Comparison of resorbable plates and titanium plates for fixation stability of combined mandibular symphysis and angle fractures

被引:8
|
作者
Lim, Ho-Yong [1 ]
Jung, Chang-Hwa [1 ]
Kim, Seong-Yong [1 ]
Cho, Jin-Yong [1 ]
Ryu, Jae-Young [1 ]
Kim, Hyeon-Min [1 ]
机构
[1] Gachon Univ, Gil Med Ctr, Dept Oral & Maxillofacial Surg, 21 Namdong Daero 774Beon Gil, Incheon 405760, South Korea
关键词
Mandibular fracture; Titanium; Resorbable;
D O I
10.5125/jkaoms.2014.40.6.285
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: We compared resorbable plates with titanium plates for treatment of combined mandibular angle and symphyseal fractures. Materials and Methods: Patients with mandibular angle and symphysis fractures were divided into two groups. The control (T) group received titanium plates while the experimental (R) group received resorbable plates. All procedures were carried out under general anesthesia using standard surgical techniques. We compared the frequency of wound dehiscence, development of infection, malocclusion, malunion, screw breakage, and any other technical difficulties between the two groups. Results: Thirteen patients were included in the R group, where 39 resorbable plates were applied. The T group consisted of 16 patients who received 48 titanium plates. The mean age in the R and T groups was 28.29 and 24.23 years, respectively. Primary healing of the fractured mandible was obtained in all patients in both groups. Postoperative complications were minor and transient. Moreover, there were no significant differences in the rates of various complications between the two groups. Breakage of 3 screws during the perioperative period was seen in the R group, while no screws or plates were broken in the T group. Conclusion: Resorbable plates can be used to stabilize combined mandibular angle and symphysis fractures.
引用
收藏
页码:285 / 290
页数:6
相关论文
共 50 条
  • [1] Comparison of Fixation Stability of Resorbable Versus Titanium Plate and Screws in Mandibular Angle Fractures
    Bayram, Burak
    Araz, Kenan
    Uckan, Sina
    Balcik, Cenk
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 67 (08) : 1644 - 1648
  • [2] Salvation for Hardware Failure of Resorbable Fixation Plates in Bilateral Mandibular Angle Fractures
    Kwon, Yong-Dae
    Lee, Deok-Won
    Choi, Byung-Joon
    Suh, Joon-Ho
    Ko, Sujin
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2012, 23 (04) : 1192 - 1194
  • [3] Resorbable plates for the fixation of mandibular fractures: A prospective study
    Laughlin, Robert M.
    Block, Michael S.
    Wilk, Randall
    Malloy, Randolph B.
    Kent, John N.
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2007, 65 (01) : 89 - 96
  • [4] Resorbable Plates for Fixation of Complicated Mandibular Fractures in Children
    Ongodia D.
    Li Z.
    Xing W.Z.
    Li Z.-B.
    [J]. Journal of Maxillofacial and Oral Surgery, 2014, 13 (2) : 99 - 103
  • [5] Resorbable Plates for the Fixation of Isolated Mandibular Angle Fracture
    Yang, Rong-tao
    Lv, Kun
    Zhou, Hai-hua
    Li, Zhi
    Li, Zu-bing
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2015, 26 (02) : 447 - 448
  • [6] Biomechanical analysis of titanium fixation plates and screws in mandibular angle fractures
    Atik, F.
    Atac, M. S.
    Ozkan, A.
    Kilinc, Y.
    Arslan, M.
    [J]. NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2016, 19 (03) : 386 - 390
  • [7] Computerized analysis of resorbable polymer plates and screws for the rigid fixation of mandibular angle fractures
    Cox, T
    Kohn, MW
    Impelluso, T
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2003, 61 (04) : 481 - 487
  • [8] Aesthetic and Functional Outcome of Zygomatic Fractures Fixation Comparison With Resorbable Versus Titanium Plates
    Mahmoud, Shadad M.
    Liao, Han-Tsung
    Chen, Chien-Tzung
    [J]. ANNALS OF PLASTIC SURGERY, 2016, 76 : S85 - S90
  • [9] Comparison of Mechanical Stabilization of the Mandibular Angulus Fracture Fixation, With Titanium Plates and Screws, Resorbable Plates and Screws, and Bone Adhesives
    Cural, Ulkem
    Atalay, Berkem
    Yildirim, Mustafa Sami
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2018, 29 (07) : 1780 - 1787
  • [10] Application of 2.0 mm Titanium Plates in Rigid Internal Fixation of Mandibular Angle Fractures
    Sabhlok S.
    Waknis P.
    Bhagwat A.
    [J]. Journal of Maxillofacial and Oral Surgery, 2010, 9 (4) : 339 - 343