Posterior avulsion fracture at adjacent vertebral body during cervical disc replacement with ProDisc-C - A case report

被引:18
|
作者
Shim, Chan Shik [1 ]
Shin, Ho-Dong [1 ]
Lee, Sang-Ho [1 ]
机构
[1] Wooridul Spine Hosp, Dept Neurosurg, Seoul 135100, South Korea
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2007年 / 20卷 / 06期
关键词
artificial disc replacement; spine arthroplasty; ProDisc-C; avulsion fracture;
D O I
10.1097/BSD.0b013e31803b95db
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: A case report. Objective: To report posterior avulsion fractures at the adjacent vertebral bodies that occurred during cervical arthroplasty with ProDisc-C, which was thought to be related to design of the prosthesis and lack of the prudence in surgical technique. Summary of Background Data: There is no previous report on the avulsion fracture of the posterior vertebral body during cervical arthroplasty. Methods: A 32-year-old-man presented with painful numbness and weakness of the right upper extremity for 2 months. On magnetic resonance imaging and computer tomography scans, a huge disc herniation at C6-C7 segment compressing the right C7 root and the spinal cord was identified. To prevent acceleration of degeneration of the C5-C6 segment, disc replacement of the C6-C7 with the ProDisc-C after anterior discectomy was planned. Results: Nearing the end of the second chiseling with a box cutting chisel, and following further mallet impaction to insert the chisel a bit more, copious bleeding occurred. Hurried hemostasis was accomplished. The posterior central vertebral bodies of C6 and C7 were found to be fractured and avulsed bony fragments were displaced posteriorly compressing the spinal cord. The bony fragments were carefully removed and the ProDisc-C was inserted as planned. The patient recovered without neurologic deficit. Conclusions: Owing to keel design of the prosthesis, the possibility of posterior avulsion fracture of the vertebral body should always be appreciated in cervical disc arthroplasty with ProDisc-C, especially during chiseling with the box cutting chisel and inserting the prosthesis. Further, insertion of either of them beyond the posterior ends of the keel cuts made by a keel cutting chisel should not be tried.
引用
收藏
页码:468 / 472
页数:5
相关论文
共 50 条
  • [42] Posterior Midline Interspinal Fixation (PMIF) to Treat Persistent Severe Pain Due to Severe Compression Fracture of Thoracic Vertebral Body: A Case Report
    V. Raikar, Soubrata
    Patil, Arun A.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (03)
  • [43] Huge Cervical Extradural Schwannoma Presenting as C2-C5 Vertebral Body Lesion A Rare Case Report and Literature Review
    Lotfinia, Iraj
    Sayyahmelli, Sima
    Vahedi, Amir
    NEUROSURGERY QUARTERLY, 2010, 20 (01) : 39 - 41
  • [44] Changes in the centre of rotation and the anterior bone loss of the vertebral body in Mobi-C artificial disc replacement segments after cervical hybrid surgery: a retrospective study
    Ma, Yukun
    Yu, Xing
    Li, Chuanhong
    Xiong, Yang
    Zhao, He
    Yang, Yongdong
    Zhao, Dingyan
    Wang, Fengxian
    Qu, Yi
    Yang, Jizhou
    Bi, Lianyong
    Yue, Xinliang
    EUROPEAN SPINE JOURNAL, 2024, 33 (03) : 1265 - 1274
  • [45] Changes in the centre of rotation and the anterior bone loss of the vertebral body in Mobi-C artificial disc replacement segments after cervical hybrid surgery: a retrospective study
    Yukun Ma
    Xing Yu
    Chuanhong Li
    Yang Xiong
    He Zhao
    Yongdong Yang
    Dingyan Zhao
    Fengxian Wang
    Yi Qu
    Jizhou Yang
    Lianyong Bi
    Xinliang Yue
    European Spine Journal, 2024, 33 : 1265 - 1274
  • [46] Contralateral C7 nerve transfer through posterior vertebral approach combined with selective posterior rhizotomy of the affected cervical nerve in the treatment of central upper limb spastic paralysis A case report
    Guan, Jingyu
    Lin, Jun
    Guan, Xueqing
    Jin, Qiang
    Zhang, Wenchuan
    MEDICINE, 2021, 100 (12) : E25061
  • [47] Intraoperative Anterior Migration of the Prestige-LP Cervical Disc Owing to an Inappropriate Implantation Sequence During Continuous 2-Level Artificial Cervical Disc Replacement: A Case Report with 8-Year Follow-Up
    Meng, Yang
    Wang, Xiaofei
    Zhao, Zhengyang
    Wang, Beiyu
    Wu, Tingkui
    Liu, Hao
    WORLD NEUROSURGERY, 2018, 116 : 194 - 200
  • [48] Persistent First Intersegmental Vertebral Artery in Association With Type II Odontoid Fracture: Surgical Treatment Utilizing a Novel C1 Posterior Arch Screw: Case Report
    Carmody, Margaret A.
    Martin, Michael D.
    Wolfla, Christopher E.
    NEUROSURGERY, 2010, 67 (01) : E210 - U231
  • [50] Posterior Cervical Fusion of Occiput-T3 for Unstable Complex Odontoid Fracture in an 80-Year-Old Male With C2-Sacrum Synostosis From Ankylosing Spondylitis: A Case Report
    Frolov, David
    Porter, Matt
    Schmitz, Miguel
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (07)