Data analysis of dose map verification for IMRT

被引:0
|
作者
Xu Yingjie [1 ]
Dai Jianrong [1 ]
Fu Guishan [1 ]
Zhang Yongqian [1 ]
Zhang Ke [1 ]
Cui Weijie [1 ]
Zhang Dandan [1 ]
Jiang Bo [1 ]
机构
[1] Chinese Acad Med Sci, Dept Radiat Oncol, Inst Canc, Beijing 100021, Peoples R China
关键词
Dose map verification; Intensity-modulated radiation therapy; Beam modeling; MODULATED RADIATION-THERAPY; QUALITY-ASSURANCE; DIODE-ARRAY;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this paper we analyze the results of dose map verifications for patient's IMRT (Intensity-modulated radiation therapy) plans and study the factors that may influence the accuracy of verification. MapCHECK, a two-dimensional diode array, was used to measure the dose maps for 1242 plans (14540 fields) from May 2004 to August 2008. The plans were designed with Pinnacle 3 planning system. The passing rate of beams was determined when the acceptance criterion was 2%/2 mm, 3%/3 mm and 4%/4 mm. And the data with 3%/3 mm criteria was analyzed in more detail. The considered factors were beam modeling, optimization mode and treatment site. The median passing rate of total fields was 93.5%, 98.8%, and 100% when the acceptance criterion was 2%/2 mm, 3%/3 mm. and 4%/4 mm, and the interquartile range were 11.1%, 3.8%, and 1.3%, respectively. The results of direct machine parameter optimization (DMPO) planning mode was better than those of multiple-step mode and beam modeling of new accelerators was better than that of old accelerators. These indicate that beam modeling is the key point of improving passing rate of IMRT verification and the influence of treatment site was little. The factors, the total number of segments, minimum area of segments and minimum monitor unit (MU) of segments, also influence the dosimetric accuracy of IMRT plan verification.
引用
收藏
页码:297 / 301
页数:5
相关论文
共 50 条
  • [31] Dosimetry Investigation of MOSFET for Clinical IMRT Dose Verification
    Deshpande, Sudesh
    Kumar, Rajesh
    Ghadi, Yogesh
    Nehru, R. M.
    Kannan, V.
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2013, 12 (03) : 193 - 198
  • [32] Impact of normalisation point on gamma factor analysis for single field IMRT dose verification
    Vollans, E
    Hilts, M
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S104 - S104
  • [33] The Impact of Measurement Uncertainty On the Dose Map Analysis of VMAT Verification Plans
    Sandhu, G.
    Kohli, K.
    Szpala, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 2894 - 2894
  • [34] A Fourier analysis of the dose grid resolution required for accurate IMRT fluence map optimization
    Dempsey, JF
    Romeijn, HE
    Li, JG
    Low, DA
    Palta, JR
    MEDICAL PHYSICS, 2005, 32 (02) : 380 - 388
  • [35] Surface dose prediction and verification for IMRT plans using line dose profiles
    Berg, R
    Klash, S
    Gossman, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S590 - S590
  • [36] Intensity map verification for IMRT QA using an a-Si EPID
    Lometti, M.
    Morin, O.
    Bellerose, J.
    Malfatti, C.
    Aubin, M.
    Pouliot, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2113 - 2113
  • [37] Verification of planned vs delivered dose for narrow IMRT segments
    Chakarova, R.
    Lindberg, A.
    Bankvall, G.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S201 - S201
  • [38] Retrospective Multicenter Study of IMRT Absorbed Dose Verification in Japan
    Kurooka, M.
    Kawachi, T.
    Tohyama, N.
    Kojima, T.
    Kumazaki, Y.
    Kitou, S.
    Okamoto, H.
    Hashimoto, S.
    Fujita, Y.
    Hayashi, N.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [39] Characterization and use of EBT radiochromic film for IMRT dose verification
    Zeidan, O.
    Stephenson, S.
    Meeks, S.
    Willoughby, T.
    Wagner, T.
    Kupelian, P.
    Langen, K.
    MEDICAL PHYSICS, 2006, 33 (06) : 2077 - 2077
  • [40] Results of a one year study of IMRT dose verification with RadCalc
    Krayenbuehl, J
    Lutters, GR
    Davis, JB
    STRAHLENTHERAPIE UND ONKOLOGIE, 2005, 181 : 144 - 144