Corneal Parameters in Healthy Subjects Assessed by Corvis ST

被引:22
|
作者
Salouti, Ramin [1 ,2 ]
Bagheri, Mansoureh [1 ]
Shamsi, Anis [1 ]
Zamani, Mohammad [2 ]
Ghoreyshi, Maryam [3 ]
Nowroozzadeh, M. Hossein [1 ]
机构
[1] Shiraz Univ Med Sci, Poostchi Ophthalmol Res Ctr, Sch Med, Dept Ophthalmol, Shiraz, Iran
[2] Salouti Eye Clin, Salouti Cornea Res Ctr, Shiraz, Iran
[3] Shiraz Univ Med Sci, Hlth Policy Res Ctr, Shiraz, Iran
关键词
Central Corneal Thickness; Corneal Biomechanics; Corvis ST; Intraocular Pressure; INTRAOCULAR-PRESSURE; BIOMECHANICAL METRICS; NONCONTACT; THICKNESS; TONOMETER;
D O I
10.18502/jovr.v15i1.5936
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate corneal biomechanics using Corvis ST in healthy eyes from Iranian keratorefractive surgery candidates. Methods: In this prospective consecutive observational case series, the intraocular pressure (IOP), central corneal thickness (CCT), and biomechanical properties of 1,304 eyes from 652 patients were evaluated using Corvis ST. Keratometric readings and manifest refraction were also recorded. Results: The mean (+/- SD) age of participants was 28 +/- 5 years, and 31.7% were male. The mean spherical equivalent refraction was -3.50 +/- 1.57 diopters (D), the mean IOP was 16.8 +/- 2.9 mmHg, and the mean CCT was 531 +/- 31.. m for the right eye. The respective means (+/- SD) corneal biomechanical parameters of the right eye were as follows: first applanation time: 7.36 +/- 0.39 milliseconds (ms); first applanation length: 1.82 +/- 0.22 mm; velocity in: 0.12 +/- 0.04 m/s; second applanation time: 20.13 +/- 0.48 ms; second applanation length: 1.34 +/- 0.55 mm; velocity out: -0.67 +/- 0.17 m/s; total time: 16.84 +/- 0.64 ms; deformation amplitude: 1.05 +/- 0.10 mm; peak distance: 4.60 +/- 1.01 mm; and concave radius of curvature: 7.35 +/- 1.39 mm. In the linear regression analysis, IOP exhibited a statistically significant association with the first and second applanation times, total time, velocity in, peak distance, deformation amplitude, and concave radius of curvature. Conclusion: Our study results can be used as a reference for the interpretation of Corvis ST parameters in healthy refractive surgery candidates in the Iranian population. Our results confirmed that IOP is a major determinant of Corvis parameters.
引用
收藏
页码:24 / 31
页数:8
相关论文
共 50 条
  • [1] Corneal Deformation Parameters Provided by the Corvis- ST Pachy- Tonometer in Healthy Subjects and Glaucoma Patients
    Salvetat, Maria L.
    Zeppieri, Marco
    Tosoni, Claudia
    Felletti, Mirella
    Grasso, Leonardo
    Brusini, Paolo
    JOURNAL OF GLAUCOMA, 2015, 24 (08) : 568 - 574
  • [2] Repeatability and Reproducibility of Intraocular Pressure and Dynamic Corneal Response Parameters Assessed by the Corvis ST
    Lopes, Bernardo T.
    Roberts, Cynthia J.
    Elsheikh, Ahmed
    Vinciguerra, Riccardo
    Vinciguerra, Paolo
    Reisdorf, Sven
    Berger, Stefanie
    Koprowski, Robert
    Ambrosio, Renato, Jr.
    JOURNAL OF OPHTHALMOLOGY, 2017, 2017
  • [3] Assessment of Corneal Biomechanical Properties by CorVis ST in Patients with Dry Eye and in Healthy Subjects
    Long, Qin
    Wang, Jingyi
    Yang, Xue
    Jin, Yumei
    Ai, Fengrong
    Li, Ying
    JOURNAL OF OPHTHALMOLOGY, 2015, 2015
  • [4] Corneal Biomechanical Metrics by Corvis ST in a Healthy Mexican Population
    Fabregas-Sanchez-Woodworth, Daniela
    Ramirez-Miranda, Arturo J.
    Mordani, Simran Mangwani
    Rivera, Jose Y. Arteaga
    Jimenez-Collado, David
    Navas, Alejandro
    Graue-Hernandez, Enrique O.
    Valdez, Jorge E.
    Mendizabal-Velazquez, Alondra
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [5] Corneal biomechanics. Corvis® ST parameters after LASIK
    Frings, A.
    Linke, S. J.
    Bauer, E. L.
    Druchkiv, V.
    Katz, T.
    Steinberg, J.
    OPHTHALMOLOGE, 2015, 112 (09): : 740 - 745
  • [6] Relationship of Corneal Biomechanics Measured by Corvis ST and Optic Nerve Head Parameters in Healthy Saudi Females
    Elagamy, Amira
    Alnasser, Hessa Khaled
    Alghamdi, Wafa Saeed
    Berika, Mohamed
    Aldisi, Dara
    CLINICAL OPHTHALMOLOGY, 2024, 18 : 2851 - 2863
  • [7] Corneal Deformation Response in Patients With Primary Open-Angle Glaucoma and in Healthy Subjects Analyzed by Corvis ST
    Wang, Wei
    Du, Shaolin
    Zhang, Xiulan
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (09) : 5557 - 5565
  • [8] Corneal biomechanical metrics of healthy Chinese adults using Corvis ST
    Wang, Wei
    He, Miao
    He, Hong
    Zhang, Chi
    Jin, He
    Zhong, Xingwu
    CONTACT LENS & ANTERIOR EYE, 2017, 40 (02): : 97 - 103
  • [9] Assessment of corneal biomechanical parameters in myopes and emmetropes using the Corvis ST
    Lee, Rachel
    Chang, Robert T.
    Wong, Ian Y. H.
    Lai, Jimmy S. M.
    Lee, Jacky W. Y.
    Singh, Kuldev
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2016, 99 (02) : 157 - 162
  • [10] Repeatability and reproducibility of corneal biomechanical parameters derived from Corvis ST
    Serbecic, Nermin
    Beutelspacher, Sven
    Markovic, Lovro
    Roy, Abhijit Sina
    Shetty, Rohit
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2020, 30 (06) : 1287 - 1294