On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry

被引:56
|
作者
Andreo, Pedro [1 ]
Palmans, Hugo [2 ,3 ]
Marteinsdottir, Maria [1 ]
Benmakhlouf, Hamza [1 ]
Carlsson-Tedgren, Asa [1 ]
机构
[1] Karolinska Univ Hosp, Dept Med Phys, SE-17176 Stockholm, Sweden
[2] Natl Phys Lab, Acoust & Ionising Radiat Div, Teddington TW11 0LW, Middx, England
[3] EBG MedAustron GmbH, Dept Med Phys, A-2700 Wiener Neustadt, Austria
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 01期
关键词
small field dosimetry; Monte Carlo; output correction factors; detector geometry accuracy; CHAMBERS;
D O I
10.1088/0031-9155/61/1/L1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Monte Carlo (MC) calculated detector-specific output correction factors for small photon beam dosimetry are commonly used in clinical practice. The technique, with a geometry description based on manufacturer blueprints, offers certain advantages over experimentally determined values but is not free of weaknesses. Independent MC calculations of output correction factors for a PTW-60019 micro-diamond detector were made using the EGSnrc and PENELOPE systems. Compared with published experimental data the MC results showed substantial disagreement for the smallest field size simulated (5 mm x 5 mm). To explain the difference between the two datasets, a detector was imaged with x rays searching for possible anomalies in the detector construction or details not included in the blueprints. A discrepancy between the dimension stated in the blueprints for the active detector area and that estimated from the electrical contact seen in the x-ray image was observed. Calculations were repeated using the estimate of a smaller volume, leading to results in excellent agreement with the experimental data. MC users should become aware of the potential differences between the design blueprints of a detector and its manufacturer production, as they may differ substantially. The constraint is applicable to the simulation of any detector type. Comparison with experimental data should be used to reveal geometrical inconsistencies and details not included in technical drawings, in addition to the well-known QA procedure of detector x-ray imaging.
引用
收藏
页码:L1 / L10
页数:10
相关论文
共 50 条
  • [41] Application of radiochromic gel dosimetry to commissioning of a megavoltage research linear accelerator for small-field animal irradiation studies
    Sunbul, Noora Ba
    Oraiqat, Ibrahim
    Rosen, Benjamin
    Miller, Cameron
    Meert, Christopher
    Matuszak, Martha M.
    Clarke, Shaun
    Pozzi, Sara
    Moran, Jean M.
    El Naqa, Issam
    MEDICAL PHYSICS, 2021, 48 (03) : 1404 - 1416
  • [42] Monte Carlo study on electron contamination and output factors of small field dosimetry in 6 MV photon beam
    Yani, Sitti
    Dirgayussa, I. Gde Eka
    Rhani, Mohamad Fahdillah
    Soh, Roger C. X.
    Haryanto, Freddy
    Arif, Idam
    SMART SCIENCE, 2016, 4 (02): : 87 - 94
  • [43] A Novel Micro-Dosimetry Technique to Advance Small-Field Radiation Therapy Treatment
    Debnath, S. B. C.
    Tonneau, D.
    Fauquet, C.
    Tallet, A. Richard
    Ferre, M.
    Darreon, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : S87 - S87
  • [44] Detailed Monte Carlo simulation of energy integrating and photon counting semiconductor detectors
    Mendesa, Hitalo R.
    Tomala, Alessandra
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [45] Monte Carlo modeling of converging small-field contrast-enhanced radiotherapy of prostate
    Garnica-Garza, H. M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (05): : 493 - 499
  • [46] Design and dosimetry validation for micro-irradiator using Monte Carlo simulation
    Rodriguez, M.
    Jeraj, R.
    MEDICAL PHYSICS, 2007, 34 (06) : 2632 - 2633
  • [47] Bone and mucosal dosimetry in skin radiation therapy: a Monte Carlo study using kilovoltage photon and megavoltage electron beams
    Chow, James C. L.
    Jiang, Runqing
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12): : 3885 - 3899
  • [48] Estimation of Cable Leakage of Ionization Chamber in Megavoltage Photon Beam by Monte Carlo Simulation and Theoretical Analysis
    Yamauchi, R.
    Igari, M.
    Kasai, Y.
    Hariu, M.
    Suda, Y.
    Kawachi, T.
    Katayose, T.
    Mizuno, N.
    Miyasaka, R.
    Saitoh, H.
    MEDICAL PHYSICS, 2020, 47 (06) : E459 - E460
  • [49] Monte Carlo linear accelerator simulation of megavoltage photon beams: Independent determination of initial beam parameters
    Almberg, Sigrun Saur
    Frengen, Jomar
    Kylling, Arve
    Lindmo, Tore
    MEDICAL PHYSICS, 2012, 39 (01) : 40 - 47
  • [50] Evaluation of the RayStation Monte Carlo Dose Algorithm for Small-Field Surface-Collimated Electrons
    Fenoli, J.
    Oakey, M.
    Andrew, S.
    McGurk, R.
    MEDICAL PHYSICS, 2020, 47 (06) : E775 - E775