Conversion factors for effective dose and organ doses with the air Kerma area product in hysterosalpingography

被引:2
|
作者
D'Alessio, Andrea [1 ]
Cannillo, Barbara [1 ]
Guzzardi, Giuseppe [2 ]
Cernigliaro, Massimiliano [2 ]
Carriero, Alessandro [2 ]
Brambilla, Marco [1 ]
机构
[1] Univ Hosp Maggiore Carita, Med Phys Dept, Novara, Italy
[2] Univ Hosp Maggiore Carita, Radiol Dept, Novara, Italy
关键词
Histerosalpingography; Conversion factors; Dosimetry; Effective dose; Organ dose; OPTIMIZATION;
D O I
10.1016/j.ejmp.2020.11.032
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Histerosalpingography (HSG) remains the dominant diagnostic tool for investigation of infertility in women. Conversion factors used to estimate effective (E) and organ doses (H-T) from air Kerma area product (KAP) are needed to estimate patient doses in HSG, performed with state-of-the-art fluoroscopic X-ray systems with digital detectors. In this study, estimates of E and H-T for six critical organs/tissues, were derived on an individual basis in 120 HSG procedures and in 1410 irradiation events, performed on two X-ray systems from information available through the radiation dose structured report using Monte Carlo methods. Mean values of E and H-ovaries were 1.0 +/- 0.9 mSv and 5.6 +/- 5.4 mGy. E/KAP conversion factors of 0.13; 0.18; 0.28 and 0.35 mSv Gy(-1) cm(-2) were established for irradiation events with a Cu filtration of 0.0; 0.1; 0.4 and 0.9 mm. A high agreement was obtained between E estimated through Monte Carlo methods and E/KAP conversion factors accounting separately for the different modes of fluoroscopy and the radiography component of HSG, with a systematic error of 0 mSv and lower/upper limits of agreement of -0.6 and 0.5 mSv. On the contrary, the use of a single coefficient of conversion did not provide accurate estimates of E, showing a bias of -0.4 mSv and lower and upper limits of agreement of -1.9 and 1.2 mSv. An algorithm for the estimation of effective and organ doses from KAP has been established in HSG procedures depending on the Cu filtration in the X-ray irradiation events.
引用
收藏
页码:40 / 46
页数:7
相关论文
共 50 条
  • [41] Determination of dose-area product from panoramic radiography using a pencil ionization chamber: Normalized data for the estimation of patient effective and organ doses
    Perisinakis, K
    Damilakis, J
    Neratzoulakis, J
    Gourtsoyiannis, N
    MEDICAL PHYSICS, 2004, 31 (04) : 708 - 714
  • [42] Dose-Area Product to Effective Dose Conversion Coefficients for Pelvic Radiography Using a Monte Carlo Program
    Chen, David S.
    Escobedo, Eva M.
    Eastman, Jonathan G.
    Bloomstein, Joshua D.
    Taylor, Sandra L.
    Seibert, J. Anthony
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 215 (03) : 679 - 684
  • [43] An approach for the online determination of Entrance Surface Air Kerma (ESAK) from Kerma Area product (PKA) in digital radiology
    Ibrahim I. Suliman
    Physical and Engineering Sciences in Medicine, 2022, 45 : 1055 - 1061
  • [44] CONVERSION COEFFICIENTS FROM AIR KERMA TO DIRECTIONAL DOSE EQUIVALENT H'(0.07) FOR PHOTONS
    Bairlein, Katharina
    Behrens, Rolf
    Hupe, Oliver
    Hranitzky, Christian
    RADIATION PROTECTION DOSIMETRY, 2019, 186 (04) : 513 - 519
  • [45] AIR KERMA TO DOSE EQUIVALENT CONVERSION COEFFICIENTS NOT INCLUDED IN ISO 4037-3
    Behrens, Rolf
    RADIATION PROTECTION DOSIMETRY, 2011, 147 (03) : 373 - 379
  • [46] EXPERIMENTAL-VERIFICATION OF THE AIR KERMA TO ABSORBED DOSE CONVERSION FACTOR CW,U
    MIJNHEER, BJ
    WITTKAMPER, FW
    AALBERS, AHL
    VANDIJK, E
    RADIOTHERAPY AND ONCOLOGY, 1987, 8 (01) : 49 - 56
  • [47] An international calibration of Kerma-Area Product meters for patient dose optimisation study
    Jankowski, J.
    Domienik, J.
    Papierz, S.
    Padovani, R.
    Vano, E.
    Faulkner, K.
    RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) : 328 - 332
  • [48] Organ Doses and Effective Doses in Pediatric Radiography: Patient-Dose Survey in Finland
    Kiljunen, T.
    Tietavainen, A.
    Parviainen, T.
    Viitala, A.
    Kortesniemi, M.
    ACTA RADIOLOGICA, 2009, 50 (01) : 114 - U16
  • [49] Analysis of Factors Affecting Air Kerma Area Product Obtained during Uterine Artery Embolization Procedures Using Logistic Regression
    Soliman, Khaled
    Almutairi, Ahmed
    AlHarbi, Murdhi
    Almutairi, Khaleel
    Almutairi, Turky
    Bakkari, Mousa
    INDIAN JOURNAL OF RADIOLOGY AND IMAGING, 2021, 31 (02): : 265 - 269
  • [50] Position Sensitivity of Calculated Dose-Length-Product-To-Effective-Dose Conversion Factors in Computed Tomography Examination
    Gao, Y.
    Ding, A.
    Caracappa, P.
    Xu, X.
    MEDICAL PHYSICS, 2012, 39 (06) : 3607 - 3607