Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy

被引:7
|
作者
Murazaki, Hiroo [1 ]
Fukunaga, Junichi [1 ]
Hirose, Taka-aki [1 ]
Funatsu, Naomi [1 ]
Matsumoto, Ryoji [1 ]
Hidaka, Kyohei [1 ]
Nagamine, Shuji [1 ]
Nakanishi, Daiki [1 ]
Kato, Toyoyuki [1 ]
机构
[1] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, 3-1-1 Maidashi, Fukuoka, Fukuoka 8128582, Japan
关键词
Metal artifact reduction; Single-energy metal artifact reduction; Radiation treatment planning; Dose distribution; ANISOTROPIC ANALYTICAL ALGORITHM; ACUROS XB; DOSE CALCULATION; CT; QUALITY; IMPACT; SIMULATION; AAA;
D O I
10.1007/s12194-019-00517-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aimed to evaluate the performance of a single-energy metal artifact reduction (SEMAR) algorithm for radiation therapy treatment using phantom cases with metal inserts, assess improvements in computed tomography (CT) number accuracy, and investigate its effects on treatment planning dosimetry. A standard electron density phantom was scanned with and without metal inserts. The numbers of tissue-equivalent materials on both uncorrected and SEMAR-corrected CT images were compared. Treatment planning accuracy was evaluated by comparing dose distributions computed using true density images (without metal inserts), uncorrected images (with metal inserts), and SEMAR-corrected images (with metal inserts) using three-dimensional gamma analysis. The numbers of the true density and uncorrected and SEMAR-corrected CT images in a muscle plug with unilateral inserts were 25.9 HU, - 281.8 HU, and 26.1 HU, respectively. A similar tendency was obtained for other tissue-equivalent materials, and the numbers on CT images were improved with the SEMAR algorithm. In cases involving 1 portal irradiation, 10-MV X-ray, and the Acuros XB algorithm, the pass ratio between the true density and uncorrected images was 89.89%, while that between the true density and SEMAR-corrected images was 95.03%. Improvements in dose distribution were evident using the SEMAR algorithm. Similar trends were found for different irradiation methods and dose calculation algorithms. The SEMAR algorithm can significantly reduce metal artifacts on CT images used for radiation treatment planning. This aspect influenced dosimetry in the region of the artifact and dose distribution was significantly improved with use of the SEMAR-corrected images.
引用
收藏
页码:268 / 276
页数:9
相关论文
共 50 条
  • [1] Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy
    Hiroo Murazaki
    Junichi Fukunaga
    Taka-aki Hirose
    Naomi Funatsu
    Ryoji Matsumoto
    Kyohei Hidaka
    Shuji Nagamine
    Daiki Nakanishi
    Toyoyuki Kato
    Radiological Physics and Technology, 2019, 12 : 268 - 276
  • [2] Single-energy metal artifact reduction for helical computed tomography of the pelvis in patients with metal hip prostheses
    Koichiro Yasaka
    Eriko Maeda
    Shouhei Hanaoka
    Masaki Katsura
    Jiro Sato
    Kuni Ohtomo
    Japanese Journal of Radiology, 2016, 34 : 625 - 632
  • [3] Single-energy metal artifact reduction for helical computed tomography of the pelvis in patients with metal hip prostheses
    Yasaka, Koichiro
    Maeda, Eriko
    Hanaoka, Shouhei
    Katsura, Masaki
    Sato, Jiro
    Ohtomo, Kuni
    JAPANESE JOURNAL OF RADIOLOGY, 2016, 34 (09) : 625 - 632
  • [4] Performance of single-energy metal artifact reduction in cardiac computed tomography: A clinical and phantom study
    Groves, Daniel W.
    Acharya, Tushar
    Steveson, Chloe
    Schuzer, John L.
    Rollison, Shirley F.
    Nelson, Evan A.
    Sirajuddin, Arlene
    Sathya, Bharath
    Bronson, Kathie
    Shanbhag, Sujata M.
    Chen, Marcus Y.
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2020, 14 (06) : 510 - 515
  • [5] Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction
    Giantsoudi, Drosoula
    De Man, Bruno
    Verburg, Joost
    Trofimov, Alexei
    Jin, Yannan
    Wang, Ge
    Gjesteby, Lars
    Paganetti, Harald
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (08): : R49 - R80
  • [6] Deep learning reconstruction with single-energy metal artifact reduction in pelvic computed tomography for patients with metal hip prostheses
    Hosoi, Reina
    Yasaka, Koichiro
    Mizuki, Masumi
    Yamaguchi, Haruomi
    Miyo, Rintaro
    Hamada, Akiyoshi
    Abe, Osamu
    JAPANESE JOURNAL OF RADIOLOGY, 2023, 41 (08) : 863 - 871
  • [7] Deep learning reconstruction with single-energy metal artifact reduction in pelvic computed tomography for patients with metal hip prostheses
    Reina Hosoi
    Koichiro Yasaka
    Masumi Mizuki
    Haruomi Yamaguchi
    Rintaro Miyo
    Akiyoshi Hamada
    Osamu Abe
    Japanese Journal of Radiology, 2023, 41 : 863 - 871
  • [8] Radiotherapy planning of spine and pelvis using single-energy metal artifact reduction corrected computed tomography sets
    Ignatius, Daliya
    Alkhatib, Zaid
    Rowshanfarzad, Pejman
    Goodall, Simon
    Ibrahim, Mounir
    Hirst, Andrew
    Croxford, Riley
    Dass, Joshua
    Sabet, Mahsheed
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2023, 26
  • [9] Added value of a single-energy projection-based metal-artifact reduction algorithm for the computed tomography evaluation of oral cavity cancers
    Hirata, Kenichiro
    Utsunomiya, Daisuke
    Oda, Seitaro
    Kidoh, Masafumi
    Funama, Yoshinori
    Yuki, Hideaki
    Yoshida, Morikatsu
    Yamashita, Yasuyuki
    JAPANESE JOURNAL OF RADIOLOGY, 2015, 33 (10) : 650 - 656
  • [10] Added value of a single-energy projection-based metal-artifact reduction algorithm for the computed tomography evaluation of oral cavity cancers
    Kenichiro Hirata
    Daisuke Utsunomiya
    Seitaro Oda
    Masafumi Kidoh
    Yoshinori Funama
    Hideaki Yuki
    Morikatsu Yoshida
    Yasuyuki Yamashita
    Japanese Journal of Radiology, 2015, 33 : 650 - 656