Conversion factors for effective dose and organ doses with the air Kerma area product in hysterosalpingography

被引:2
|
作者
D'Alessio, Andrea [1 ]
Cannillo, Barbara [1 ]
Guzzardi, Giuseppe [2 ]
Cernigliaro, Massimiliano [2 ]
Carriero, Alessandro [2 ]
Brambilla, Marco [1 ]
机构
[1] Univ Hosp Maggiore Carita, Med Phys Dept, Novara, Italy
[2] Univ Hosp Maggiore Carita, Radiol Dept, Novara, Italy
关键词
Histerosalpingography; Conversion factors; Dosimetry; Effective dose; Organ dose; OPTIMIZATION;
D O I
10.1016/j.ejmp.2020.11.032
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Histerosalpingography (HSG) remains the dominant diagnostic tool for investigation of infertility in women. Conversion factors used to estimate effective (E) and organ doses (H-T) from air Kerma area product (KAP) are needed to estimate patient doses in HSG, performed with state-of-the-art fluoroscopic X-ray systems with digital detectors. In this study, estimates of E and H-T for six critical organs/tissues, were derived on an individual basis in 120 HSG procedures and in 1410 irradiation events, performed on two X-ray systems from information available through the radiation dose structured report using Monte Carlo methods. Mean values of E and H-ovaries were 1.0 +/- 0.9 mSv and 5.6 +/- 5.4 mGy. E/KAP conversion factors of 0.13; 0.18; 0.28 and 0.35 mSv Gy(-1) cm(-2) were established for irradiation events with a Cu filtration of 0.0; 0.1; 0.4 and 0.9 mm. A high agreement was obtained between E estimated through Monte Carlo methods and E/KAP conversion factors accounting separately for the different modes of fluoroscopy and the radiography component of HSG, with a systematic error of 0 mSv and lower/upper limits of agreement of -0.6 and 0.5 mSv. On the contrary, the use of a single coefficient of conversion did not provide accurate estimates of E, showing a bias of -0.4 mSv and lower and upper limits of agreement of -1.9 and 1.2 mSv. An algorithm for the estimation of effective and organ doses from KAP has been established in HSG procedures depending on the Cu filtration in the X-ray irradiation events.
引用
收藏
页码:40 / 46
页数:7
相关论文
共 50 条