Computation of residence time in the simulation of pulsatile ventricular assist devices

被引:80
|
作者
Long, C. C. [2 ]
Esmaily-Moghadam, M. [2 ]
Marsden, A. L. [2 ]
Bazilevs, Y. [1 ]
机构
[1] Univ Calif San Diego, Dept Struct Engn, La Jolla, CA 92093 USA
[2] Univ Calif San Diego, Dept Mech & Aerosp Engn, La Jolla, CA 92093 USA
关键词
PVAD; Residence time; FSI; Isogeometric analysis; Biomechanics; FEM; Blood flow; FLUID-STRUCTURE INTERACTION; FINITE-ELEMENT COMPUTATION; DERIVATIVE-FREE OPTIMIZATION; STAGNATION POINT FLOW; PLATELET DEPOSITION; CONSTRAINED OPTIMIZATION; ISOGEOMETRIC ANALYSIS; BLOOD-FLOW; FORMULATION; ALGORITHMS;
D O I
10.1007/s00466-013-0931-y
中图分类号
O1 [数学];
学科分类号
0701 ; 070101 ;
摘要
A continuum-based model of particle residence time for moving-domain fluid mechanics and fluid-structure interaction (FSI) computations is proposed, analyzed, and applied to the simulation of an adult pulsatile ventricular assist device (PVAD). Residence time is a quantity of clinical interest for blood pumps because it correlates with thrombotic risk. The proposed technique may be easily implemented in any flow or FSI solver. In the context of PVADs the results of the model may be used to assess how efficiently the pump moves the blood through its interior. Three scalar measures of particle residence time are also proposed. These scalar quantities may be used in the PVAD design with the goal of reducing thrombotic risk.
引用
收藏
页码:911 / 919
页数:9
相关论文
共 50 条
  • [21] NON-PULSATILE VENTRICULAR ASSIST DEVICES - A COMPREHENSIVE CLINICOPATHOLOGIC STUDY
    SCHWAEGERLE, SM
    MARCHANT, KK
    RATLIFF, NB
    GOLDING, LR
    ESTES, ME
    [J]. LABORATORY INVESTIGATION, 1989, 60 (01) : A85 - A85
  • [22] Pulsatile mode of operation of left ventricular assist devices and pulmonary haemodynamics
    Poullis, Michael
    [J]. INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2014, 19 (01) : 11 - 15
  • [23] Pulsatile flow assist devices
    Aurelio Sarralde, Jose
    Negueruela, Carolina P.
    [J]. CIRUGIA CARDIOVASCULAR, 2009, 16 (02): : 125 - 130
  • [24] Experimental measurement and numerical modelling of dye washout for investigation of blood residence time in ventricular assist devices
    Molteni, Alessandra
    Masri, Zubair P. H.
    Low, Kenny W. Q.
    Yousef, Haitham N.
    Sienz, Johann
    Fraser, Katharine H.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2018, 41 (04): : 201 - 212
  • [25] Neurocognitive function in patients with ventricular assist devices: A comparison of pulsatile and continuous blood flow devices
    Zimpfer, D
    Wieselthaler, G
    Czerny, M
    Fakin, R
    Haider, D
    Zrunek, P
    Roethy, W
    Schima, H
    Wolner, E
    Grimm, M
    [J]. ASAIO JOURNAL, 2006, 52 (01) : 24 - 27
  • [26] Pneumatic pulsatile ventricular assist devices in children under 1 year of age
    Stiller, B
    Weng, YG
    Hübler, M
    Lemmer, J
    Nagdyman, N
    Redlin, M
    Lange, PE
    Hetzer, R
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 28 (02) : 234 - 239
  • [27] In vivo proof of concept of a pulsatile physiologic controller framework for ventricular assist devices
    Salesch, Tobias
    Gesenhues, Jonas
    Habigt, Moriz
    Mechelinck, Mare
    Hein, Marc
    Abel, Dirk
    [J]. 5TH IEEE CONFERENCE ON CONTROL TECHNOLOGY AND APPLICATIONS (IEEE CCTA 2021), 2021, : 1056 - 1062
  • [28] Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devices
    Crow, Sheri
    John, Ranjit
    Boyle, Andrew
    Shumway, Sara
    Liao, Kenneth
    Colvin-Adams, Monica
    Toninato, Carol
    Missov, Emil
    Pritzker, Marc
    Martin, Cindy
    Garry, Daniel
    Thomas, William
    Joyce, Lyle
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (01): : 208 - 215
  • [29] Corrugated diaphragm shape design study for hemocompatible pulsatile ventricular assist devices
    Loosli, C.
    Moy, L.
    Kress, G.
    Mazza, E.
    Ermanni, P.
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2018, 21 (05) : 399 - 407
  • [30] Impact of continuous versus pulsatile left ventricular assist devices on renal function
    Sandner, S. E.
    Zimpfer, D.
    Zrunek, P.
    Schima, H.
    Rajek, A.
    Grimm, M.
    Wolner, E.
    Wieselthaler, G. M.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2007, 26 (02): : S200 - S201