Posterior avulsion fracture at adjacent vertebral body during cervical disc replacement with ProDisc-C - A case report

被引:18
|
作者
Shim, Chan Shik [1 ]
Shin, Ho-Dong [1 ]
Lee, Sang-Ho [1 ]
机构
[1] Wooridul Spine Hosp, Dept Neurosurg, Seoul 135100, South Korea
来源
JOURNAL OF SPINAL DISORDERS & TECHNIQUES | 2007年 / 20卷 / 06期
关键词
artificial disc replacement; spine arthroplasty; ProDisc-C; avulsion fracture;
D O I
10.1097/BSD.0b013e31803b95db
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: A case report. Objective: To report posterior avulsion fractures at the adjacent vertebral bodies that occurred during cervical arthroplasty with ProDisc-C, which was thought to be related to design of the prosthesis and lack of the prudence in surgical technique. Summary of Background Data: There is no previous report on the avulsion fracture of the posterior vertebral body during cervical arthroplasty. Methods: A 32-year-old-man presented with painful numbness and weakness of the right upper extremity for 2 months. On magnetic resonance imaging and computer tomography scans, a huge disc herniation at C6-C7 segment compressing the right C7 root and the spinal cord was identified. To prevent acceleration of degeneration of the C5-C6 segment, disc replacement of the C6-C7 with the ProDisc-C after anterior discectomy was planned. Results: Nearing the end of the second chiseling with a box cutting chisel, and following further mallet impaction to insert the chisel a bit more, copious bleeding occurred. Hurried hemostasis was accomplished. The posterior central vertebral bodies of C6 and C7 were found to be fractured and avulsed bony fragments were displaced posteriorly compressing the spinal cord. The bony fragments were carefully removed and the ProDisc-C was inserted as planned. The patient recovered without neurologic deficit. Conclusions: Owing to keel design of the prosthesis, the possibility of posterior avulsion fracture of the vertebral body should always be appreciated in cervical disc arthroplasty with ProDisc-C, especially during chiseling with the box cutting chisel and inserting the prosthesis. Further, insertion of either of them beyond the posterior ends of the keel cuts made by a keel cutting chisel should not be tried.
引用
收藏
页码:468 / 472
页数:5
相关论文
共 50 条
  • [31] Results of the prospective, randomized, controlled multicenter Food and Drug Administration investigational, device exemption study of the ProDisc-C total disc replacement versus anterior discectomy and fusion for the treatment of 1-level symptomatic cervical disc disease
    Murrey, Daniel
    Janssen, Michael
    Delamarter, Rick
    Goldstein, Jeffrey
    Zigler, Jack
    Tay, Bobby
    Darden, Bruce
    SPINE JOURNAL, 2009, 9 (04): : 275 - 286
  • [32] CORRECTION OF SPINAL INSTABILITY AND RECOVERY OF NEUROLOGIC LOSS FOLLOWING CERVICAL VERTEBRAL BODY REPLACEMENT - A CASE-REPORT
    LIGHT, TR
    WAGNER, FC
    JOHNSON, RM
    SOUTHWICK, WO
    SPINE, 1980, 5 (04) : 392 - 394
  • [33] ProDisc-C Total Disc Replacement Versus Anterior Cervical Discectomy and Fusion for Single-Level Symptomatic Cervical Disc Disease Seven-Year Follow-up of the Prospective Randomized US Food and Drug Administration Investigational Device Exemption Study
    Janssen, Michael E.
    Zigler, Jack E.
    Spivak, Jeffrey M.
    Delamarter, Rick B.
    Darden, Bruce V., II
    Kopjar, Branko
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (21): : 1738 - 1747
  • [34] Failure of an expandable cage-plate cervical vertebral body replacement: case report of a device related complication
    Trieb, Klemens
    Koch, Stephan
    EKLEM HASTALIKLARI VE CERRAHISI-JOINT DISEASES AND RELATED SURGERY, 2017, 28 (03): : 214 - 218
  • [35] Endovascular Treatment of Vertebral Artery Injury during Cervical Posterior Fusion (C1 Lateral Mass Screw) A Case Report
    Yoon, Kyeong-Wook
    Ko, Jung-Ho
    Cho, Chun-Sung
    Lee, Sang-Koo
    Kim, Young-Joon
    Kim, Young Jin
    INTERVENTIONAL NEURORADIOLOGY, 2013, 19 (03) : 370 - 376
  • [36] Vertebral Artery Injury During Cervical Discectomy and Fusion in a Patient With Bilateral Anomalous Arteries in the Disc Space: Case Report
    Gantwerker, Brian R.
    Baaj, Ali A.
    Maughan, Peter H.
    McDougall, Cameron G.
    White, William L.
    NEUROSURGERY, 2010, 67 (03) : E874 - U355
  • [37] A novel approach to fix an avulsion fracture of posterior cruciate ligament from the tibia when associated with a type C distal femoral fracture and a Hoffa fracture: a case report
    Mounasamy, Varatharaj
    Zuelzer, Wilhelm A.
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2012, 22 (03): : 251 - 254
  • [38] A novel approach to fix an avulsion fracture of posterior cruciate ligament from the tibia when associated with a type C distal femoral fracture and a Hoffa fracture: a case report
    Varatharaj Mounasamy
    Wilhelm A. Zuelzer
    European Journal of Orthopaedic Surgery & Traumatology, 2012, 22 : 251 - 254
  • [39] Vertebroplasty of the C2 Vertebral Body and Dens Using an Anterior Cervical Approach: Technical Case Report
    Dorman, John K.
    NEUROSURGERY, 2010, 67 (04) : 1143 - 1144
  • [40] Unicameral bone cyst of a cervical vertebral body and lateral mass with associated pathological fracture in a child - Case report and review of the literature
    Snell, BE
    Adesina, A
    Wolfla, CE
    JOURNAL OF NEUROSURGERY, 2001, 95 (02) : 243 - 245