Improved Accuracy with 3D Planning and Patient-Specific Instruments During Simulated Pelvic Bone Tumor Surgery

被引:131
|
作者
Cartiaux, Olivier [1 ]
Paul, Laurent [1 ]
Francq, Bernard G. [2 ]
Banse, Xavier [1 ,3 ]
Docquier, Pierre-Louis [1 ,3 ]
机构
[1] UCL, IREC, B-1200 Brussels, Belgium
[2] Catholic Univ Louvain, Inst Stat Biostat & Sci Actuarielles, B-1348 Louvain, Belgium
[3] Clin Univ St Luc, Serv Orthope & Traumatol Appareil Locomoteur, B-1200 Brussels, Belgium
关键词
Tumor resection; Safe margin; Patient-specific cutting guides; Quantitative surgery; ISO-based evaluation; ASSISTED ORTHOPEDIC-SURGERY; TOTAL KNEE ARTHROPLASTY; RESECTION; RECONSTRUCTION; OSTEOSARCOMA; PERSPECTIVES; TEMPLATES; OSTEOTOMY; ALIGNMENT;
D O I
10.1007/s10439-013-0890-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In orthopaedic surgery, resection of pelvic bone tumors can be inaccurate due to complex geometry, limited visibility and restricted working space of the pelvis. The present study investigated accuracy of patient-specific instrumentation (PSI) for bone-cutting during simulated tumor surgery within the pelvis. A synthetic pelvic bone model was imaged using a CT-scanner. The set of images was reconstructed in 3D and resection of a simulated periacetabular tumor was defined with four target planes (ischium, pubis, anterior ilium, and posterior ilium) with a 10-mm desired safe margin. Patient-specific instruments for bone-cutting were designed and manufactured using rapid-prototyping technology. Twenty-four surgeons (10 senior and 14 junior) were asked to perform tumor resection. After cutting, ISO1101 location and flatness parameters, achieved surgical margins and the time were measured. With PSI, the location accuracy of the cut planes with respect to the target planes averaged 1 and 1.2 mm in the anterior and posterior ilium, 2 mm in the pubis and 3.7 mm in the ischium (p < 0.0001). Results in terms of the location of the cut planes and the achieved surgical margins did not reveal any significant difference between senior and junior surgeons (p = 0.2214 and 0.8449, respectively). The maximum differences between the achieved margins and the 10-mm desired safe margin were found in the pubis (3.1 and 5.1 mm for senior and junior surgeons respectively). Of the 24 simulated resection, there was no intralesional tumor cutting. This study demonstrates that using PSI technology during simulated bone cuts of the pelvis can provide good cutting accuracy. Compared to a previous report on computer assistance for pelvic bone cutting, PSI technology clearly demonstrates an equivalent value-added for bone cutting accuracy than navigation technology. When in vivo validated, PSI technology may improve pelvic bone tumor surgery by providing clinically acceptable margins.
引用
收藏
页码:205 / 213
页数:9
相关论文
共 50 条
  • [41] Accuracy of a patient-specific 3D printed drill guide for placement of cervical transpedicular screws
    Hamilton-Bennett, Sinead E.
    Oxley, Bill
    Behr, Sebastien
    VETERINARY SURGERY, 2018, 47 (02) : 236 - 242
  • [42] Feasibility and accuracy of 3D printed patient-specific skull contoured brain biopsy guides
    Shinn, Richard
    Park, Clair
    DeBose, Kyrille
    Hsu, Fang-Chi
    Cecere, Thomas
    Rossmeisl, John
    VETERINARY SURGERY, 2021, 50 (05) : 933 - 943
  • [43] Quality Control in 3D Printing: Accuracy Analysis of 3D-Printed Models of Patient-Specific Anatomy
    Dorweiler, Bernhard
    Baque, Pia Elisabeth
    Chaban, Rayan
    Ghazy, Ahmed
    Salem, Oroa
    MATERIALS, 2021, 14 (04) : 1 - 13
  • [44] Craniosynostosis surgery: workflow based on virtual surgical planning, intraoperative navigation and 3D printed patient-specific guides and templates
    David García-Mato
    Santiago Ochandiano
    Mónica García-Sevilla
    Carlos Navarro-Cuéllar
    Juan V. Darriba-Allés
    Roberto García-Leal
    José A. Calvo-Haro
    Rubén Pérez-Mañanes
    José I. Salmerón
    Javier Pascau
    Scientific Reports, 9
  • [45] Development of a 3D patient-specific planning platform for interstitial and transurethral ultrasound thermal therapy
    Prakash, Punit
    Diederich, Chris J.
    9TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2010, 1215 : 396 - 399
  • [46] Craniosynostosis surgery: workflow based on virtual surgical planning, intraoperative navigation and 3D printed patient-specific guides and templates
    Garcia-Mato, David
    Ochandiano, Santiago
    Garcia-Sevilla, Monica
    Navarro-Cuellar, Carlos
    Darriba-Alles, Juan V.
    Garcia-Leal, Roberto
    Calvo-Haro, Jose A.
    Perez-Mananes, Ruben
    Salmeron, Jose I.
    Pascau, Javier
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [47] 3D Observation of Pelvic Organs with Dynamic MRI Segmentation: A Bridge Toward Patient-Specific Models
    Omouri, Adel
    Rapacchi, Stanislas
    Duclos, Julie
    Niddam, Raphael
    Bellemare, Marc-Emmanuel
    Pirro, Nicolas
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2024, 35 (07) : 1389 - 1397
  • [48] Does functional planning, 3D templating and patient-specific instrumentation improve accuracy in total hip replacement?— a randomized controlled trial
    Christopher Thomas
    Vatsal Gupta
    Helen Parsons
    Andrew Metcalfe
    Pedro Foguet
    Richard King
    Arthroplasty, 4
  • [49] Does functional planning, 3D templating and patient-specific instrumentation improve accuracy in total hip replacement?- a randomized controlled trial
    Thomas, Christopher
    Gupta, Vatsal
    Parsons, Helen
    Metcalfe, Andrew
    Foguet, Pedro
    King, Richard
    ARTHROPLASTY, 2022, 4 (01)
  • [50] 3D Modelling and Printing Technology to Produce Patient-Specific 3D Models
    Birbara, Nicolette S.
    Otton, James M.
    Pather, Nalini
    HEART LUNG AND CIRCULATION, 2019, 28 (02): : 302 - 313