Enhancement and validation of Geant4 Brachytherapy application on clinical HDR 192Ir source

被引:9
|
作者
Ababneh, Eshraq [1 ]
Dababneh, Saed [1 ]
Qatarneh, Sharif [2 ]
Wadi-Ramahi, Shada [3 ]
机构
[1] Al Balqa Appl Univ, Dept Phys, Fac Sci, Amman 11941, Jordan
[2] North West Armed Forces Hosp, Prince Sultan Oncol Ctr, Tabuk, Saudi Arabia
[3] King Hussein Canc Ctr, Amman, Jordan
关键词
Brachytherapy; HDR; Monte Carlo simulation; Geant4; Scatter dose; MONTE-CARLO; DOSIMETRY PARAMETERS; DOSE CALCULATIONS; SIMULATIONS; ATTENUATION; ESTRO; SEEDS; WATER; AAPM;
D O I
10.1016/j.radphyschem.2014.05.040
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The Geant4 Monte Carlo MC associated Brachytherapy example was adapted, enhanced and several analysis techniques have been developed. The simulation studies the isodose distribution of the total, primary and scattered doses around a Nucletron microSelectron Ir-192 source. Different phantom materials were used (water, tissue and bone) and the calculation was conducted at various depths and planes. The work provides an early estimate of the required number of primary events to ultimately achieve a given uncertainty at a given distance, in the otherwise CPU and time consuming clinical MC calculation. The adaptation of the Geant4 toolkit and the enhancements introduced to the code are all validated including the comprehensive decay of the Ir-192 source, the materials used to build the geometry, the geometry itself and the calculated scatter to primary dose ratio. The simulation quantitatively illustrates that the scattered dose in the bone medium is larger than its value in water and tissue. As the distance away from the source increases, scatter contribution to dose becomes more significant as the primary dose decreases. The developed code could be viewed as a platform that contains detailed dose calculation model for clinical application of HDR Ir-192 in Brachytherapy. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:57 / 66
页数:10
相关论文
共 50 条
  • [31] An experimental MOSFET approach to characterize 192Ir HDR source anisotropy
    Toye, W. C.
    Das, K. R.
    Todd, S. P.
    Kenny, M. B.
    Franich, R. D.
    Johnston, P. N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (17): : 5329 - 5339
  • [32] Positional and angular tracking of HDR 192Ir source for brachytherapy quality assurance using radiochromic film dosimetry
    Aldelaijan, Saad
    Devic, Slobodan
    Bekerat, Hamed
    Papaconstadopoulos, Pavlos
    Schneider, James
    Seuntjens, Jan
    Cormack, Robert A.
    Buzurovic, Ivan M.
    [J]. MEDICAL PHYSICS, 2020, 47 (12) : 6122 - 6139
  • [33] Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations
    Woo Sang Ahn
    Wonsik Choi
    Seong Soo Shin
    Jinhong Jung
    [J]. Journal of the Korean Physical Society, 2014, 65 : 557 - 564
  • [34] Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations
    Ahn, Woo Sang
    Choi, Wonsik
    Shin, Seong Soo
    Jung, Jinhong
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (04) : 557 - 564
  • [35] Dose perturbation effects by metal hip prosthesis in gynaecological 192Ir HDR brachytherapy
    Ab Shukor, N. S.
    Abdullah, R.
    Aziz, M. Z. Abdul
    Samson, D. O.
    Musarudin, M.
    [J]. APPLIED RADIATION AND ISOTOPES, 2023, 196
  • [36] On the Significance of Model Based Dosimetry for Breast and Head and Neck 192Ir HDR Brachytherapy
    Peppa, V.
    Pappas, E.
    Pantelis, E.
    Major, T.
    Polgar, C.
    Papagiannis, P.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3469 - 3469
  • [37] Validation of electromagnetic physics models and electron range in Geant4 Brachytherapy application
    Albqoor, A.
    Ababneh, E.
    Okoor, S.
    Zahran, I.
    [J]. NUCLEAR ENGINEERING AND TECHNOLOGY, 2023, 55 (01) : 229 - 237
  • [38] MAASTRO applicator, a novel rectal applicator for contact brachytherapy with 192Ir HDR sources
    Bellezzo, M.
    Fonseca, G. P.
    Voncken, R.
    Verrijssen, A.
    Van Beveren, C.
    Roelofs, E.
    Yoriyaz, H.
    Reniers, B.
    Van Limbergen, E. J.
    Barbee, M.
    Verhaegen, F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S1102 - S1103
  • [39] On the accuracy of techniques for obtaining the calibration coefficient NK of 192Ir HDR brachytherapy sources
    Mainegra-Hing, Ernesto
    Rogers, D. W. O.
    [J]. MEDICAL PHYSICS, 2006, 33 (09) : 3340 - 3347
  • [40] Simulation of Ir-192 Source Functions with Geant4 Monte Carlo (MC) Methods
    Walke, M.
    Hass, P.
    Senz, S.
    Scheermann, J.
    Gademann, G.
    Ricke, J.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 : S103 - S103