A new passive dosimetric system with thermoluminescent LiF: Mg, Cu, P detectors applied in individual radiation monitoring

被引:0
|
作者
Stochioiu, A. [1 ]
Mihai, F.
Scarlat, F.
机构
[1] Natl Inst Phys & Nucl Engn Horia Hulubei, Bucharest, Romania
[2] Natl Inst Laser Plasma & Radiat Phys, Bucharest, Romania
来源
关键词
dosimetry; thermoluminescent detectors; biological protection; radioactivity monitoring;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The paper presents the results of measuring the main parameters and characteristics of a new passive dosimetric system outfitted with thermo luminescent (TL) LiF Mg Cu P detectors for individual radiation monitoring, constructed in IFIN - HH. The presented dosimetric system is employed for measuring the equivalent absorbed dose of X, y and beta radiations within the energy range of 30 KeV - 3 MeV for a 1 to 90 days monitoring period. Among the parameters and characteristics of the new dosimetric system note: the homogeneity of the dosimetric response for a batch of 30 dosimeters is max. 10%; the dosimetric response repeatability is max. 3%; the measurement uncertainty within 7.5 mu Sv - 100 mSv range is max. 8%; the minimum traceable equivalent dose 70 nSv; loss of dosimetric response information in time is max. 2% per month and the variation of the dosimetric response with the incident radiation energy is max. 8%. The measurements and the system characterization were developed according to CEI 1066 standards previsions. The obtained values fall in the standard provisions. The paper also presents some constructive elements of the dosimetric system.
引用
收藏
页码:1545 / 1551
页数:7
相关论文
共 50 条
  • [31] The low energy X-ray response of the LiF:Mg:Cu:P thermoluminescent dosemeter:: a comparison with LiF:Mg:Ti
    Edwards, CR
    Mountford, PJ
    Green, S
    Palethorpe, JE
    Moloney, AJ
    BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (930): : 543 - 547
  • [32] INFLUENCE OF THE COOLING RATE ON REPEATABILITY OF LIF-MG,CU,P THERMOLUMINESCENT CHIPS
    PITERS, TM
    BOS, AJJ
    RADIATION PROTECTION DOSIMETRY, 1990, 33 (1-4) : 91 - 94
  • [33] Influence of annealing on kinetic trapping parameters in LiF:Mg,Cu,P thermoluminescent phosphors
    Chen, TC
    Stoebe, TG
    RADIATION MEASUREMENTS, 1998, 29 (01) : 39 - 44
  • [34] Preparation, thermoluminescent and electron spin resonance characteristics of LiF:Mg,Cu,P phosphor
    Shinde, SS
    Dhabekar, BS
    Rao, TKG
    Bhatt, BC
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2001, 34 (17) : 2683 - 2689
  • [35] Alpha particle radiation detection with LiF:Mg,Cu,P TL detectors using the TSEE technique
    Inst. de Pesq. Energeticas e Nucl., Comissao Nac. Ener. Nucl., C.P., São Paulo, Brazil
    Appl. Radiat. Isot., 4 (769-772):
  • [36] Alpha particle radiation detection with LiF:Mg,Cu,P TL detectors using the TSEE technique
    Rocha, FDG
    Caldas, LVE
    APPLIED RADIATION AND ISOTOPES, 1999, 50 (04) : 769 - 772
  • [37] Effect of dopants on TL characteristics of LiF:Mg,Cu,P detectors
    Knezevic, Z.
    Ranogajec-Komor, M.
    Miljanic, S.
    RADIATION MEASUREMENTS, 2010, 45 (3-6) : 573 - 575
  • [38] Nanorods of LiF:Mg,Cu,P as Detectors for Mixed Field Radiations
    Salah, Numan
    Habib, Sami S.
    Khan, Zishan H.
    Lochab, S. P.
    Kanjilal, D.
    Ranjan, Ranju
    Aleynikov, V. E.
    Rupasov, A. A.
    IEEE TRANSACTIONS ON NANOTECHNOLOGY, 2008, 7 (06) : 749 - 753
  • [39] THERMOLUMINESCENT RESPONSE OF (LIF)-LI-6 (MG,CU,P) AND (LIF)-LI-7 (MG,CU,P) TL CHIPS IN NEUTRON AND GAMMA-RAY MIXED RADIATION-FIELDS
    WANG, SS
    CAI, GG
    ZHOU, KQ
    ZHOU, RX
    RADIATION PROTECTION DOSIMETRY, 1990, 33 (1-4) : 247 - 250
  • [40] Application of LiF:Mg,Cu,P (MCP-N) thermoluminescent detectors (TLD) for experimental verification of radial dose distribution models
    Gieszczyk, Wojciech
    Olko, Pawel
    Bilski, Pawel
    Grzanka, Leszek
    Obryk, Barbara
    Horwacik, Tomasz
    NUKLEONIKA, 2012, 57 (04) : 507 - 512