Comparison of compressive strength between three different plates for mandibular angle fractures fixation

被引:4
|
作者
Lyrio, Mariana C. Negreiros [1 ]
Monnazzi, Marcelo Silva [1 ]
De Moraes, Marcio [1 ]
Hochuli-Vieira, Eduardo [1 ]
Nunes Reis, Jose Mauricio [1 ]
Pereira-Filho, Valfrido Antonio [1 ]
机构
[1] UNESP, Univ Estadual Paulista, Dept Diag & Oral & Maxillofacial Surg, Dent Sch Araraquara, BR-14801903 Araraquara, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
Bone plate; Bite force; Compressive strength; Mandibular fracture; BIOMECHANICAL EVALUATION; MINIPLATE; COMPLICATIONS;
D O I
10.1016/j.jcms.2013.10.005
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The present study aims to compare three types of internal fixation for fractures of the mandibular angle. Mechanical testing was performed on replicas of polyurethane hemimandibles sectioned at the angle region to simulate a fracture and fixed with three different hardwares. Fixation devices enrolled on this survey included the grid plates with and without an intermediate bar and the method described by Champy and colleagues in 1978 and the sample consisted of 10 hemimandibles for each group. Vertical loadings were applied on each hemimandible and recorded after a vertical displacement of 3 and 5 mm. Statistical analysis was made by means of the variance analysis (ANOVA) and the Duncan test with a significance level of 5%. The Champy technique showed a statistically significant increased resistance when compared to the grid plates after vertical displacements of 3 and 5 mm. The results of this survey suggest that the Champy technique, when compared to the grid plate positioned at the middle of the mandibular bone (placement site selected for this study), is more resistant than the grid plate and that the inclusion or not of an intermediate bar to the grid plates does not improve its resistance after linear vertical loadings. (C) 2013 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E277 / E280
页数:4
相关论文
共 50 条
  • [1] Comparison of resorbable plates and titanium plates for fixation stability of combined mandibular symphysis and angle fractures
    Lim, Ho-Yong
    Jung, Chang-Hwa
    Kim, Seong-Yong
    Cho, Jin-Yong
    Ryu, Jae-Young
    Kim, Hyeon-Min
    [J]. JOURNAL OF THE KOREAN ASSOCIATION OF ORAL AND MAXILLOFACIAL SURGEONS, 2014, 40 (06) : 285 - 290
  • [2] Biomechanical analysis of titanium fixation plates and screws in mandibular angle fractures
    Atik, F.
    Atac, M. S.
    Ozkan, A.
    Kilinc, Y.
    Arslan, M.
    [J]. NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2016, 19 (03) : 386 - 390
  • [3] A computer study of biodegradable plates for internal fixation of mandibular angle fractures
    Tams, J
    Van Loon, JP
    Otten, B
    Bos, RRM
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2001, 59 (04) : 404 - 407
  • [4] Internal Fixation of Mandibular Angle Fractures: A Comparison of 2 Techniques
    Mehra, Pushkar
    Murad, Haitham
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2008, 66 (11) : 2254 - 2260
  • [5] Salvation for Hardware Failure of Resorbable Fixation Plates in Bilateral Mandibular Angle Fractures
    Kwon, Yong-Dae
    Lee, Deok-Won
    Choi, Byung-Joon
    Suh, Joon-Ho
    Ko, Sujin
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2012, 23 (04) : 1192 - 1194
  • [6] Comparison of Management of Mandibular Angle Fractures by Three Approaches
    Sudhakar G.V.S.
    Rajasekhar G.
    Dhanala S.
    Vura N.
    Ramisetty S.
    [J]. Journal of Maxillofacial and Oral Surgery, 2015, 14 (4) : 979 - 985
  • [7] A comparison of three treatment methods for fractures of the mandibular angle
    Kim, M-Y.
    Kim, C-H.
    Han, S-J.
    Lee, J-H.
    [J]. INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2016, 45 (07) : 878 - 883
  • [8] Biomechanical Study in Polyurethane Mandibles of Different Metal Plates and Internal Fixation Techniques, Employed in Mandibular Angle Fractures
    Semeghini Guastaldi, Fernando Pozzi
    Hochuli-Vieira, Eduardo
    Guastaldi, Antonio Carlos
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2014, 25 (06) : 2246 - 2250
  • [9] Application of 2.0 mm Titanium Plates in Rigid Internal Fixation of Mandibular Angle Fractures
    Sabhlok S.
    Waknis P.
    Bhagwat A.
    [J]. Journal of Maxillofacial and Oral Surgery, 2010, 9 (4) : 339 - 343
  • [10] Fixation of mandibular fractures with biodegradable plates and screws
    Yerit, KC
    Enislidis, G
    Schopper, C
    Turhani, D
    Wanschitz, F
    Wagner, A
    Watzinger, F
    Ewers, R
    [J]. ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2002, 94 (03): : 294 - 300