Evaluation of a photon-counting breast tomosynthesis imaging system

被引:17
|
作者
Maidment, Andrew D. A. [1 ]
Ullberg, Christer [1 ]
Lindman, Karin [1 ]
Adelow, Leif [1 ]
Egerstrom, Johan [1 ]
Eklund, Mathias [1 ]
Francke, Tom [1 ]
Jordung, Ulf [1 ]
Kristoffersson, Tomas [1 ]
Lindqvist, Lars [1 ]
Marchal, Daniel [1 ]
Olla, Hans [1 ]
Penton, Erik [1 ]
Rantanen, Juha [1 ]
Solokov, Skiff [1 ]
Weber, Niclas [1 ]
Westerberg, Hans [1 ]
机构
[1] Univ Penn, Philadelphia, PA 19104 USA
关键词
digital breast tomosynthesis; synthetic tomography; digital mammography; modulation transfer function (MTF); noise power spectra (NPS); clinical trial;
D O I
10.1117/12.654651
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Digital breast tomosynthesis promises solutions to many of the problems associated with projection mammography, including elimination of artifactual densities due to the superposition of normal tissues and increasing the conspicuity of true lesions that would otherwise be masked by superimposed normal tissue. We have investigated tomosynthesis using a digital camera containing 48 photon counting, orientation sensitive, linear detectors which are precisely aligned with the focal spot of the x-ray source. The x-ray source and the digital detectors are scanned in a continuous motion across the object (patient), each linear detector collecting an image at a distinct angle. A preliminary assessment of tomosynthesis image quality has been performed with both qualitative and quantitative methods. Measured values of MTF and NPS appear concordant with theoretical values. The MTF in the scanning direction is dominated by scanning unsharpness and geometric factors, while the NPS is white. The MTF and NPS in the strip direction are somewhat lower than in the scan direction. The NPS of tomographic images show a slight decrease with increasing spatial frequency, related to the sampling and interpolation in the reconstruction process. A phase I clinical trial is ongoing; 9 women have been recruited. Breast positioning is comparable to other imaging systems. The visualization of breast anatomy appears to be superior to screen-film mammography, at the same average glandular dose. Examination of images reconstructed with a sub-sampled set of projection images appears to support the hypothesis that image quality is superior when more projection images are used in the reconstruction.
引用
收藏
页码:B1420 / B1420
页数:11
相关论文
共 50 条
  • [41] Performance Characteristics of Photon-Counting Breast CT
    Kalender, W.
    MEDICAL PHYSICS, 2016, 43 (06) : 3829 - 3830
  • [42] PHOTON-COUNTING
    LONGWORTH, JW
    PHOTOCHEMISTRY AND PHOTOBIOLOGY, 1977, 26 (06) : 665 - 668
  • [43] Spectral photon-counting CT in cardiovascular imaging
    Sandfort, Veit
    Persson, Mats
    Pourmorteza, Amir
    Noel, Peter B.
    Fleischmann, Dominik
    Willemink, Martin J.
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2021, 15 (03) : 218 - 225
  • [44] Musculoskeletal imaging with a prototype photon-counting detector
    Gruber, M.
    Homolka, P.
    Chmeissani, M.
    Uffmann, M.
    Pretterklieber, M.
    Kainberger, F.
    EUROPEAN RADIOLOGY, 2012, 22 (01) : 205 - 210
  • [45] Depth imaging denoising of photon-counting lidar
    Huang, Pengwei
    He, Weiji
    Gu, Guohua
    Chen, Qian
    APPLIED OPTICS, 2019, 58 (16) : 4390 - 4394
  • [46] First Clinical Photon-counting Detector CT System: Technical Evaluation
    Rajendran, Kishore
    Petersilka, Martin
    Henning, Andre
    Shanblatt, Elisabeth R.
    Schmidt, Bernhard
    Flohr, Thomas G.
    Ferrero, Andrea
    Baffour, Francis
    Diehn, Felix E.
    Yu, Lifeng
    Rajiah, Prabhakar
    Fletcher, Joel G.
    Leng, Shuai
    McCollough, Cynthia H.
    RADIOLOGY, 2022, 303 (01) : 130 - 138
  • [47] Evaluation of a new photon-counting imaging detector (PCD) with various acquisition modes
    Shankar, A.
    Krebs, J.
    Bednarek, D. R.
    Rudin, S.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [48] Lightweight Omnidirectional Radiation Protection for a Photon-Counting Imaging System in Space Applications
    Han, Zhen-Wei
    Song, Ke-Fei
    Liu, Shi-Jie
    Guo, Quan-Feng
    Ding, Guang-Xing
    He, Ling-Ping
    Li, Cheng-Wei
    Zhang, Hong-Ji
    Liu, Yang
    Chen, Bo
    APPLIED SCIENCES-BASEL, 2023, 13 (10):
  • [49] PHOTON-COUNTING SYSTEM FOR CORRELATION AMPLITUDE MEASUREMENTS
    EVTYUSHENKOV, AM
    KIYACHENKO, YF
    OLEFIRENKO, GI
    YUDIN, IK
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1981, 24 (05) : 1265 - 1267
  • [50] THE MEPSICRON - A NEW BIDIMENSIONAL PHOTON-COUNTING SYSTEM
    FIRMANI, C
    RUIZ, E
    BOHIGAS, J
    BISIACCHI, GF