Measurement Errors Related to Contact Angle Analysis of Hydrogel and Silicone Hydrogel Contact Lenses

被引:44
|
作者
Read, Michael L. [1 ]
Morgan, Philip B. [1 ]
Maldonado-Codina, Carole [1 ]
机构
[1] Univ Manchester, Fac Life Sci, Manchester, Lancs, England
基金
英国医学研究理事会;
关键词
contact angle; contact lens; hydrogel; silicone hydrogel; measurement error; coefficient of repeatability; BACTERIAL ADHESION; WETTABILITY; WEAR; WORN;
D O I
10.1002/jbm.b.31442
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work sought to undertake a comprehensive investigation of the measurement errors associated with contact angle assessment of curved hydrogel contact lens surfaces. The contact angle coefficient of repeatability (COR) associated with three measurement conditions (image analysis COR, intralens COR, and interlens COR) was determined by measuring the contact angles (using both sessile drop and captive bubble methods) for three silicone hydrogel lenses (senofilcon A, balafilcon A, lotrafilcon A) and one conventional hydrogel lens (etafilcon A). Image analysis COR values were about 2 degrees, whereas intralens COR values (95% confidence intervals) ranged from 4.0 degrees (3.3 degrees, 4.7 degrees) (lotrafilcon A, captive bubble) to 10.2 degrees (8.4 degrees, 12.1 degrees) (senofilcon A, sessile drop). Interlens COR values ranged from 4.5 degrees (3.7 degrees, 5.2 degrees) (lotrafilcon A, captive bubble) to 16.5 degrees (13.6 degrees, 19.4 degrees) (senofilcon A, sessile drop). Measurement error associated with image analysis was shown to be small as an absolute measure, although proportionally more significant for lenses with low contact angle. Sessile drop contact angles were typically less repeatable than captive bubble contact angles. For sessile drop measures, repeatability was poorer with the silicone hydrogel lenses when compared with the conventional hydrogel lens; this phenomenon was not observed for the captive bubble method, suggesting that methodological factors related to the sessile drop technique (such as surface dehydration and blotting) may play a role in the increased variability of contact angle measurements observed with silicone hydrogel contact lenses. (C) 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 91B: 662-668, 2009
引用
收藏
页码:662 / 668
页数:7
相关论文
共 50 条
  • [31] The impact of silicone hydrogel contact lenses on the measurement of intraocular pressure using non-contact tonometry
    Pesic, Snezana
    Jovanovic, Svetlana
    Mitrasevic, Milos
    Vuletic, Biljana
    Jovanovic, Milena
    Jovanovic, Zorica
    VOJNOSANITETSKI PREGLED, 2017, 74 (08) : 763 - 766
  • [32] Extended cyclosporine delivery by silicone-hydrogel contact lenses
    Peng, Cheng-Chun
    Chauhan, Anuj
    JOURNAL OF CONTROLLED RELEASE, 2011, 154 (03) : 267 - 274
  • [33] Impact of Cosmetics on the Surface Properties of Silicone Hydrogel Contact Lenses
    Srinivasan, Sruthi
    Otchere, Heinz
    Yu, Mili
    Yang, Jeffery
    Luensmann, Doerte
    Jones, Lyndon
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (04): : 228 - 235
  • [34] Bacterial populations on silicone hydrogel and hydrogel contact lenses after swimming in a chlorinated pool
    Choo, J
    Vuu, K
    Bergenske, P
    Burnham, K
    Smythe, J
    Caroline, P
    OPTOMETRY AND VISION SCIENCE, 2005, 82 (02) : 134 - 137
  • [35] Changes in Ultraviolet Transmittance of Hydrogel and Silicone-Hydrogel Contact Lenses Induced by Wear
    Osuagwu, Uchechukwu L.
    Ogbuehi, Kelechi C.
    AlMubrad, Turki M.
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2014, 40 (01): : 28 - 36
  • [36] In Response: Changes in Ultraviolet Transmittance of Hydrogel and Silicone Hydrogel Contact Lenses Induced by Wear
    Uchechukwu, Osuagwu L.
    Ogbuehi, Kelechi C.
    Almubrad, Turki M.
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (04): : 252 - 254
  • [37] Continuous wear silicone hydrogel contact lenses and microbial keratitis
    M A N Whiting
    M K Raynor
    P B Morgan
    P Galloway
    D M Tole
    A Tullo
    Eye, 2004, 18 : 935 - 937
  • [38] Extended elution of phospholipid from silicone hydrogel contact lenses
    Pitt, William G.
    Zhao, Yibei
    Jack, Daniel R.
    Perez, Krystian X.
    Jones, Peter W.
    Marelli, Ryan
    Nelson, Jared L.
    Pruitt, John D.
    JOURNAL OF BIOMATERIALS SCIENCE-POLYMER EDITION, 2015, 26 (04) : 224 - 234
  • [39] Modification of Silicone Hydrogel Contact Lenses for the Selective Adsorption of Ofloxacin
    Xiao, Wenke
    Zhang, Xiaojuan
    Hao, Lingyun
    Zhao, Yuan
    Lin, Qing
    Wang, Zhao
    Song, Wenli
    Liang, Dong
    JOURNAL OF NANOMATERIALS, 2022, 2022
  • [40] Continuous wear silicone hydrogel contact lenses and microbial keratitis
    Whiting, MAN
    Raynor, MK
    Morgan, PB
    Galloway, P
    Tole, DM
    Tullo, A
    EYE, 2004, 18 (09) : 935 - 937