Sorting a large set of heavily used LiF:Mg,Ti thermoluminescent detectors into repeatable subsets of similar response

被引:8
|
作者
Kearfott, Kimberlee J. [1 ]
Newton, Jill P. [1 ]
Rafique, Muhammad [1 ,2 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
[2] Univ Azad Jammu & Kashmir Muzaffarbad, Dept Phys, Azad Kashmir 13100, Pakistan
关键词
Thermoluminescent dosimeters; Cs-137; source; Sensitivity; Statistics; Calibration; PEAK FADING PROPERTIES; DOSIMETERS; PERFORMANCE; CALIBRATION; READINGS; DESIGN; SYSTEM; DRIFT; CT;
D O I
10.1016/j.apradiso.2014.10.019
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A set of 920 heavily used LiF:Mg,Ti thermoluminescent dosimeters (TLDs) was placed into a polymethyl methacrylate (PMMA) plate attached to a 40 x 40 x 15 cm(3) PMMA phantom and irradiated to 4.52 mGy using a Cs-137 source. This was repeated three times to determine the mean and standard deviation of each TLD's sensitivity. Reader drift was tracked over time with 10 control dosimeters. Two test sets of 100 TLDs were divided into subsets with sensitivities within +/- 1% of their subset means. All dosimeters were re-irradiated four times to test the TLDs' response repeatability and determine the sensitivity uniformity within the subsets. Coefficients of variation revealed that, within a given subset, the dosimeters responded within +/- 2.5% of their subset mean in all calibrations. The coefficient of variation in any of the 200 TLDs' calibrations was below 6% across the four calibrations. The work validates the approach of performing three calibrations to separate heavily used and aged TLDs with overall sensitivity variations of +/- 25% into subsets that reproducibly respond within +/- 2.5%. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:180 / 187
页数:8
相关论文
共 37 条
  • [1] Thermoluminescent response of LiF:Mg,Ti to 20 keV electrons
    Mercado-Uribe, H
    Brandan, ME
    SOLID STATE DOSIMETRY, PTS 1 AND 2, PROCEEDINGS, 2002, : 179 - 182
  • [2] Thermoluminescent response of LiF:Mg,Ti to 20 keV electrons
    Mercado-Uribe, H
    Brandan, ME
    RADIATION PROTECTION DOSIMETRY, 2002, 100 (1-4) : 179 - 182
  • [3] Thermoluminescent response of LiF:Mg,Ti to low energy electrons
    Mercado-Uribe, H
    Brandan, ME
    MEDICAL PHYSICS, 2000, 538 : 235 - 239
  • [4] PASSIVE NEUTRON SPECTROMETER AND DOSIMETER USING LIF = MG, TI THERMOLUMINESCENT DETECTORS
    SINGH, D
    BURGKHARDT, B
    PIESCH, E
    NUCLEAR INSTRUMENTS & METHODS, 1977, 142 (03): : 409 - 415
  • [5] Glow curve deconvolution for the routine readout of LiF:Mg,Ti thermoluminescent detectors
    Stadtmann, H.
    Wilding, G.
    RADIATION MEASUREMENTS, 2017, 106 : 278 - 284
  • [6] Photon energy response of LiF:Mg,Ti (MTS) and LiF:Mg,Cu,P (MCP) thermoluminescent detectors: Experimental measurements and microdosimetric modeling
    Parisi, Alessio
    Dabin, Jeremie
    Schoonjans, Werner
    Van Hoey, Olivier
    Megret, Patrice
    Vanhavere, Filip
    RADIATION PHYSICS AND CHEMISTRY, 2019, 163 : 67 - 73
  • [7] Behaviour of LiF:Mg,Cu,P and LiF:Mg,Ti thermoluminescent detectors for electron doses up to 1 MGy
    Bilski, Pawel
    Obryk, Barbara
    Stuglik, Zofia
    RADIATION MEASUREMENTS, 2010, 45 (3-6) : 576 - 578
  • [8] The Response of LiF:Mg, Ti Thermoluminescent Dosimeters to Low-Energy Photons
    Rasmussen, B.
    Davis, S.
    Micka, J.
    DeWerd, L.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [9] The investigation of LiF:Mg,Cu,P thermoluminescent detectors; The dose response to gamma radiation
    Ramlo, MJ
    Bicron, NE
    Karpov, NA
    IRPA9 - 1996 INTERNATIONAL CONGRESS ON RADIATION PROTECTION / NINTH INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, PROCEEDINGS, VOL 4, 1996, : D421 - D421
  • [10] Fading study and readout optimization for routinely use of LiF:Mg,Ti thermoluminescent detectors for personal dosimetry
    Sorger, D.
    Stadtmann, H.
    Sprengel, W.
    RADIATION MEASUREMENTS, 2020, 135 (135)