In Vitro Evaluation of the Gore Viabahn Balloon-Expandable Stent-Graft for Fenestrated Endovascular Aortic Repair

被引:18
|
作者
Torsello, Giovanni Federico [1 ,2 ]
Herten, Monika [2 ,3 ]
Mueller, Markus [4 ]
Andre-Frank [4 ]
Torsello, Giovanni B. [2 ]
Austermann, Martin [2 ]
机构
[1] Charite Univ Med Berlin, Charite Campus Virchow Klinikum, Dept Diagnost & Intervent Radiol & Nucl Med, Augustenburger Pl 1, D-13353 Berlin, Germany
[2] St Franziskus Hosp Munster, Dept Vasc Surg, Munster, Germany
[3] Essen Univ Hosp, Dept Orthoped & Trauma Surg, Essen, Germany
[4] Univ Hosp Munster, Dept Hand Trauma & Reconstruct Surg, Biomech Lab, Munster, Germany
关键词
bridging stent-graft; displacement force; fenestrated stent-graft; in vitro experiment; in vitro model; material fatigue; permeability; stent fracture; ENDOGRAFTS; COMPONENTS; OUTCOMES; SECURE;
D O I
10.1177/1526602819842569
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To evaluate the safety and integrity of a new stent-graft as a potential bridging device in fenestrated stent-grafts using an in vitro fenestrated model. Materials and Methods: Polyester test sheets with ten 6-mm- or 8-mm-diameter fenestrations were used to simulate a fenestrated main body endoprosthesis. In total, 50 Viabahn balloon-expandable (VBX) stent-grafts of varying lengths (29 and 39 mm) and diameters (6, 7, and 8 mm) were implanted in the fitting fenestrations. After release, the 6- and 7-mm-diameter stent-grafts were flared with a 10x20-mm angioplasty balloon; a 12x20-mm balloon was used in the 8-mm-diameter devices. Safety of the devices was defined as absence of fractures detected on radiography or computed tomography (CT), as well as material failure detected by microscopy and water permeability testing. The forces (in Newtons, N) needed for perpendicular dislocation (pullout force) and axial dislocation (shear stress force) were also evaluated. Results: Forty VBX stent-grafts were subjected to digital radiographic imaging and multiplanar CT. None showed any stent fracture. Subsequent microscopy indicated no damage to the fabric or separation of the graft after flaring. Ten VBX stent-grafts underwent water permeability testing after flaring; no water passed through the graft wall during a 10-minute period under an intraluminal pressure at 120 mm Hg. Testing of 25 VBX stent-grafts revealed initial pullout forces between 11.3 and 31 N. Shear stress tests showed that the average force needed to dislocate the stent-grafts by 50% of their diameter ranged between 5.75 and 6.91 N (mean 6.1 +/- 0.5 N) for the 6-mm stents and between 3.31 and 5.4 N (mean 4.4 +/- 0.8) for the 8-mm stents. Conclusion: This preliminary study demonstrated the applicability of the VBX as a bridging stent-graft in a simulated fenestration model. A comparison with other stent-grafts and clinical assessment are required.
引用
收藏
页码:361 / 368
页数:8
相关论文
共 50 条
  • [1] First/Preliminary Experience of Gore Viabahn Balloon-Expandable Endoprosthesis as Bridging Stent in Fenestrated and Branched Endovascular Aortic Repair
    Gallitto, Enrico
    Faggioli, Gianluca
    Pini, Rodolfo
    Mascoli, Chiara
    Sonetto, Alessia
    Abualhin, Mohammad
    Logiacco, Antonino
    Ricco, Jean-Baptiste
    Gargiulo, Mauro
    [J]. ANNALS OF VASCULAR SURGERY, 2019, 61 : 299 - 309
  • [2] Preliminary Outcomes of Viabahn Balloon-Expandable Endoprosthesis as Bridging Stent in Renal Arteries During Fenestrated Endovascular Aortic Repair
    Mezzetto, Luca
    Mastrorilli, Davide
    Leone, Nicola
    Gennai, Stefano
    Silingardi, Roberto
    Veraldi, Gian Franco
    Piazza, Michele
    Squizzato, Francesco
    Antonello, Michele
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2021, 28 (04) : 575 - 584
  • [3] Treatment of Coral Reef Aorta by Endovascular VIABAHN VBX Balloon-Expandable Stent-Graft Placement
    Myouchin, Kaoru
    Takayama, Katsutoshi
    Wada, Takeshi
    Taguchi, Hidehiko
    Tanaka, Toshihiro
    Kichikawa, Kimihiko
    [J]. ANNALS OF VASCULAR DISEASES, 2021, 14 (03) : 244 - 248
  • [4] Performance of Viabahn Balloon-Expandable Stent Compared With Self-expanding Covered Stents for Fenestrated-Branched Endovascular Aortic Repair
    Motta, Fernando
    Parodi, Frederico E.
    Crowner, Jason R.
    Pas-Carella, Luigi
    McGinigle, Katharine L.
    Marston, William A.
    Kibbe, Melina R.
    Farber, Mark A.
    [J]. JOURNAL OF VASCULAR SURGERY, 2020, 71 (01) : E28 - E29
  • [5] Performance of the Gore VBX Balloon Expandable Endoprosthesis as Bridging Stent-Graft in Branched Endovascular Aortic Repair for Thoracoabdominal Aneurysms
    Torsello, Giovanni B.
    Pitoulias, Apostolos
    Litterscheid, Sarah
    Berekoven, Barbel
    Torsello, Giovanni-Federico
    Austermann, Martin
    Bosiers, Michel J.
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2021, 28 (04) : 549 - 554
  • [6] Performance of Viabahn balloon-expandable stent compared with self-expandable covered stents for branched endovascular aortic repair
    Motta, Fernando
    Parodi, F. Ezequiel
    Knowles, Martyn
    Crowner, Jason R.
    Pascarella, Luigi
    McGinigle, Katharine L.
    Marston, William A.
    Kibbe, Melina R.
    Ohana, Elad
    Farber, Mark A.
    [J]. JOURNAL OF VASCULAR SURGERY, 2021, 73 (02) : 410 - +
  • [7] Fatigue Resistance of the Advanta V12/iCast and Viabahn Balloon-Expandable Stent-Graft as Bridging Stents in Experimental Fenestrated Endografting
    Torsello, Giovanni
    Mueller, Marcus
    Litterscheid, Sarah
    Berekoven, Baerbel
    Austermann, Martin
    Torsello, Giovanni-Federico
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2021, 28 (03) : 463 - 468
  • [8] Endovascular Repair of Ascending Aortic Rupture: Effectiveness of a Fenestrated Stent-Graft
    Uchida, Keiji
    Imoto, Kiyotaka
    Yanagi, Hiromasa
    Machida, Daisuke
    Okiyama, Makoto
    Yasuda, Shota
    Sugiura, Tadahisa
    Kawaguchi, Satoshi
    Yokoi, Yoshihiko
    Shigematsu, Hiroshi
    Masuda, Munetaka
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2010, 17 (03) : 395 - 398
  • [9] Ventana Fenestrated Stent-Graft System for Endovascular Repair of Juxtarenal Aortic Aneurysms
    Mertens, Renato
    Bergoeing, Michel
    Marine, Leopoldo
    Valdes, Francisco
    Kraemer, Albrecht
    Vergara, Jeanette
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2012, 19 (02) : 173 - 178
  • [10] Endovascular repair of abdominal aortic aneurysm using a pararenal fenestrated stent-graft
    Faruqi, RM
    Chuter, TAM
    Reilly, LM
    Sawhney, R
    Wall, S
    Canto, C
    Messina, LM
    [J]. JOURNAL OF ENDOVASCULAR SURGERY, 1999, 6 (04): : 354 - 358