Data analysis of dose map verification for IMRT

被引:0
|
作者
Xu Yingjie [1 ]
Dai Jianrong [1 ]
Fu Guishan [1 ]
Zhang Yongqian [1 ]
Zhang Ke [1 ]
Cui Weijie [1 ]
Zhang Dandan [1 ]
Jiang Bo [1 ]
机构
[1] Chinese Acad Med Sci, Dept Radiat Oncol, Inst Canc, Beijing 100021, Peoples R China
关键词
Dose map verification; Intensity-modulated radiation therapy; Beam modeling; MODULATED RADIATION-THERAPY; QUALITY-ASSURANCE; DIODE-ARRAY;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this paper we analyze the results of dose map verifications for patient's IMRT (Intensity-modulated radiation therapy) plans and study the factors that may influence the accuracy of verification. MapCHECK, a two-dimensional diode array, was used to measure the dose maps for 1242 plans (14540 fields) from May 2004 to August 2008. The plans were designed with Pinnacle 3 planning system. The passing rate of beams was determined when the acceptance criterion was 2%/2 mm, 3%/3 mm and 4%/4 mm. And the data with 3%/3 mm criteria was analyzed in more detail. The considered factors were beam modeling, optimization mode and treatment site. The median passing rate of total fields was 93.5%, 98.8%, and 100% when the acceptance criterion was 2%/2 mm, 3%/3 mm. and 4%/4 mm, and the interquartile range were 11.1%, 3.8%, and 1.3%, respectively. The results of direct machine parameter optimization (DMPO) planning mode was better than those of multiple-step mode and beam modeling of new accelerators was better than that of old accelerators. These indicate that beam modeling is the key point of improving passing rate of IMRT verification and the influence of treatment site was little. The factors, the total number of segments, minimum area of segments and minimum monitor unit (MU) of segments, also influence the dosimetric accuracy of IMRT plan verification.
引用
收藏
页码:297 / 301
页数:5
相关论文
共 50 条
  • [1] Data analysis of dose map verification for IMRT
    XU Yingjie DAI Jianrong FU Guishan ZHANG Yongqian ZHANG Ke CUI Weijie ZHANG Dandan JIANG Bo Department of Radiation Oncology
    Nuclear Science and Techniques, 2009, 20 (05) : 297 - 301
  • [2] The Analysis of Dose Verification Results for 924 IMRT Plans
    Wu, S.
    Chen, J.
    Li, Z.
    Li, C.
    Yin, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E284 - E284
  • [3] The Quantitative Analysis of Dose Verification Results for IMRT Plans
    Wu, S.
    Chen, J.
    Li, Z.
    Qiu, Q.
    Wang, X.
    Li, C.
    Yin, Y.
    MEDICAL PHYSICS, 2019, 46 (06) : E611 - E611
  • [4] Dose Verification in IMRT and VMAT
    Feygelman, Vladimir
    Nelms, Benjamin E.
    CONCEPTS AND TRENDS IN MEDICAL RADIATION DOSIMETRY, 2011, 1345 : 145 - +
  • [5] Alanine dosimetry in dose verification for IMRT
    Schaeken, B.
    De Ost, B.
    Van Gestel, D.
    Meijnders, P.
    Jacobs, I.
    Goor, C.
    Vanderkam, S.
    Coelmont, A.
    Van den Weyngaert, D.
    Schaeken, B.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S365 - S366
  • [6] Dose Verification of IMRT Plans: Reconsideration of Acceptance Criteria and Analysis Method
    Rezaee, M.
    Wong, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E609 - E609
  • [7] Dose verification in clinical IMRT prostate incidents
    Sandilos, P
    Angelopoulos, A
    Baras, P
    Dardoufas, K
    Karaiskos, P
    Kipouros, P
    Kozicki, M
    Rosiak, JM
    Sakelliou, L
    Seimenis, I
    Vlahos, L
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (05): : 1540 - 1547
  • [8] Dose verification for rotating multileaf collimator IMRT
    Schmuland, M
    Otto, K
    MEDICAL PHYSICS, 2005, 32 (07) : 2425 - 2425
  • [9] A quality assurance phantom for IMRT dose verification
    Ma, CM
    Jiang, SB
    Pawlicki, T
    Chen, Y
    Li, JS
    Deng, J
    Lee, MC
    Mok, E
    Boyer, AL
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 1180 - 1183
  • [10] Practical dose delivery verification of craniospinal IMRT
    Lee, Young K.
    Kim, Anthony T.
    Zhao, Peiying
    Karotki, Aliaksandr
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (06): : 76 - 83