Numerical investigation on stress intensity around Bone-Implant interface by 3-Dimensional FEA and experimental verification by optical technique

被引:12
|
作者
Dhatrak, Pankaj [1 ]
Shirsat, Uddhav [2 ]
Sumanth, S. [3 ]
Deshmukh, Vijay [4 ]
机构
[1] Dr Vishwanath Karad MIT WPU, MAEERS MIT, Sch Mech Engn, Pune 38, Maharashtra, India
[2] Bhivarabai Swanat Coll Engn & Researach, Dept Mech Engn, Pune 41, Maharashtra, India
[3] MA Rangoonwal Coll Dent Sci & Res Ctr, Dept Periodont, Pune 01, Maharashtra, India
[4] Int Clin Dent Res Org, Pune 38, Maharashtra, India
关键词
Cancellous bone; Dental Implant; FEA; Optical Technique; Photoelastic stress;
D O I
10.1016/j.matpr.2020.06.097
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The aim of the current study is to examine numerically the stress intensity along bone-implant interface for selected implant systems under occlusal load using finite element analysis (FEA) and confirming the result experimentally by optical technique. To see the biomechanical performance at bone-implant interface the five types of implant systems are chosen for the current study. A three-dimensional (3D) computer aided design (CAD) is prepared using CreO Parametric modeling software. A non-linear contact between implant and surrounding bone is defined to analyze the stress intensity in the cancellous bone under the applied load of 100 N axial directions, 40 N laterals (Bucco-lingual) direction and 100 N oblique at 45 degree to the longitudinal axis of implant. An experimental photoelastic technique (Optical technique) is used to predict the stress intensity in a cancellous bone near the bone-implant interface using Tardy's method of compensation. A homogeneous blend of Araldite and hardener material is used to prepare a photoelastic block model. The maximum value of stress intensity is noted in Type-I implant system at apical part of implant whereas minimum value was noted in Type-III implant systemunder lateral and oblique loads. Stress intensity in cancellous bone at the bone interface is found to be more significant to the applied oblique load. (C) 2020 The Authors. Published by Elsevier Ltd.
引用
收藏
页码:35 / 41
页数:7
相关论文
共 50 条
  • [41] Influence of Implant Positions and Occlusal Forces on Peri-Implant Bone Stress in Mandibular Two-Implant Overdentures: A 3-Dimensional Finite Element Analysis
    Alvarez-Arenal, Angel
    Gonzalez-Gonzalez, Ignacio
    deLlanos-Lanchares, Hector
    Brizuela-Velasco, Aritza
    Martin-Fernandez, Elena
    Ellacuria-Echebarria, Joseba
    JOURNAL OF ORAL IMPLANTOLOGY, 2017, 43 (06) : 419 - 428
  • [42] MEASUREMENT OF 3-DIMENSIONAL MODE-I STRESS INTENSITY FACTOR USING MULTIPLE EMBEDDED GRID MOIRE TECHNIQUE
    DHAR, S
    CLOUD, GL
    PALEEBUT, S
    THEORETICAL AND APPLIED FRACTURE MECHANICS, 1989, 12 (02) : 141 - 147
  • [43] EXPERIMENTAL AND NUMERICAL INVESTIGATION OF 3-DIMENSIONAL VISCOUS FLOWS AND VORTEX MOTION INSIDE AN ANNULAR COMPRESSOR BLADE ROW
    GALLUS, HE
    HAH, C
    SCHULZ, HD
    JOURNAL OF TURBOMACHINERY-TRANSACTIONS OF THE ASME, 1991, 113 (02): : 198 - 206
  • [44] A Three-Dimensional Finite Element Analysis of the Stress Distribution Around the Bone Mini-Implant Interface Based on the Mini-Implant Angle of Insertion, Diameter, and Length
    Sarika, K.
    Kumaran, Navaneethakrishnan Kurunji
    Seralathan, Sakthidaran
    Sathishkumar, Rathinavel Kumar
    Preethi, Somasundaram Kanmani
    JOURNAL OF PHARMACY AND BIOALLIED SCIENCES, 2023, 15 : S535 - S539
  • [45] DETERMINATION OF 3-DIMENSIONAL STRESS INTENSITY FACTOR USING VIRTUAL CRACK EXTENSION TECHNIQUE BASED ON SINGULAR INTEGRAL-EQUATIONS
    SOHN, GH
    HONG, CS
    ENGINEERING FRACTURE MECHANICS, 1992, 41 (02) : 177 - 190
  • [46] Effects of custom-made and thread dental implant systems on the stress distribution in alveolar bone: a 3-dimensional finite element analysis
    Zhang, Xiao
    Xie, Zhangong
    Feng, Wei
    Chen, Xianshuai
    Chen, Jianyu
    SENSORS, MEASUREMENT AND INTELLIGENT MATERIALS II, PTS 1 AND 2, 2014, 475-476 : 1487 - +
  • [47] Stress and Strain Patterns of 1-Piece and 2-Piece Implant Systems in Bone: A 3-Dimensional Finite Element Analysis
    Ormianer, Zeev
    Ben Amar, Arial
    Duda, Mariusz
    Marku-Cohen, Sharon
    Lewinstein, Israel
    IMPLANT DENTISTRY, 2012, 21 (01) : 39 - 45
  • [48] DIRECT NUMERICAL-SIMULATION OF 3-DIMENSIONAL NAVIER-STOKES EQUATIONS FOR A SLIT NOZZLE FREE JET AND EXPERIMENTAL-VERIFICATION
    YUU, S
    NISHIOKA, T
    UMEKAGE, T
    JSME INTERNATIONAL JOURNAL SERIES B-FLUIDS AND THERMAL ENGINEERING, 1993, 36 (01) : 113 - 120
  • [49] Effects of Impression Coping Design, Impression Technique, and Dental Undercuts on the Accuracy of Implant Impressions Assessed by 3-Dimensional Optical Scanning: An In Vitro Study
    Sabouhi, Mahmoud
    Bajoghli, Farshad
    Dakhilalian, Mansour
    Beygi, Ali
    Abolhasani, Majid
    IMPLANT DENTISTRY, 2016, 25 (02) : 238 - 246
  • [50] Stress Distribution Around Maxillary Anterior Implants as a Factor of Labial Bone Thickness and Occlusal Load Angles: A 3-Dimensional Finite Element Analysis
    Alikhasi, Marzieh
    Siadat, Hakimeh
    Geramy, Allahyar
    Hassan-Ahangari, Ahmad
    JOURNAL OF ORAL IMPLANTOLOGY, 2014, 40 (01) : 37 - 41