Improvements in Carotid Plaque Imaging Using a New Eight-Element Phased Array Coil at 3T

被引:37
|
作者
Balu, Niranjan [1 ]
Yarnykh, Vasily L. [1 ]
Scholnick, Joshua [2 ]
Chu, Baocheng [1 ]
Yuan, Chun [1 ]
Hayes, Cecil [1 ]
机构
[1] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
[2] Univ Washington, Dept Cardiol, Seattle, WA 98195 USA
基金
美国国家卫生研究院;
关键词
atherosclerosis; phased array coil; black blood MRI; vessel wall imaging; carotid bifurcation; vulnerable atheroma; HIGH-RESOLUTION MRI; IN-VIVO MRI; FIBROUS CAP; ATHEROSCLEROTIC PLAQUES; INVERSION-RECOVERY; ARTERIES; HEMORRHAGE; COMPONENTS; SYMPTOMS; IMAGES;
D O I
10.1002/jmri.21890
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To design and compare an eight channel phased array (PA) coil for carotid imaging to an established four-channel PA design at 3T. Materials and Methods: An eight-channel PA (8PA) coil was designed specifically for imaging the carotid bifurcation and compared with the existing four-channel (4PA) design using a phantom and by in vivo black-blood magnetic resonance imaging (MRI). The 8PA and 4PA were compared in terms of coverage, signal-to-noise ratio (SNR), and contrast-to-noise (CNR). Results: The 8PA showed up to 1.7-fold improvement in SNR at a depth of 3.5 cm and greater longitudinal coverage at a given SNR on a phantom. The 8PA showed improved vessel wall SNR for high spatial resolution (0.63/mm(2)), PD. T1, and T2(1.7, 1.7, 1.6 times, respectively; P <= 0.002) and improved CNR(1.7, 1.6, 1.5 times respectively; P <= 0.002). Ultrahigh resolution (0.27 mm(2)). T1 weighted images showed better SNR and CNR(1.4 times, P <= 0.0001) on 8PA compared to 4PA. Conclusion: Carotid imaging studies may benefit from the improved SNR and larger coverage provided by use of the SPA.
引用
收藏
页码:1209 / 1214
页数:6
相关论文
共 50 条
  • [21] Numerical Design of Dual Resonant Phased Array RF Coil for MRI 3T and 7T System
    Basari
    Yuliyanti, Sri
    Rahardjo, Eko Tjipto
    Zulkifli, Fitri Yuli
    2015 2ND INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY, COMPUTER, AND ELECTRICAL ENGINEERING (ICITACEE), 2015, : 275 - 278
  • [22] Design of 8 channel insertional pTx array coil for 3T body imaging
    Kim, Young Beom
    Ryu, Yeunchul
    Transactions of the Korean Institute of Electrical Engineers, 2014, 63 (04): : 546 - 550
  • [23] Continuous ASL (CASL) perfusion MRI with an array coil and parallel imaging at 3T
    Wang, Z
    Wang, JJ
    Connick, TJ
    Wetmore, GS
    Detre, JA
    MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (03) : 732 - 737
  • [24] Prostate MRI using a rigid two-channel phased-array endorectal coil: comparison with phased array coil acquisition at 3 T
    Sara Lewis
    Aasrith Ganti
    Pamela Argiriadi
    Ally Rosen
    Stefanie Hectors
    Sahar Semaan
    Christopher Song
    Steve Peti
    Maxwell Segall
    Kezia George
    Vaneela Jaikaran
    Sebastian Villa
    David Kestenbaum
    Nicholas Voutsinas
    John Doucette
    Ashutosh Tewari
    Ardeshir R. Rastinehad
    Bachir Taouli
    Cancer Imaging, 22
  • [25] Prostate MRI using a rigid two-channel phased-array endorectal coil: comparison with phased array coil acquisition at 3 T
    Lewis, Sara
    Ganti, Aasrith
    Argiriadi, Pamela
    Rosen, Ally
    Hectors, Stefanie
    Semaan, Sahar
    Song, Christopher
    Peti, Steve
    Segall, Maxwell
    George, Kezia
    Jaikaran, Vaneela
    Villa, Sebastian
    Kestenbaum, David
    Voutsinas, Nicholas
    Doucette, John
    Tewari, Ashutosh
    Rastinehad, Ardeshir R.
    Taouli, Bachir
    CANCER IMAGING, 2022, 22 (01)
  • [26] Three-Element Phased-Array Coil for Imaging of Rat Spinal Cord at 7T
    Mogatadakala, Kishore V.
    Bankson, James A.
    Narayana, Ponnada A.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (06) : 1498 - 1505
  • [27] Eight-channel decoupled array for cervical spinal cord imaging at 3T: six-channel posterior and two-channel anterior array coil
    Sapkota, Nabraj
    Thapa, Bijaya
    Lee, YouJung
    Kim, Taeho
    Bisson, Erica F.
    Shah, Lubdha M.
    Rose, John W.
    Jeong, Eun-Kee
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2016, 46 (02) : 90 - 99
  • [28] Human cardiac imaging at 3 T using phased array coils
    Noeske, R
    Seifert, F
    Rhein, KH
    Rinneberg, H
    MAGNETIC RESONANCE IN MEDICINE, 2000, 44 (06) : 978 - 982
  • [29] 3T Multiparametric MRI (3T mMRI) With a Phased-Array Torso Coil in Staging of Prostate Cancer Patients (PCP) Prior to Radical Prostatectomy: Preliminary Results
    Counago, F.
    Recio, M.
    Del Cerro, E.
    Diaz, A. A.
    Marcos, F. J.
    Murillo, R.
    Rodriguez, J. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S409 - S410
  • [30] Validation of PI-RADS v.2 for prostate cancer diagnosis with MRI at 3T using an external phased-array coil
    Baldisserotto, Matteo
    Dornelles Neto, Eurico J.
    Carvalhal, Gustavo
    de Toledo, Aloyso F.
    de Almeida, Clovis M.
    Cairoli, Carlos E. D.
    de Silva, Daniel O.
    Carvalhal, Eduardo
    Paganin, Ricardo P.
    Agra, Alexandre
    de Santos, Francisco S.
    Noronha, Jorge A. P.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 44 (05) : 1354 - 1359