Evaluation of a photon-counting breast tomosynthesis imaging system

被引:17
|
作者
Maidment, Andrew D. A. [1 ]
Ullberg, Christer [1 ]
Lindman, Karin [1 ]
Adelow, Leif [1 ]
Egerstrom, Johan [1 ]
Eklund, Mathias [1 ]
Francke, Tom [1 ]
Jordung, Ulf [1 ]
Kristoffersson, Tomas [1 ]
Lindqvist, Lars [1 ]
Marchal, Daniel [1 ]
Olla, Hans [1 ]
Penton, Erik [1 ]
Rantanen, Juha [1 ]
Solokov, Skiff [1 ]
Weber, Niclas [1 ]
Westerberg, Hans [1 ]
机构
[1] Univ Penn, Philadelphia, PA 19104 USA
关键词
digital breast tomosynthesis; synthetic tomography; digital mammography; modulation transfer function (MTF); noise power spectra (NPS); clinical trial;
D O I
10.1117/12.654651
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Digital breast tomosynthesis promises solutions to many of the problems associated with projection mammography, including elimination of artifactual densities due to the superposition of normal tissues and increasing the conspicuity of true lesions that would otherwise be masked by superimposed normal tissue. We have investigated tomosynthesis using a digital camera containing 48 photon counting, orientation sensitive, linear detectors which are precisely aligned with the focal spot of the x-ray source. The x-ray source and the digital detectors are scanned in a continuous motion across the object (patient), each linear detector collecting an image at a distinct angle. A preliminary assessment of tomosynthesis image quality has been performed with both qualitative and quantitative methods. Measured values of MTF and NPS appear concordant with theoretical values. The MTF in the scanning direction is dominated by scanning unsharpness and geometric factors, while the NPS is white. The MTF and NPS in the strip direction are somewhat lower than in the scan direction. The NPS of tomographic images show a slight decrease with increasing spatial frequency, related to the sampling and interpolation in the reconstruction process. A phase I clinical trial is ongoing; 9 women have been recruited. Breast positioning is comparable to other imaging systems. The visualization of breast anatomy appears to be superior to screen-film mammography, at the same average glandular dose. Examination of images reconstructed with a sub-sampled set of projection images appears to support the hypothesis that image quality is superior when more projection images are used in the reconstruction.
引用
收藏
页码:B1420 / B1420
页数:11
相关论文
共 50 条
  • [21] Dedicated Photon-Counting CT for Detection and Classification of Microcalcifications: An Intraindividual Comparison With Digital Breast Tomosynthesis
    Huck, Luisa Charlotte
    Bode, Maike
    Zanderigo, Eloisa
    Wilpert, Caroline
    Raaff, Vanessa
    Dethlefsen, Ebba
    Wenkel, Evelyn
    Kuhl, Christiane Katharina
    INVESTIGATIVE RADIOLOGY, 2024, 59 (12) : 838 - 844
  • [22] Development of Low-Dose Photon-counting Contrast-enhanced Tomosynthesis with Spectral Imaging
    Schmitzberger, Florian F.
    Fallenberg, Eva Maria
    Lawaczeck, Ruediger
    Hemmendorff, Magnus
    Moa, Elin
    Danielsson, Mats
    Bick, Ulrich
    Diekmann, Susanne
    Poellinger, Alexander
    Engelken, Florian J.
    Diekmann, Felix
    RADIOLOGY, 2011, 259 (02) : 558 - 564
  • [23] Photon-counting CT in lung imaging
    Tarnoki, adam Domonkos
    Tarnoki, David Laszlo
    Maurovich-Horvat, Pal
    EXPERT REVIEW OF RESPIRATORY MEDICINE, 2025, 19 (02) : 77 - 81
  • [24] PHOTON-COUNTING IMAGING AND ITS APPLICATION
    TSUCHIYA, Y
    INUZUKA, E
    KURONO, T
    HOSODA, M
    ADVANCES IN ELECTRONICS AND ELECTRON PHYSICS, 1985, 64 : 21 - 31
  • [25] Observer-model optimization of X-ray system in photon-counting breast imaging
    Cederstrom, Bjorn
    Fredenberg, Erik
    Lundqvist, Mats
    Ericson, Tove
    Aslund, Magnus
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S54 - S57
  • [26] Breast Tissue Characterization with Photon-counting Spectral CT Imaging: A Postmortem Breast Study
    Ding, Huanjun
    Klopfer, Michael J.
    Ducote, Justin L.
    Masaki, Fumitaro
    Molloi, Sabee
    RADIOLOGY, 2014, 272 (03) : 731 - 738
  • [27] Case Report: Spectral Tomosynthesis-Galactography with photon-counting Detector
    Diekmann, F.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2013, 185 (09): : 880 - 881
  • [28] Spiral breast computed tomography with a photon-counting detector (SBCT): The future of breast imaging?
    Wetzl, Matthias
    Dietzel, Matthias
    Ohlmeyer, Sabine
    Uder, Michael
    Wenkel, Evelyn
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 157
  • [29] Volumetric Breast-Density Measurement Using Spectral Photon-Counting Tomosynthesis: First Clinical Results
    Fredenberg, Erik
    Berggren, Karl
    Bartels, Matthias
    Erhard, Klaus
    BREAST IMAGING, IWDM 2016, 2016, 9699 : 576 - 584
  • [30] PHOTON-COUNTING
    WILLIAMS, MD
    SCIENCE, 1982, 215 (4534) : 745 - 745