Martinez, Jose
    Lamas Pelayo, Joana
    Carrillo Garcia, Celia
    Penarrocha Diago, Miguel
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2008, 13 (04): : E244 - E247
  • [30] Measurement of mini-implant stability using resonance frequency analysis
    Nienkemper, Manuel
    Wilmes, Benedict
    Panayotidis, Agamemnon
    Pauls, Alexander
    Golubovic, Vladimir
    Schwarz, Frank
    Drescher, Dieter
    ANGLE ORTHODONTIST, 2013, 83 (02) : 230 - 238