Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose

被引:2
|
作者
Khasanah, Miratul [1 ]
Widati, Alfa Akustia [1 ]
Handajani, Usreg Sri [1 ]
Shofiyyah, Masfah Raudlotus [1 ]
Rakhma, Sabrina Aulia [1 ]
Predianto, Herwin [1 ]
机构
[1] Airlangga Univ, Fac Sci & Technol, Chem Dept, Kampus C Unair, Surabaya 60115, Indonesia
来源
关键词
CONSTRUCTION;
D O I
10.1063/5.0005231
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
Imprinted zeolite modified carbon paste electrode as a potentiometric sensor for glucose analysis in blood serum sample has been developed. In this study, the type of used zeolite was TS-1. Imprinted zeolite (IZ) was synthesized with mole ratio of glucose/Si of 0.0306. The modified electrode prepared by mixing activated carbon, imprinted zeolite, and paraffin granule with mass ratio 9:4:7 from the measurement in range of 10(-5) - 10(-2) M is with Nernst factor of 28.6 mV/decade, and the detection limit of 4.79 x 10(-5) M. The electrode was able to respond the glucose molecules in solution quickly (<30 s), stable for more than 9 weeks (130 times usage) and selective to glucose molecules. The urea, uric acid, and creatine with various concentrations did not interfere to glucose analysis by potentiometry using the electrode. Applying of the electrode for glucose analysis in spiked blood serum samples showed that it had recovery of 91.73 +/- 3.48% (n=5), while its accuracy against spectrophotometry method as commonly used method in medical field was 90.58 +/- 4.65% (n=5). Based on its performance, potentiometry using the imprinted zeolite modified carbon paste electrode is recommended as an alternative method for routine blood glucose analysis in the medical field.
引用
收藏
页数:8
相关论文
共 50 条