Evaluation of the IRMA TRUpoint and i-STAT creatinine assays

被引:18
|
作者
Nichols, James H.
Bartholomew, Cathy
Bonzagi, Arianna
Garb, Jane L.
Jin, Long
机构
[1] Baystate Hlth Syst, Clin Chem, Dept Surg, Springfield, MA 01199 USA
[2] Tufts Univ, Sch Med, Baystate Hlth Syst, Dept Pathol, Springfield, MA 01199 USA
关键词
IRMA TRUpoint analyzer; iSTAT analyzer; Jaffe creatinine; enzymatic creatinine; point-of-care testing; estimated GFR;
D O I
10.1016/j.cca.2006.09.024
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The i-STAT((R))(Abbott Diagnostics, East Windsor, NJ) and IRMA TRUpoint (TM) (ITC, Edison, NJ) POCT analyzers were evaluated in an oncology center. Methods: Precision and agreement with our core laboratory creatinine was judged by comparison of 50 consecutive chemotherapy patient results against the Roche rate-blanked Jaffe and enzymatic creatinine methods. Glomerular filtration rate (GFR) was estimated using the Cockroft-Gault (CG) calculation and Modification of Diet in Renal Disease Study (MDRD) equation. Results: Precision varied from 1% (enzymatic)-6.1% (TRUpoint). Correlation was good (r > 0.9948) with slopes within 5% of the Jaffe and enzymatic methods. Intercepts were < 15.9 mu mol/l (< 0.18 mg/dl), and statistically significant bias (p < 0.0025) was noted between the mean of patient specimens for i-STAT correlations to both the Jaffe and enzymatic laboratory creatinine methods. There was statistically significant concordance of estimated GFR between all methods, however, the agreement of estimated GFR to either the Jaffe or enzymatic creatinine laboratory methods was better for the TRUpoint (by either MDRD or CG estimation)and i-STAT (by MDRD equation) (Kappa > 0.60) than the i-STAT (by CG estimation) (Kappa = 0.41-0.60). Conclusion: Small biases in the calibration of analytical creatinine methods can lead to differences in clinical concordance using estimated GFR. Selecting an optimal POCT method depends on the institution's current creatinine method and tolerance for analytical performance and clinical concordance. (c) 2006 Elsevier B.V All rights reserved.
引用
收藏
页码:201 / 205
页数:5
相关论文
共 50 条
  • [31] Evaluation of the next generation I-STAT® point-of-care prothrombin time test (I-STAT PTPLUS) against other commercially available prothrombin time tests
    Frank, P.
    Ball, S.
    Thibodeau, L.
    Perry, R.
    Ditullio, K.
    Wilson, K.
    Trivedi, A.
    Villeneuve, A.
    CLINICA CHIMICA ACTA, 2019, 493 : S645 - S645
  • [32] Effects of extreme temperature, altitude, and humidity on clinical chemistry assays using i-STAT cartridges.
    不详
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2005, 35 (02): : 210 - 210
  • [33] Analytical performance of the i-STAT cardiac troponin I assay
    Apple, FS
    Murakami, MM
    Christenson, RH
    Campbell, JL
    Miller, CJ
    Hock, KG
    Scott, MG
    CLINICA CHIMICA ACTA, 2004, 345 (1-2) : 123 - 127
  • [34] Evaluation of the CoaguChek S, CoaguChek XS and i-STAT for Point of Care INR testing
    Beyer, L. K.
    Wockenfus, A.
    McBane, R.
    Chaudhry, R.
    Karon, B.
    Santrach, P.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A198 - A199
  • [35] Evaluation of the i-STAT Alinity point-of-care analyzer in pediatric patient populations
    Di Meo, Ashley
    Lee, Elsie
    Quach, Kim
    Nemec, Rose
    Zappitelli, Michael
    Licht, Christoph
    Adeli, Khosrow
    CLINICAL BIOCHEMISTRY, 2022, 109 : 104 - 104
  • [36] EVALUATION OF A NOVEL POINT-OF-CARE SYSTEM, THE I-STAT PORTABLE CLINICAL ANALYZER
    ERICKSON, KA
    WILDING, P
    CLINICAL CHEMISTRY, 1993, 39 (02) : 283 - 287
  • [37] Accuracy of the i-STAT™ bedside blood gas analyser
    Sediame, S
    Zerah-Lancner, F
    d'Ortho, MP
    Adnot, S
    Harf, A
    EUROPEAN RESPIRATORY JOURNAL, 1999, 14 (01) : 214 - 217
  • [38] Evaluation of i-STAT portable clinical analyzer in a neonatal and pediatric intensive care unit
    Murthy, JN
    Hicks, JM
    Soldin, SJ
    CLINICAL BIOCHEMISTRY, 1997, 30 (05) : 385 - 389
  • [39] Use of the i-STAT portable clinical analyzer in mice
    Tinkey, Peggy
    Lembo, Thomas
    Craig, Suzanne
    West, Cheri
    Van Pelt, Carolyn
    LAB ANIMAL, 2006, 35 (02) : 45 - 50
  • [40] Use of the i-STAT portable clinical analyzer in mice
    Peggy Tinkey
    Thomas Lembo
    Suzanne Craig
    Cheri West
    Carolyn Van Pelt
    Lab Animal, 2006, 35 : 45 - 50