Improved Accuracy with 3D Planning and Patient-Specific Instruments During Simulated Pelvic Bone Tumor Surgery

被引:131
|
作者
Cartiaux, Olivier [1 ]
Paul, Laurent [1 ]
Francq, Bernard G. [2 ]
Banse, Xavier [1 ,3 ]
Docquier, Pierre-Louis [1 ,3 ]
机构
[1] UCL, IREC, B-1200 Brussels, Belgium
[2] Catholic Univ Louvain, Inst Stat Biostat & Sci Actuarielles, B-1348 Louvain, Belgium
[3] Clin Univ St Luc, Serv Orthope & Traumatol Appareil Locomoteur, B-1200 Brussels, Belgium
关键词
Tumor resection; Safe margin; Patient-specific cutting guides; Quantitative surgery; ISO-based evaluation; ASSISTED ORTHOPEDIC-SURGERY; TOTAL KNEE ARTHROPLASTY; RESECTION; RECONSTRUCTION; OSTEOSARCOMA; PERSPECTIVES; TEMPLATES; OSTEOTOMY; ALIGNMENT;
D O I
10.1007/s10439-013-0890-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In orthopaedic surgery, resection of pelvic bone tumors can be inaccurate due to complex geometry, limited visibility and restricted working space of the pelvis. The present study investigated accuracy of patient-specific instrumentation (PSI) for bone-cutting during simulated tumor surgery within the pelvis. A synthetic pelvic bone model was imaged using a CT-scanner. The set of images was reconstructed in 3D and resection of a simulated periacetabular tumor was defined with four target planes (ischium, pubis, anterior ilium, and posterior ilium) with a 10-mm desired safe margin. Patient-specific instruments for bone-cutting were designed and manufactured using rapid-prototyping technology. Twenty-four surgeons (10 senior and 14 junior) were asked to perform tumor resection. After cutting, ISO1101 location and flatness parameters, achieved surgical margins and the time were measured. With PSI, the location accuracy of the cut planes with respect to the target planes averaged 1 and 1.2 mm in the anterior and posterior ilium, 2 mm in the pubis and 3.7 mm in the ischium (p < 0.0001). Results in terms of the location of the cut planes and the achieved surgical margins did not reveal any significant difference between senior and junior surgeons (p = 0.2214 and 0.8449, respectively). The maximum differences between the achieved margins and the 10-mm desired safe margin were found in the pubis (3.1 and 5.1 mm for senior and junior surgeons respectively). Of the 24 simulated resection, there was no intralesional tumor cutting. This study demonstrates that using PSI technology during simulated bone cuts of the pelvis can provide good cutting accuracy. Compared to a previous report on computer assistance for pelvic bone cutting, PSI technology clearly demonstrates an equivalent value-added for bone cutting accuracy than navigation technology. When in vivo validated, PSI technology may improve pelvic bone tumor surgery by providing clinically acceptable margins.
引用
收藏
页码:205 / 213
页数:9
相关论文
共 50 条
  • [31] The validity and accuracy of 3D-printed patient-specific instruments for high tibial osteotomy: a cadaveric study
    Zhuang Miao
    Songlin Li
    Desu Luo
    Qunshan Lu
    Peilai Liu
    Journal of Orthopaedic Surgery and Research, 17
  • [32] The validity and accuracy of 3D-printed patient-specific instruments for high tibial osteotomy: a cadaveric study
    Miao, Zhuang
    Li, Songlin
    Luo, Desu
    Lu, Qunshan
    Liu, Peilai
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2022, 17 (01)
  • [33] Optimization of 3D Print Material for the Recreation of Patient-Specific Temporal Bone Models
    Haffner, Max
    Quinn, Austin
    Hsieh, Tsung-yen
    Strong, E. Bradley
    Steele, Toby
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2018, 127 (05): : 338 - 343
  • [34] OPTIMISATION OF 3D PRINT MATERIAL FOR THE RECREATION OF PATIENT-SPECIFIC TEMPORAL BONE MODELS
    Quinn, A.
    Haffner, M. R.
    Hsieh, T.
    Steele, T.
    Strong, E.
    JOURNAL OF INVESTIGATIVE MEDICINE, 2018, 66 (01) : 130 - 130
  • [35] Bioprinting of 3D patient-specific auricular scaffolds
    Martinez, H.
    Sundberg, J.
    Prakash, S.
    Seniger, M.
    Gatenholm, P.
    JOURNAL OF TISSUE ENGINEERING AND REGENERATIVE MEDICINE, 2012, 6 : 153 - 153
  • [36] Applications of patient-specific 3D printing in medicine
    Heller, M.
    Bauer, H-K.
    Goetze, E.
    Gielisch, M.
    Roth, K. E.
    Maier, G. S.
    Drees, P.
    Dorweiler, B.
    Ghazy, A.
    Neufurth, M.
    Mueller, W. E. G.
    Schoerder, H. C.
    Wang, X. H.
    Vahl, C-F
    Al-Nawas, B.
    INTERNATIONAL JOURNAL OF COMPUTERIZED DENTISTRY, 2016, 19 (04) : 323 - 339
  • [37] Intraoperative patient-specific volumetric reconstruction and 3D visualization for laparoscopic liver surgery
    Boretto, Luca
    Pelanis, Egidijus
    Regensburger, Alois
    Petkov, Kaloian
    Palomar, Rafael
    Fretland, Asmund Avdem
    Edwin, Bjrn
    Elle, Ole Jakob
    HEALTHCARE TECHNOLOGY LETTERS, 2024, 11 (06) : 374 - 383
  • [38] Patient-Specific 3D Burn Size Estimation
    Nguyen, Kim-Ngan
    Yong, Han Ching
    Sim, Terence
    CLINICAL IMAGE-BASED PROCEDURES, CLIP 2024, 2024, 15196 : 53 - 62
  • [39] An overview of 3D printing and the orthopaedic application of patient-specific models in malunion surgery
    Thomas, Claire N.
    Mavrommatis, Sophia
    Schroder, Lisa K.
    Cole, Peter A.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (03): : 977 - 983
  • [40] Computer-assisted planning and navigation improves cutting accuracy during simulated bone tumor surgery of the pelvis
    Cartiaux, Olivier
    Banse, Xavier
    Paul, Laurent
    Francq, Bernard G.
    Aubin, Carl-Eric
    Docquier, Pierre-Louis
    COMPUTER AIDED SURGERY, 2013, 18 (1-2) : 19 - 26