Dosimetry studies with TLDs for stereotactic radiation techniques for intraocular tumours

被引:16
|
作者
Ertl, A
Zehetmayer, M
Schoggl, A
Kindl, P
Hartl, R
机构
[1] UNIV VIENNA, DEPT OPHTHALMOL, A-1090 VIENNA, AUSTRIA
[2] GRAZ TECH UNIV, INST TECH PHYS, A-8010 GRAZ, AUSTRIA
[3] TECH UNIV VIENNA, AUSTRIAN ATOM INST, A-1060 VIENNA, AUSTRIA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1997年 / 42卷 / 11期
关键词
D O I
10.1088/0031-9155/42/11/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Between March 1993 and January 1997, stereotactic radiation techniques were used to irradiate 66 intraocular tumour patients with the Gamma Knife (Leksell Gamma Knife(R), model B unit) at the University of Vienna, Austria. This study investigates the dosimetry for stereotactic irradiation of ocular structures. For the dosimetry program KULA 4.4, Gamma Knife stereotactic irradiation of the eye represents an extreme frontal skull position. In addition, irradiation of the eye may be performed in the usual supine position in exceptional cases only. With the patient in the prone position, the dose planning program has to calculate with a significantly large number of single-beam extrapolations. In our first experiment we measured the isocentre dose for eight different gamma-angle positions, both in prone and supine positions, using TLD measurements in an Alderson head phantom. We found a maximum deviation of +/-1.6% using these individually calibrated TLDs. In the second experiment we examined the dose cross profiles for the two most frequently used treatment positions (supine position, gamma = 65 degrees, and prone position, gamma = 140 degrees). For this purpose we implanted a specially designed TLD array into the orbit of a human cadaver head. We found excellent agreement of the dose values measured for the isocentre as well as the posterior part of the eye with orbit with deviations of less than -2.7%. However, for the anterior part of the eye, deviations between computer-generated calculations and the TLD measurements were found to range up to -30%. These differences were noticed both for supine and prone positions. For the Gamma Knife stereotactic irradiation of ocular tumours or pathologies, precautions should be taken to avoid significant underdosage in the anterior part of the radiation field.
引用
收藏
页码:2137 / 2145
页数:9
相关论文
共 50 条
  • [31] Technical and Quality Considerations for Stereotactic Radiation Treatment Techniques
    Covington, Elizabeth L.
    Popple, Richard A.
    CANCER JOURNAL, 2024, 30 (06): : 372 - 376
  • [32] EVALUATION OF STEREOTACTIC RADIATION THERAPY TECHNIQUES FOR INTRACRANIAL TUMORS
    Seoane, A.
    Puertas, E.
    Hermida, M.
    Toribio, I.
    Benavente, S.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S259 - S260
  • [33] RADIATION DOSIMETRY FOR BIOLOGICAL STUDIES WITH CYCLOTRON NEUTRONS
    SHUMWAY, B
    TOCHILIN, E
    KOHLER, G
    RADIATION RESEARCH, 1955, 3 (03) : 346 - 347
  • [34] Dosimetry protocol for the forthcoming clinical trials in synchrotron stereotactic radiation therapy (SSRT)
    Prezado, Y.
    Vautrin, M.
    Martinez-Rovira, I.
    Bravin, A.
    Esteve, F.
    Elleaume, H.
    Berkvens, P.
    Adam, J. F.
    MEDICAL PHYSICS, 2011, 38 (03) : 1709 - 1717
  • [35] Evaluation of therapeutic response after stereotactic body radiation therapy for liver tumours
    Janoray, G.
    Barillot, I.
    Calais, G.
    CANCER RADIOTHERAPIE, 2014, 18 (04): : 320 - 324
  • [36] PERSONNEL DOSIMETRY TECHNIQUES FOR EXTERNAL RADIATION THEIR APPLICATION IN NUCLEAR INSTALLATIONS
    OBERHOFE.M
    KERNTECHNIK ISOTOPENTECHNIK UND CHEMIE, 1965, 7 (06): : 297 - +
  • [37] Dosimetry and techniques for simultaneous hyperthermia and external beam radiation therapy
    Straube, WL
    Klein, EE
    Moros, EG
    Low, DA
    Myerson, RJ
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2001, 17 (01) : 48 - 62
  • [38] REVIEW OF RETROSPECTIVE DOSIMETRY TECHNIQUES FOR EXTERNAL IONISING RADIATION EXPOSURES
    Ainsbury, E. A.
    Bakhanova, E.
    Barquinero, J. F.
    Brai, M.
    Chumak, V.
    Correcher, V.
    Darroudi, F.
    Fattibene, P.
    Gruel, G.
    Guclu, I.
    Horn, S.
    Jaworska, A.
    Kulka, U.
    Lindholm, C.
    Lloyd, D.
    Longo, A.
    Marrale, M.
    Monteiro Gil, O.
    Oestreicher, U.
    Pajic, J.
    Rakic, B.
    Romm, H.
    Trompier, F.
    Veronese, I.
    Voisin, P.
    Vral, A.
    Whitehouse, C. A.
    Wieser, A.
    Woda, C.
    Wojcik, A.
    Rothkamm, K.
    RADIATION PROTECTION DOSIMETRY, 2011, 147 (04) : 573 - 592
  • [39] Dosimetry comparison of irradiation with conformal radiotherapy, intensity modulated radiotherapy, conformal radiotherapy in stereotactic conditions and robotic stereotactic radiotherapy for benign brain tumours
    Spasic, E.
    Buchheit, I.
    Bernier, V.
    Noel, A.
    CANCER RADIOTHERAPIE, 2011, 15 (04): : 287 - 293
  • [40] PHYSICAL DOSIMETRY TECHNIQUES UTILIZED IN RADIOBIOLOGICAL STUDIES WITH PROTONS
    HARDY, KH