Measurement Errors Related to Contact Angle Analysis of Hydrogel and Silicone Hydrogel Contact Lenses

被引:44
|
作者
Read, Michael L. [1 ]
Morgan, Philip B. [1 ]
Maldonado-Codina, Carole [1 ]
机构
[1] Univ Manchester, Fac Life Sci, Manchester, Lancs, England
基金
英国医学研究理事会;
关键词
contact angle; contact lens; hydrogel; silicone hydrogel; measurement error; coefficient of repeatability; BACTERIAL ADHESION; WETTABILITY; WEAR; WORN;
D O I
10.1002/jbm.b.31442
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work sought to undertake a comprehensive investigation of the measurement errors associated with contact angle assessment of curved hydrogel contact lens surfaces. The contact angle coefficient of repeatability (COR) associated with three measurement conditions (image analysis COR, intralens COR, and interlens COR) was determined by measuring the contact angles (using both sessile drop and captive bubble methods) for three silicone hydrogel lenses (senofilcon A, balafilcon A, lotrafilcon A) and one conventional hydrogel lens (etafilcon A). Image analysis COR values were about 2 degrees, whereas intralens COR values (95% confidence intervals) ranged from 4.0 degrees (3.3 degrees, 4.7 degrees) (lotrafilcon A, captive bubble) to 10.2 degrees (8.4 degrees, 12.1 degrees) (senofilcon A, sessile drop). Interlens COR values ranged from 4.5 degrees (3.7 degrees, 5.2 degrees) (lotrafilcon A, captive bubble) to 16.5 degrees (13.6 degrees, 19.4 degrees) (senofilcon A, sessile drop). Measurement error associated with image analysis was shown to be small as an absolute measure, although proportionally more significant for lenses with low contact angle. Sessile drop contact angles were typically less repeatable than captive bubble contact angles. For sessile drop measures, repeatability was poorer with the silicone hydrogel lenses when compared with the conventional hydrogel lens; this phenomenon was not observed for the captive bubble method, suggesting that methodological factors related to the sessile drop technique (such as surface dehydration and blotting) may play a role in the increased variability of contact angle measurements observed with silicone hydrogel contact lenses. (C) 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 91B: 662-668, 2009
引用
收藏
页码:662 / 668
页数:7
相关论文
共 50 条
  • [21] Diffusion of Antimicrobials Across Silicone Hydrogel Contact Lenses
    Zambelli, Alison M.
    Brothers, Kimberly M.
    Hunt, Kristin M.
    Romanowski, Eric G.
    Nau, Amy C.
    Dhaliwal, Deepinder K.
    Shanks, Robert M. Q.
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (05): : 277 - 280
  • [22] Mechanical Complications Induced by Silicone Hydrogel Contact Lenses
    Lin, Meng C.
    Yeh, Thao N.
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2013, 39 (01): : 115 - 124
  • [23] Concentrations of Silicon at the Surface of Silicone Hydrogel Contact Lenses
    Rex, Jessica
    Perry, Scott
    Lemp, Jessie
    Maissa, Cecile A.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [24] Refractive and biometric changes with silicone hydrogel contact lenses
    Santodomingo-Rubido, J
    Gilmartin, B
    Wolffsohn, J
    OPTOMETRY AND VISION SCIENCE, 2005, 82 (06) : 481 - 489
  • [25] Analysis of free volumes and light transmission in hydrogel and silicone-hydrogel polymer contact lenses
    Kocela, Agnieszka
    Miedzinski, Rafal
    Filipecka, Katarzyna
    Filipecki, Jacek
    OPTICA APPLICATA, 2016, 46 (01) : 35 - 45
  • [26] Micromechanical measurement of adhesion of dehydrating silicone hydrogel contact lenses to corneal tissue
    Zhu, Dongkai
    Liu, Yangping
    Gilbert, Jeremy L.
    ACTA BIOMATERIALIA, 2021, 127 : 242 - 251
  • [27] Influence of environmental factors in the in vitro dehydration of hydrogel and silicone hydrogel contact lenses
    Martin-Montanez, Vicente
    Lopez-Miguel, Alberto
    Arroyo, Cristina
    Mateo, Maria E.
    Gonzalez-Meijome, Jose M.
    Calonge, Margarita
    Gonzalez-Garcia, Maria J.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2014, 102 (04) : 764 - 771
  • [28] Changes in Ultraviolet Transmittance of Hydrogel and Silicone Hydrogel Contact Lenses Induced by Wear
    Shedden, Arthur
    Pall, Brian
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (01): : 64 - 64
  • [29] A comparative study of biweekly disposable contact lenses: silicone hydrogel versus hydrogel
    Cheung, Sin-wan
    Cho, Pauline
    Chan, Ben
    Choy, Camus
    Ng, Vincent
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2007, 90 (02) : 124 - 131
  • [30] Improving contact-lens related dryness symptoms with silicone hydrogel lenses
    Chalmers, Robin
    Long, Bill
    Dillehay, Sally
    Begley, Carolyn
    OPTOMETRY AND VISION SCIENCE, 2008, 85 (08) : 778 - 784