Volumetric modulated arc therapy versus conventional intensity modulated radiation therapy for stereotactic spine radiotherapy: A planning study and early clinical data

被引:59
|
作者
Kuijper, Ingrid T. [1 ]
Dahele, Max [1 ]
Senan, Suresh [1 ]
Verbakel, Wilko F. A. R. [1 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Radiat Oncol, NL-1081 HV Amsterdam, Netherlands
关键词
Stereotactic radiotherapy; Vertebral metastases; Volumetric modulated arc radiotherapy; Intensity-modulated radiotherapy; CORD COMPRESSION; RANDOMIZED-TRIAL; RADIOSURGERY; METASTASES; CANCER; COLUMN;
D O I
10.1016/j.radonc.2009.12.027
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Outcomes for selected patients with spinal metastases may be improved by dose escalation using stereotactic body radiation therapy (SBRT). As target geometry is complex, we compared SBRT plans using volumetric modulated arc radiotherapy (RapidArc (R), RA) and conventional intensity-modulated radiotherapy (IMRT). Materials and methods Materials and methods: RA and IMRT plans to deliver a fraction of 16 Gy to at least 90% of planning target volume (PTV) were compared for PTV coverage, normal organ sparing and estimated delivery times. Group 1 consisted of PTVs to only vertebral body (n = 3), while group 2 had PTVs encompassing the entire vertebra (n = 4). Finally, RA delivery parameters in four patients were assessed. Results: Both techniques delivered 16 Gy to a mean of 95% and 85% of the PTV in groups 1 and 2, respectively. Spinal cord sparing was comparable; mean V(10-partial cord) for RA and IMRT in group 1 was 3.6%, and was 9.4% versus 11.5%, respectively, in group 2. Estimated mean treatment times for RA with 2-3 arcs and IMRT were comparable. Clinical RA beam-on times ranged from 11 to 15.4 min. Conclusions: Both RA and conventional IMRT plans deliver high quality vertebral SBRT, but plan quality was poorer when the PTV consisted of the entire vertebra. (C) 2010 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and Oncology 94 (2010) 224-228
引用
收藏
页码:224 / 228
页数:5
相关论文
共 50 条
  • [31] VOLUMETRIC ARC INTENSITY-MODULATED THERAPY FOR SPINE BODY RADIOTHERAPY: COMPARISON WITH STATIC INTENSITY-MODULATED TREATMENT
    Wu, Q. Jackie
    Yoo, Sua
    Kirkpatrick, John P.
    Thongphiew, Danthai
    Yin, Fang-Fang
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (05): : 1596 - 1604
  • [32] Comparative dosimetric analysis of intensity modulated radiotherapy and volumetric modulated arc therapy
    Ali, M. Athar
    Babaiah, M.
    George, G.
    Madhusudhan, N.
    Ramalingam, K.
    Kumar, S. Ashok
    Karthikeyan, K.
    Anantharaman, A.
    Shanu, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S188 - S188
  • [33] A Planning Benchmark Study for SBRT: Helical Radiation Therapy Versus Intensity Modulated Arc Therapy
    Maha, F. Ebrahimi Tazeh
    Moustakis, C.
    Haverkamp, U.
    Eich, H. T. T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E655 - E656
  • [34] Comparison of spinal Stereotactic Body Radiotherapy (SBRT) planning techniques: intensity-modulated radiation therapy, modulated arc therapy, and helical tomotherapy
    Kim, Jina
    Jang, Hong Seok
    Kim, Yeon Sil
    Choi, Byung Ock
    Kang, Young-nam
    [J]. MEDICAL DOSIMETRY, 2017, 42 (03) : 210 - 215
  • [35] Evaluation of Spine Stereotactic Body Radiotherapy (SBRT) Using Volumetric Modulated Arc Therapy (Rapid Arc)
    Sharma, P. K.
    Srivastava, R.
    Noufal, M.
    Garg, M.
    Nangia, S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S875 - S875
  • [36] Dosimetric Comparison of Volumetric Modulated Arc Therapy and Helical Tomotherapy for Spine Stereotactic Body Radiation Therapy
    Dandekar, V. K.
    Young, J. L.
    Turian, J.
    Diaz, A.
    Sher, D. J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S725 - S725
  • [37] The role of volumetric modulated arc therapy (VMAT) in gynaecological radiation therapy: A dosimetric comparison of intensity modulated radiation therapy versus VMAT
    Knapp, Penelope
    Eva, Belinda
    Reseigh, Gemma
    Gibbs, Adrian
    Sim, Lucy
    Daly, Tiffany
    Cox, Judith
    Bernard, Anne
    [J]. JOURNAL OF MEDICAL RADIATION SCIENCES, 2019, 66 (01) : 44 - 53
  • [38] VOLUMETRIC MODULATED INTENSITY MODULATED ARC THERAPY FOR STEREOTACTIC LUNG RADIOTHERAPY: CONSTANT DOSE RATE VERSUS VARIABLE DOSE RATE DELIVERY
    Dogan, N.
    Wu, Y.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S499 - S499
  • [39] FusionArc optimization: A hybrid volumetric modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) planning strategy
    Matuszak, Martha M.
    Steers, Jennifer M.
    Long, Troy
    McShan, Daniel L.
    Fraass, Benedick A.
    Romeijn, H. Edwin
    Ten Haken, Randall K.
    [J]. MEDICAL PHYSICS, 2013, 40 (07)
  • [40] VOLUMETRIC-MODULATED ARC THERAPY FOR STEREOTACTIC BODY RADIOTHERAPY OF LUNG TUMORS: A COMPARISON WITH INTENSITY-MODULATED RADIOTHERAPY TECHNIQUES
    Holt, Andrea
    van Vliet-Vroegindeweij, Corine
    Mans, Anton
    Belderbos, Jose S.
    Damen, Eugene M. F.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (05): : 1560 - 1567