Effective dose to staff from interventional procedures: estimations from single and double dosimetry

被引:14
|
作者
Kuipers, Gerritjan [1 ]
Velders, Xandra L. [1 ]
机构
[1] Acad Med Ctr, NL-1100 DD Amsterdam, Netherlands
关键词
CARDIAC-CATHETERIZATION; FOLLOW-UP; RADIOLOGY;
D O I
10.1093/rpd/ncp155
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The exposure of 11 physicians performing interventional procedures was measured by means of two personal dosemeters. One personal dosemeter was worn outside the lead apron and an additional under the lead apron. The study was set up in order to determine the added value of a dosemeter worn under the lead apron. With the doses measured, the effective doses of the physicians were estimated using an algorithm for single dosimetry and two algorithms for double dosimetry. The effective doses calculated with the single dosimetry algorithm ranged from 0.11 to 0.85 mSv in 4 weeks. With the double dosimetry algorithms, the effective doses ranged from 0.02 mSv to 0.47 mSv. The statistical analysis revealed no significant differences in the accuracy of the effective doses calculated with single or double dosimetry algorithms. It was concluded that the effective dose cannot be considered a more accurate estimate when two dosemeters are used instead of one.
引用
收藏
页码:95 / 100
页数:6
相关论文
共 50 条
  • [1] Overview of double dosimetry procedures for the determination of the effective dose to the interventional radiology staff
    Jarvinen, H.
    Buls, N.
    Clerinx, P.
    Jansen, J.
    Miljanic, S.
    Nikodemova, D.
    Ranogajec-Komor, M.
    d'Errico, F.
    RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) : 333 - 339
  • [2] COMPARISON OF DOUBLE DOSIMETRY ALGORITHMS FOR ESTIMATING THE EFFECTIVE DOSE IN OCCUPATIONAL DOSIMETRY OF INTERVENTIONAL RADIOLOGY STAFF
    Jarvinen, H.
    Buls, N.
    Clerinx, P.
    Miljanic, S.
    Nikodemova, D.
    Ranogajec-Komor, M.
    Struelens, L.
    d'Errico, F.
    RADIATION PROTECTION DOSIMETRY, 2008, 131 (01) : 80 - 86
  • [3] Use of secondary air kerma from dose area product and fluoroscopy time for preventive effective dose estimation in interventional radiology procedures for staff
    Indovina, Luca
    Moretti, Roberto
    Breschi, Laura
    Cusumano, Davide
    Meffe, Guenda
    Placidi, Lorenzo
    Nardini, Matteo
    Reale, Roberta
    Quici, Arturo
    Placidi, Elisa
    Stimato, Gerardina
    Capotosti, Amedeo
    JOURNAL OF RADIOLOGICAL PROTECTION, 2023, 43 (02)
  • [4] Simulation of Hp(10) and effective dose received by the medical staff in interventional radiology procedures
    Ferrari, P.
    Becker, F.
    Jovanovic, Z.
    Khan, S.
    Bakhanova, E.
    Principi, S.
    Kristic, D.
    Pierotti, L.
    Mariotti, F.
    Faj, D.
    Turk, T.
    Nikezic, D.
    Bertolini, M.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2019, 39 (03) : 809 - 824
  • [5] DOSIMETRY FOR THE LENS OF THE EYE, APPLICATIONS FOR MEDICAL STAFF INVOLVED IN INTERVENTIONAL PROCEDURES
    Geber, Therese
    Gunnarsson, Mikael
    Mattsson, Soren
    MEDICAL PHYSICS IN THE BALTIC STATES, 2010, (02): : 105 - 108
  • [6] Dose Evaluation in Medical Staff during Interventional Radiology Procedures
    Bacchim Neto, F. A.
    Alves, A. F. F.
    Alvarez, M.
    Rosa, M. E. D.
    Macedo de Freitas, C. C.
    Moura, R.
    Pina, D. R.
    Miranda, J. R. A.
    VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 : 927 - 930
  • [7] Effective-dose estimation in interventional radiological procedures
    Falco M.D.
    Masala S.
    Stefanini M.
    Bagalà P.
    Morosetti D.
    Calabria E.
    Tonnetti A.
    Verona-Rinati G.
    Radiological Physics and Technology, 2018, 11 (2) : 149 - 155
  • [8] Correlation Between Routine Personal Dosimetry Reading and the Dose to the Brain of Interventional Staff
    Smeulders, Jelle
    da Silva, Edilaine Honorio
    Struelens, Lara
    Vanhavere, Filip
    De Mey, Johan
    Martin, Colin J.
    Buls, Nico
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (06) : 349 - 357
  • [9] Medical staff dose estimation during pediatric cardiac interventional procedures
    Roberto, Maryanna Regina
    Leite, Luiz M. A.
    Cruz, Aricia R. P.
    Khoury, Helen
    de Barros, Vinicius S. M.
    Neves, Juliana R.
    Cartaxo, Jessica M.
    Ubeda, Carlos
    Asfora, Viviane K.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2024, 44 (02)
  • [10] Estimation of effective dose in some digital angiographic and interventional procedures
    Cruces, RR
    García-Granados, J
    Romero, FJD
    Armas, JH
    BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841): : 42 - 47