Evaluation of the IRMA TRUpoint and i-STAT creatinine assays

被引:18
|
作者
Nichols, James H.
Bartholomew, Cathy
Bonzagi, Arianna
Garb, Jane L.
Jin, Long
机构
[1] Baystate Hlth Syst, Clin Chem, Dept Surg, Springfield, MA 01199 USA
[2] Tufts Univ, Sch Med, Baystate Hlth Syst, Dept Pathol, Springfield, MA 01199 USA
关键词
IRMA TRUpoint analyzer; iSTAT analyzer; Jaffe creatinine; enzymatic creatinine; point-of-care testing; estimated GFR;
D O I
10.1016/j.cca.2006.09.024
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The i-STAT((R))(Abbott Diagnostics, East Windsor, NJ) and IRMA TRUpoint (TM) (ITC, Edison, NJ) POCT analyzers were evaluated in an oncology center. Methods: Precision and agreement with our core laboratory creatinine was judged by comparison of 50 consecutive chemotherapy patient results against the Roche rate-blanked Jaffe and enzymatic creatinine methods. Glomerular filtration rate (GFR) was estimated using the Cockroft-Gault (CG) calculation and Modification of Diet in Renal Disease Study (MDRD) equation. Results: Precision varied from 1% (enzymatic)-6.1% (TRUpoint). Correlation was good (r > 0.9948) with slopes within 5% of the Jaffe and enzymatic methods. Intercepts were < 15.9 mu mol/l (< 0.18 mg/dl), and statistically significant bias (p < 0.0025) was noted between the mean of patient specimens for i-STAT correlations to both the Jaffe and enzymatic laboratory creatinine methods. There was statistically significant concordance of estimated GFR between all methods, however, the agreement of estimated GFR to either the Jaffe or enzymatic creatinine laboratory methods was better for the TRUpoint (by either MDRD or CG estimation)and i-STAT (by MDRD equation) (Kappa > 0.60) than the i-STAT (by CG estimation) (Kappa = 0.41-0.60). Conclusion: Small biases in the calibration of analytical creatinine methods can lead to differences in clinical concordance using estimated GFR. Selecting an optimal POCT method depends on the institution's current creatinine method and tolerance for analytical performance and clinical concordance. (c) 2006 Elsevier B.V All rights reserved.
引用
收藏
页码:201 / 205
页数:5
相关论文
共 50 条
  • [21] Which point-of-care creatinine analyser for radiology: direct comparison of the i-Stat and StatStrip creatinine methods with different sample types
    Dimeski, Goce
    Tilley, Vera
    Jones, Brock W.
    Brown, Nigel N.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2013, 50 (01) : 47 - 52
  • [22] AN EVALUATION OF BLOOD-GAS AND IONIZED CALCIUM MEASUREMENT OF THE I-STAT ANALYZER
    MOCK, T
    YATSCOFF, RW
    CLINICAL CHEMISTRY, 1995, 41 (06) : S182 - S182
  • [23] The reliability of the i-STAT clinical portable analyser
    Dascombe, B. J.
    Reaburn, P. R. J.
    Sirotic, A. C.
    Coutts, A. J.
    JOURNAL OF SCIENCE AND MEDICINE IN SPORT, 2007, 10 (03) : 135 - 140
  • [24] Analytical performance evaluation of the i-STAT Total β-human chorionic gonadotropin immunoassay
    Sowder, Aleksandra M.
    Yarbrough, Melanie L.
    Nerenz, Robert D.
    Mitsios, John V.
    Mortensen, Rachel
    Gronowski, Ann M.
    Grenache, David G.
    CLINICA CHIMICA ACTA, 2015, 446 : 165 - 170
  • [25] QUALITY-CONTROL FOR THE I-STAT ANALYZER
    UTZ, C
    GIBSON, C
    HORTIN, GL
    CLINICAL CHEMISTRY, 1995, 41 (06) : S157 - S157
  • [26] I-STAT - INTERACTIVE STATISTICAL PROGRAM PACKAGE
    ROSENBLO.L
    CHAPMAN, K
    BEHAVIOR RESEARCH METHODS & INSTRUMENTATION, 1974, 6 (01): : 56 - 56
  • [27] Evaluation of the i-STAT analyzer for determination of ionized calcium concentrations in bovine blood
    Yilmaz, Oznur
    Karapinar, Tolga
    VETERINARY CLINICAL PATHOLOGY, 2019, 48 (01) : 31 - 35
  • [28] CARTRIDGE TEMPERATURE SIGNIFICANTLY EFFECTS PAO2 RESULTS USING THE I-STAT PORTABLE CLINICAL ANALYSER (I-STAT)
    Harnett, Lauren A.
    Johnston, Sonya D.
    Thornton, Andrew T.
    Yeo, Aeneas
    RESPIROLOGY, 2012, 17 : 6 - 6
  • [29] Comparison of GFAP and UCH-L1 Measurements from Two Prototype Assays: The Abbott i-STAT and ARCHITECT Assays
    Korley, Frederick K. A.
    Datwyler, Saul A.
    Jain, Sonia
    Sun, Xiaoying
    Beligere, Gangamani
    Chandran, Raj A.
    Marino, Jaime A.
    McQuiston, Beth
    Zhang, Hongwei L.
    Caudle, Krista L.
    Wang, Kevin K. W. M.
    Puccio, Ava M. O.
    Okonkwo, David O. K.
    Yue, John K. R.
    Taylor, Sabrina R.
    Markowitz, Amy T.
    Manley, Geoffrey T.
    Diaz-Arrastia, Ramon
    TRACK-TBI Study Investigators
    NEUROTRAUMA REPORTS, 2021, 2 (01): : 193 - 199
  • [30] i-STAT biosensor system to be distributed in Europe
    不详
    BIOSENSORS & BIOELECTRONICS, 1996, 11 (09): : R2 - R2