Calibration-free wavelength-modulation spectroscopy for measurements of gas temperature and concentration in harsh environments

被引:452
|
作者
Rieker, Gregory B. [1 ]
Jeffries, Jay B. [1 ]
Hanson, Ronald K. [1 ]
机构
[1] Stanford Univ, Dept Mech Engn, High Temp Gasdynam Lab, Stanford, CA 94305 USA
关键词
TUNABLE DIODE-LASER; WATER-VAPOR; HIGH-PRESSURE; ABSORPTION-SPECTROSCOPY; FREQUENCY-MODULATION; MU-M; SENSOR; SPECTROMETRY; COMBUSTION; FLOWS;
D O I
10.1364/AO.48.005546
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We present a practical implementation of calibration-free wavelength-modulation spectroscopy with second harmonic detection (WMS-2f) for measurements of gas temperature and concentration in harsh environments. The method is applicable to measurements using lasers with synchronous wavelength and intensity modulation (such as injection current-tuned diode lasers). The key factors that enable measurements without the on-site calibration normally associated with WMS are (1) normalization of the WMS-2f signal by the first harmonic (1f) signal to account for laser intensity, and (2) the inclusion of laser-specific tuning characteristics in the spectral-absorption model that is used to compare with measured 1f-normalized, WMS-2f signals to infer gas properties. The uncertainties associated with the calibration-free WMS method are discussed, with particular emphasis on the influence of pressure and optical depth on the WMS signals. Many of these uncertainties are also applicable to calibrated WMS measurements. An example experimental setup that combines six tunable diode laser sources between 1.3 and 2.0 mu m into one probe beam for measurements of temperature, H2O, and CO2 is shown. A hybrid combination of wavelength and frequency demultiplexing is used to distinguish among the laser signals, and the optimal set of laser-modulation waveforms is presented. The system is demonstrated in the harsh environment of a ground-test scramjet combustor. A comparison of direct absorption and 1f-normalized, WMS-2f shows a factor of 4 increase in signal-to-noise ratio with the WMS technique for measurements of CO2 in the supersonic flow. Multidimensional computational fluid-dynamics (CFD) calculations are compared with measurements of temperature and H2O using a simple method that accounts for the influence of line-of-sight (LOS) nonuniformity on the absorption measurements. The comparisons show the ability of the LOS calibration-free technique to gain useful information about multidimensional CFD models. (C) 2009 Optical Society of America
引用
收藏
页码:5546 / 5560
页数:15
相关论文
共 50 条
  • [41] Simulation technique enabling calibration-free frequency-modulation spectroscopy measurements of gas conditions and lineshapes with modulation frequencies spanning kHz to GHz
    Goldenstein, Christopher S.
    Mathews, Garrett C.
    APPLIED OPTICS, 2020, 59 (05) : 1491 - 1500
  • [42] Gas temperature measurements using wavelength modulation spectroscopy at 1.39 μm
    Cai, Tingdong
    Tan, Tu
    Wang, Guishi
    Chen, Weidong
    Gao, Xiaoming
    OPTICA APPLICATA, 2009, 39 (01) : 13 - 29
  • [43] Towards an Optical Gas Standard for Traceable Calibration-Free and Direct NO2 Concentration Measurements
    Nwaboh, Javis A.
    Qu, Zhechao
    Werhahn, Olav
    Ebert, Volker
    APPLIED SCIENCES-BASEL, 2021, 11 (12):
  • [44] In situ H2O and temperature detection close to burning biomass pellets using calibration-free wavelength modulation spectroscopy
    Qu, Zhechao
    Schmidt, Florian M.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2015, 119 (01): : 45 - 53
  • [45] In situ H2O and temperature detection close to burning biomass pellets using calibration-free wavelength modulation spectroscopy
    Zhechao Qu
    Florian M. Schmidt
    Applied Physics B, 2015, 119 : 45 - 53
  • [46] Method for Measuring High Temperature Spectral Line Parameters Based on Calibration-Free Wavelength Modulation Technology
    Zang Yipeng
    Xu Zhenyu
    Xia Huihui
    Huang An
    He Yabai
    Kan Ruifeng
    CHINESE JOURNAL OF LASERS-ZHONGGUO JIGUANG, 2020, 47 (10):
  • [47] Method for Measuring High Temperature Spectral Line Parameters Based on Calibration-Free Wavelength Modulation Technology
    Zang Y.
    Xu Z.
    Xia H.
    Huang A.
    He Y.
    Kan R.
    1600, Science Press (47):
  • [48] Calibration-free breath acetone sensor with interference correction based on wavelength modulation spectroscopy near 8.2 μm
    Schwarm, Kevin K.
    Strand, Christopher L.
    Miller, Victor A.
    Spearrin, R. Mitchell
    APPLIED PHYSICS B-LASERS AND OPTICS, 2020, 126 (01):
  • [49] A Wide-Range and Calibration-Free Spectrometer Which Combines Wavelength Modulation and Direct Absorption Spectroscopy with Cavity Ringdown Spectroscopy
    Wang, Zhen
    Du, Yanjun
    Ding, Yanjun
    Peng, Zhimin
    SENSORS, 2020, 20 (03)
  • [50] Photoacoustic correlation spectroscopy for calibration-free absolute quantification of particle concentration
    Zhou, Yong
    Yao, Junjie
    Maslov, Konstantin I.
    Wang, Lihong V.
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2014, 2014, 8943