Development of calibration techniques for ultrasonic hydrophone probes in the frequency range from 1 to 100 MHz

被引:42
|
作者
Umchid, S. [1 ]
Gopinath, R. [2 ]
Srinivasan, K. [2 ]
Lewin, P. A. [2 ,3 ]
Daryoush, A. S. [2 ]
Bansal, L. [4 ]
El-Sherif, M. [4 ]
机构
[1] King Mongkuts Univ Technol N Bangkok, Fac Sci Appl, Dept Ind Phys & Med Instrumentat, Bangkok 10800, Thailand
[2] Drexel Univ, Dept Elect & Comp Engn, Philadelphia, PA 19104 USA
[3] Drexel Univ, Sch Biomed Engn Sci & Hlth Syst, Philadelphia, PA 19104 USA
[4] Photon Labs Inc, Philadelphia, PA 19125 USA
关键词
Nonlinear propagation; Spatial averaging; Piezoelectric polymer ultrasound hydrophone probes; Fiber optic hydrophone probe; Hydrophone calibration; NONLINEAR PROPAGATION MODEL; FIELDS;
D O I
10.1016/j.ultras.2008.09.011
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The primary objective of this work was to develop and optimize the calibration techniques for ultrasonic hydrophone probes used in acoustic field measurements up to 100 MHz. A dependable, 100 MHz calibration method was necessary to examine the behavior of a sub-millimeter spatial resolution fiber optic (FO) sensor and assess the need for such a sensor as an alternative tool for high frequency characterization of ultrasound fields. Also, it was of interest to investigate the feasibility of using FO probes in high intensity fields such as those employed in HIFU (high intensity focused ultrasound) applications. In addition to the development and validation of a novel, 100 MHz calibration technique the innovative elements of this research include implementation and testing of a prototype FO sensor with an active diameter of about 10 mu m that exhibits uniform sensitivity over the considered frequency range and does not require any spatial averaging corrections up to about 75 MHz. The results of the calibration measurements are presented and it is shown that the optimized calibration technique allows the sensitivity of the hydrophone probes to be determined as a virtually continuous function of frequency and is also well suited to verify the uniformity of the FO sensor frequency response. As anticipated, the overall uncertainty of the calibration was dependent on frequency and determined to be about +/- 12% (+/- 1 dB) up to 40 MHz, +/- 20% (+/- 1.5 dB) from 40 to 60 MHz and +/- 25% (+/- 2 dB) from 60 to 100 MHz. The outcome of this research indicates that once fully developed and calibrated, the combined acousto-optic system will constitute a universal reference tool in the wide, 100 MHz bandwidth. (C) 2008 Elsevier B. V. All rights reserved.
引用
收藏
页码:306 / 311
页数:6
相关论文
共 50 条
  • [31] Development of a high sensitivity hydrophone at a spectrum range from 0.1 Hz to 100 kHz
    Motsi, Nefeli
    Stamou, Georgia
    Angelopoulos, Spyridon
    Ktena, Aphrodite
    Hristoforou, Evangelos
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2024, 372
  • [32] CONSTRUCTION AND CALIBRATION TECHNIQUES OF HIGH FREQUENCY MAGNETIC PROBES
    PHILLIPS, RC
    TURNER, EB
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1965, 36 (12): : 1822 - &
  • [33] ULTRASOUND CHARACTERIZATION OF NORMAL OCULAR TISSUE IN THE FREQUENCY-RANGE FROM 50 MHZ TO 100 MHZ
    YE, SG
    HARASIEWICZ, KA
    PAVLIN, CJ
    FOSTER, FS
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1995, 42 (01) : 8 - 14
  • [34] Interlaboratory Comparisons of Hydrophone Calibration in the 3-500 kHz Frequency Range
    Malarkodi, A.
    Latha, G.
    Atmanand, M. A.
    Isaev, A. E.
    Matveev, A. N.
    Shcherbliuk, N. G.
    Mann, F.
    Jansen, C.
    Milkert, D.
    [J]. MEASUREMENT TECHNIQUES, 2016, 59 (01) : 99 - 103
  • [35] TECHNIQUES FOR MEASUREMENT OF INPUT IMPEDANCES OF ELECTRONIC EQUIPMENT IN THE FREQUENCY-RANGE FROM 1 MHZ TO 1 GHZ
    KWASNIOK, PJ
    BUI, MD
    KOZLOWSKI, AJ
    STUCHLY, SS
    [J]. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 1992, 34 (04) : 486 - 490
  • [36] Design and calibration of electric field probes in the range 10-120 MHz
    Taylor, HC
    Burl, M
    Hand, JW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (07): : 1387 - 1394
  • [37] Attenuation measurement system in the frequency range of 10 to 100 MHz
    Widarta, A
    Tomoteru, K
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2003, 52 (02) : 302 - 305
  • [38] Experimental Study of Electrical Properties of Oil-Based Mud in the Frequency Range From 1 to 100 MHz
    Patil, Parimal A.
    Gorek, Matthias
    Folberth, Martin
    Hartmann, Andreas
    Forgang, Stan
    Fulda, Christian
    Reinicke, Kurt M.
    [J]. SPE DRILLING & COMPLETION, 2010, 25 (03) : 380 - 390
  • [39] A comparison of two- and four-electrode techniques to characterize blood impedance for the frequency range of 100 Hz to 100 MHz
    Chang, Zu-yao
    Pop, Gheorghe A. M.
    Meijer, Gerard C. M.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (03) : 1247 - 1249
  • [40] Primary calibration of hydrophones with extended frequency range 1 to 70 MHz using optical interferometry
    Koch, C
    Molkenstruck, W
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1999, 46 (05) : 1303 - 1314