A large-scale multicentre study of patient skin doses in interventional cardiology: dose-area product action levels and dose reference levels

被引:66
|
作者
Bogaert, E. [1 ]
Bacher, K. [1 ]
Lemmens, K. [1 ]
Carlier, M. [2 ]
Desmet, W. [3 ]
De Wagter, X. [4 ]
Djian, D. [5 ]
Hanet, C. [6 ]
Heyndrickx, G. [7 ]
Legrand, V. [8 ]
Taeymans, Y. [9 ]
Thierens, H. [1 ]
机构
[1] Univ Ghent, Dept Med Phys & Radiat Protect, B-9000 Ghent, Belgium
[2] Hop St Joseph, Dept Intervent Cardiol, Gilly, Belgium
[3] Univ Ziekenhuis Gasthuisberg, Louvain, Belgium
[4] Algemeen Ziekenhuis Maria Middelares, Ghent, Belgium
[5] Ctr Hosp Namur, Namur, Belgium
[6] Clin Univ St Luc, B-1200 Brussels, Belgium
[7] Onze Lieve Vrouw Hosp, Aalst, Belgium
[8] Ctr Hosp Univ, Liege, Belgium
[9] Univ Ziekenhuis, Ghent, Belgium
来源
BRITISH JOURNAL OF RADIOLOGY | 2009年 / 82卷 / 976期
关键词
FLUOROSCOPICALLY GUIDED PROCEDURES; CORONARY ANGIOPLASTY; RADIATION-EXPOSURE; CARDIAC PROCEDURES; INJURIES; ANGIOGRAPHY; RADIOLOGY; CATHETERIZATION; VALUES;
D O I
10.1259/bjr/29449648
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
For 318 patients in 8 different Belgian hospitals, the entire skin-dose distribution was mapped using a grid of 70 thermoluminescence dosemeters per patient, allowing an accurate determination of the maximum skin dose (MSD). Dose-area product (DAP) values, exposure parameters and geometry, together with procedure, patient and cardiologist characteristics, were also registered. Procedures were divided into two groups: diagnostic procedures (coronary angiography) and therapeutic procedures (dilatation, stent, combined procedures (e.g. coronary angiography + dilatation + stent)). The mean value of the MSD was 0.310 Gy for diagnostic and 0.699 Gy for therapeutic procedures. The most critical projection for receiving the MSD is the LAO90 (left anterior oblique) geometry. In 3% of cases, the MSD exceeded the 2 Gy dose threshold for deterministic effects. Action levels in terms of DAP values as the basis for a strategy for follow-up of patients for deterministic radiation skin effects were derived from measured MSD and cumulative DAP values. Two DAP action levels are proposed. A first DAP action level of 125 Gy cm(2) corresponding to the dose threshold of 2 Gy would imply an optional radiopathological follow-up depending on the cardiologist's decision. A second DAP action level of 250 Gy cm(2) corresponding to the 3 Gy skin dose would imply a systematic follow-up. Dose reference levels - 71.3 Gy cm(2) for diagnostic and 106.0 Gy cm(2) for therapeutic procedures - were derived from the 75 percentile of the DAP distributions. As a conclusion, we propose that total DAP is registered in patient's record file, as it can serve to improve the follow-up of. patients for radiation-induced skin injuries.
引用
收藏
页码:303 / 312
页数:10
相关论文
共 45 条
  • [41] DOSE AREA PRODUCT VALUES OF FLUOROSCOPICALLY GUIDED PAIN MANAGEMENT PROCEDURES: COMPARISON OF A BELGIAN TEACHING HOSPITAL WITH NATIONAL DIAGNOSTIC REFERENCE LEVELS
    Hustinx, Nicolas
    Steyaert, Arnaud
    de Waroux, Bernard le Polain
    Forget, Patrice
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (03) : 271 - 274
  • [42] Multicentre survey on patient dose in paediatric imaging and proposal for updated diagnostic reference levels for France. Part 2: plain radiography and diagnostic fluoroscopy
    David Célier
    Patrice Roch
    Cécile Etard
    Hubert Ducou Le Pointe
    Hervé J. Brisse
    European Radiology, 2020, 30 : 1182 - 1190
  • [43] Multicentre survey on patient dose in paediatric imaging and proposal for updated diagnostic reference levels for France. Part 2: plain radiography and diagnostic fluoroscopy
    Celier, David
    Roch, Patrice
    Etard, Cecile
    Ducou Le Pointe, Hubert
    Brisse, Herve J.
    EUROPEAN RADIOLOGY, 2020, 30 (02) : 1182 - 1190
  • [44] Do patients with skin allergies have higher levels of anxiety than patients with allergic respiratory diseases? Results of a large-scale cross-sectional study in a French population
    Annesi-Maesano, I.
    Beyer, A.
    Marmouz, F.
    Mathelier-Fusade, P.
    Vervloet, D.
    Bauchau, V.
    BRITISH JOURNAL OF DERMATOLOGY, 2006, 154 (06) : 1128 - 1136
  • [45] Diagnostic Reference Levels (DRLs) and Achievable Doses (ADs) for the 11 Most Commonly Performed Pediatric CT Examinations in the United States (USA) as a Function of Patient Age and Size Using 1.5 Million Examinations in the American College of Radiology (ACR) CT Dose Index Registry
    Kanal, K.
    Butler, P.
    Chatfield, M.
    Wells, J.
    Samei, E.
    Simanowith, M.
    Golden, D.
    Gress, D.
    Burleson, J.
    Sensakovic, W.
    Strauss, K.
    Frush, D.
    MEDICAL PHYSICS, 2021, 48 (06)