Left Ventricular Assist Device Inflow Cannula Angle and Thrombosis Risk

被引:73
|
作者
Chivukula, Venkat Keshav [2 ]
Beckman, Jennifer A. [1 ]
Prisco, Anthony R. [4 ]
Dardas, Todd [1 ]
Lin, Shin [1 ]
Smith, Jason W. [3 ]
Mokadam, Nahush A. [3 ]
Aliseda, Alberto [2 ]
Mahr, Claudius [1 ]
机构
[1] Univ Washington, Div Cardiol, 1959 NE Pacific St, Seattle, WA 98195 USA
[2] Univ Washington, Dept Mech Engn, Seattle, WA 98195 USA
[3] Univ Washington, Div Cardiothorac Surg, Seattle, WA 98195 USA
[4] Univ Minnesota, Dept Med, Box 736 UMHC, Minneapolis, MN 55455 USA
关键词
cannula; heart assist devices; heart ventricles; stroke; thrombosis; CONTINUOUS-FLOW; PUMP; MANAGEMENT; TRANSPLANT; THERAPY; SYSTEM; BRIDGE;
D O I
10.1161/CIRCHEARTFAILURE.117.004325
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: As heart failure prevalence continues to increase in the setting of a static donor supply, left ventricular assist device (LVAD) therapy for end-stage heart failure continues to grow. Anecdotal evidence suggests that malalignment of the LVAD inflow cannula may increase thrombosis risk, but this effect has not been explored mechanistically or quantified statistically. Our objective is to elucidate the impact of surgical angulation of the inflow cannula on thrombogenicity. METHODS AND RESULTS: Unsteady computational fluid dynamics is used in conjunction with computational modeling and virtual surgery to model flow through the left ventricle for 5 different inflow cannula angulations. We use a holistic approach to evaluate thrombogenicity: platelet-based (Lagrangian) metrics to evaluate the platelet mechanical environment, combined with flow-based (Eulerian) metrics to investigate intraventricular hemodynamics. The thrombogenic potential of each LVAD inflow cannula angulation is quantitatively evaluated based on platelet shear stress history and residence time. Intraventricular hemodynamics are strongly influenced by LVAD inflow cannula angulation. Platelet behavior indicates elevated thrombogenic potential for certain inflow cannula angles, potentially leading to platelet activation. Our analysis demonstrates that the optimal range of inflow angulation is within 0 +/- 7 degrees of the left ventricular apical axis. CONCLUSIONS: Angulation of the inflow cannula >7 degrees from the apical axis (axis connecting mitral valve and ventricular apex) leads to markedly unfavorable hemodynamics as determined by computational fluid dynamics. Computational hemodynamic simulations incorporating Lagrangian and Eulerian metrics are a powerful tool for studying optimization of LVAD implantation strategies, with the long-term potential of improving outcomes.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] The influence of left ventricular assist device inflow cannula position on thrombosis risk
    Ghodrati, Mojgan
    Maurer, Alexander
    Schloeglhofer, Thomas
    Khienwad, Thananya
    Zimpfer, Daniel
    Beitzke, Dietrich
    Zonta, Francesco
    Moscato, Francesco
    Schima, Heinrich
    Aigner, Philipp
    ARTIFICIAL ORGANS, 2020, 44 (09) : 939 - 946
  • [2] Left Ventricular Assist Device Inflow Cannula Insertion Depth Influences Thrombosis Risk
    Chivukula, Venkat Keshav
    Beckman, Jennifer A.
    Li, Song
    Masri, S. Carolina
    Levy, Wayne C.
    Lin, Shin
    Cheng, Richard K.
    Farris, Stephen D.
    Wood, Gregory
    Dardas, Todd F.
    Kirkpatrick, James N.
    Koomalsingh, Kevin
    Zimpfer, Daniel
    Mackensen, G. Burkhard
    Chassagne, Fanette
    Mahr, Claudius
    Aliseda, Alberto
    ASAIO JOURNAL, 2020, 66 (07) : 766 - 773
  • [3] Understanding the influence of left ventricular assist device inflow cannula alignment and the risk of intraventricular thrombosis
    Neidlin, Michael
    Liao, Sam
    Li, Zhiyong
    Simpson, Benjamin
    Kaye, David M.
    Steinseifer, Ulrich
    Gregory, Shaun
    BIOMEDICAL ENGINEERING ONLINE, 2021, 20 (01)
  • [4] Understanding the influence of left ventricular assist device inflow cannula alignment and the risk of intraventricular thrombosis
    Michael Neidlin
    Sam Liao
    Zhiyong Li
    Benjamin Simpson
    David M. Kaye
    Ulrich Steinseifer
    Shaun Gregory
    BioMedical Engineering OnLine, 20
  • [5] Left ventricular assist device and pump thrombosis: the importance of the inflow cannula position
    Kortekaas, Kirsten A.
    de Graaf, Michiel A.
    Palmen, Meindert
    Braun, Jerry
    Mertens, Bart J. A.
    Tops, Laurens F.
    Beeres, Saskia L. M. A.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2022, 38 (12): : 2771 - 2779
  • [6] Left ventricular assist device and pump thrombosis: the importance of the inflow cannula position
    Kirsten A. Kortekaas
    Michiel A. de Graaf
    Meindert Palmen
    Jerry Braun
    Bart J. A. Mertens
    Laurens F. Tops
    Saskia L. M. A. Beeres
    The International Journal of Cardiovascular Imaging, 2022, 38 : 2771 - 2779
  • [7] Replacement of Left Ventricular Assist Device Inflow Cannula Using a Right Angle Atrial Cannula
    Greenleaf, Christopher E.
    Binsalamah, Ziyad M.
    Adachi, Iki
    ANNALS OF THORACIC SURGERY, 2020, 109 (05): : E389 - E389
  • [8] Left Ventricular Assist Device Inflow Cannula Position May Contribute to the Development of HeartMate II Left Ventricular Assist Device Pump Thrombosis
    Bhama, Jay K.
    Bansal, Aditya
    OCHSNER JOURNAL, 2018, 18 (02): : 131 - 135
  • [10] Inflow Cannula Proximity to Ventricular Myocardium on Cardiac CT is Associated with Left Ventricular Assist Device Pump Thrombosis
    Vallabhaneni, Srilakshmi
    Kancherla, Kalyan
    Ahmed, Sara
    Najjar, Samer S.
    Majure, David T.
    Weissman, Gaby
    JOURNAL OF CARDIAC FAILURE, 2016, 22 (08) : S23 - S23