Technical note: Continuous fluorescence-based monitoring of seawater pH in situ

被引:7
|
作者
Runcie, John W. [1 ,2 ,3 ]
Krause, Christian [4 ]
Gabarda, Sergio A. Torres [1 ,3 ]
Byrne, Maria [1 ,3 ]
机构
[1] Univ Sydney, Sch Life & Environm Sci, Sydney, NSW, Australia
[2] Aquat Pty Ltd, Umina Beach, Australia
[3] Sydney Inst Marine Sci, Mosman, Australia
[4] Presens Precis Sensing GmbH, Regensburg, Germany
关键词
OCEAN ACIDIFICATION; CARBONIC-ACID; DISSOCIATION; CONSTANTS;
D O I
10.5194/bg-15-4291-2018
中图分类号
Q14 [生态学(生物生态学)];
学科分类号
071012 ; 0713 ;
摘要
Electrical conductivity (salinity), temperature and fluorescence-based measurements of pH were employed to examine diel fluctuations in seawater carbonate chemistry of surface waters in Sydney Harbour over two multiple-day periods. A proof-of-concept device employing the fluorescence-based technique provided a useful time series for pH. Alkalinity with pH and temperature were used to calculate the degree of calcite and aragonite saturation (Omega Ca and Omega Ar, respectively). Alkalinity was determined from a published alkalinity-salinity relationship. The fluctuations observed in pH over intervals of minutes to hours could be distinguished from background noise. While the stated phase angle resolution of the lifetime fluorometer translated into pH units was +/- 0.0028 pH units, the repeatability standard deviation of calculated pH was 0.007 to 0.009. Diel variability in pH, Omega Ar and Omega Ca showed a clear pattern that appeared to correlate with both salinity and temperature. Drift due to photodegradation of the fluorophore was minimized by reducing exposure to ambient light. The Omega Ca and Omega Ar fluctuated on a daily cycle. The net result of changes in pH, salinity and temperature combined to influence seawater carbonate chemistry. The fluorescence-based pH monitoring technique is simple, provides good resolution and is unaffected by moving parts or leaching of solutions over time. The use of optics is pressure insensitive, making this approach to ocean acidification monitoring well suited to deepwater applications.
引用
收藏
页码:4291 / 4299
页数:9
相关论文
共 50 条
  • [31] Indirect fluorescence-based in situ geometry measurement for laser chemical machining
    Fischer, Andreas
    Mikulewitsch, Merlin
    Stoebener, Dirk
    [J]. CIRP ANNALS-MANUFACTURING TECHNOLOGY, 2020, 69 (01) : 481 - 484
  • [32] A novel fluorescence-based method in forensic science for the detection of blood in situ
    Thorogate, Richard
    Moreira, Joana C. S.
    Jickells, Sue
    Miele, Margherita M. P.
    Daniel, Barbara
    [J]. FORENSIC SCIENCE INTERNATIONAL-GENETICS, 2008, 2 (04) : 363 - 371
  • [33] Insertion Site Performance Differences for a Fluorescence-Based Continuous Glucose Monitor
    Chen, Xiaoxiao
    Wang, Xiaolin
    Rajaraman, Srinivasan
    Dehennis, Andrew
    Whitehurst, Todd
    [J]. DIABETES, 2013, 62 : A45 - A45
  • [34] A sensor for in situ indicator-based measurements of seawater pH
    Seidel, Matthew P.
    DeGrandpre, Michael D.
    Dickson, Andrew G.
    [J]. MARINE CHEMISTRY, 2008, 109 (1-2) : 18 - 28
  • [35] A Technical Note on Quantum Dots for Multi-Color Fluorescence in situ Hybridization
    Mueller, S.
    Cremer, M.
    Neusser, M.
    Grasser, F.
    Cremer, T.
    [J]. CYTOGENETIC AND GENOME RESEARCH, 2009, 124 (3-4) : 351 - 359
  • [36] Angled rigid neuroendoscope for continuous intraoperative visual monitoring: technical note
    Tamaki, N
    Hara, Y
    Takaishi, Y
    Shimada, S
    [J]. JOURNAL OF CLINICAL NEUROSCIENCE, 2001, 8 (02) : 148 - 150
  • [37] FLUORESCENCE-BASED MONITORING OF INTERLEUKIN-2 GENE-EXPRESSION
    EMILIE, D
    MATTHES, M
    GREGOIRE, C
    LETOURNEUR, F
    WEGENER, AM
    NICOLAS, JF
    MALISSEN, B
    [J]. IMMUNOLOGICAL REVIEWS, 1991, 119 : 95 - 103
  • [38] FLUORESCENCE-BASED METHODS FOR MONITORING OF NUCLEOTIDE BINDING TO G-PROTEINS
    Rinken, A.
    Tontson, L.
    Kopanchuk, S.
    [J]. BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2014, 115 : 262 - 263
  • [39] A dansyl fluorescence-based assay for monitoring kinetics of lipid extraction and transfer
    Ran, Yong
    Fanucci, Gail E.
    [J]. ANALYTICAL BIOCHEMISTRY, 2008, 382 (02) : 132 - 134
  • [40] Optical monitoring of therapeutic drugs with a novel fluorescence-based POCT device
    Berrettoni, C.
    Berneschi, S.
    Bernini, R.
    Giannetti, A.
    Grimaldi, I. A.
    Persichetti, G.
    Testa, G.
    Tombelli, S.
    Trono, C.
    Baldini, F.
    [J]. 28TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS (EUROSENSORS 2014), 2014, 87 : 392 - 395