Positioning of a plane-parallel ionization chamber in clinical electron beams and the impact on perturbation factors

被引:24
|
作者
Zink, K. [1 ]
Wulff, J. [1 ]
机构
[1] Univ Appl Sci Giessen Friedberg, Inst Med Phys & Strahlenschutz IMPS, D-35390 Giessen, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 08期
关键词
MONTE-CARLO-SIMULATION; STOPPING-POWER RATIOS; QUALITY CORRECTION FACTORS; ION CHAMBERS; MEGAVOLTAGE PHOTON; DOSIMETRY; PLATE; TRANSPORT; PENELOPE; PROTOCOL;
D O I
10.1088/0031-9155/54/8/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Current dosimetry protocols recommend the use of plane-parallel ionization chambers for the dosimetry of clinical electron beams. The necessary perturbation corrections p(wall) and p(cav) are assumed to be unity, independent of the depth of measurement and the energy of the primary electrons. To verify these assumptions detailed Monte Carlo studies of a Roos chamber in clinical electron beams with energies in the range of 6-21 MeV are performed at different depths in water and analyzed in terms of Spencer-Attix cavity theory. Separate simulations for the perturbation corrections p(wall) and p(cav) indicate quite different properties of both correction factors with depth. Dose as well as fluence calculations show a nearly depth-independent wall correction factor for a shift of the Roos chamber Delta z = -0.017 cm toward the focus. This value is in good agreement with the positioning recommendation given in all dosimetry protocols. Regarding the fluence perturbation p(cav) the simulation of the electron fluence inside the air cavity in comparison to water unambiguously reveals an in-scattering of low energy electrons, despite the fact, that the cavity is 'well guarded'. For depths beyond the reference depth z(ref) this effect is superimposed by an increased loss of primary electrons from the beam resulting in p(cav) > 1. This effect is largest for low electron energies but present for all electron energies involved in this study. Based on the different depth dependences of p(wall) and p(cav) it is possible to choose a chamber shift Delta z in a way to minimize the depth dependence of the overall perturbation factor p. For the Roos chamber this shift is Delta z = -0.04 cm independent of electron energy.
引用
收藏
页码:2421 / 2435
页数:15
相关论文
共 50 条
  • [31] Underlying mechanism of the magnetic field dependence of a plane-parallel chamber in proton beams
    Blum, Isabel
    Wong, Jing Syuen
    Padre, Krishna Godino
    Stolzenberg, Jessica
    Baumann, Kilian-Simon
    Fuchs, Hermann
    Poppe, Bjoern
    Looe, Hui Khee
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S3386 - S3388
  • [32] Monte Carlo determination of electron beam quality correction factors for plane-parallel ionization chambers
    Aldana, J
    Sempau, J
    Mazurier, J
    Andreo, P
    Salvat, C
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S27 - S27
  • [33] Correction of the measured current of a small-gap plane-parallel ionization chamber in proton beams in the presence of charge multiplication
    Rossomme, Severine
    Lorentini, Stefano
    Vynckier, Stefaan
    Delor, Antoine
    Vidal, Marie
    Lourenco, Ana
    Palmans, Hugo
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2021, 31 (02): : 192 - 202
  • [34] CALIBRATION OF PLANE-PARALLEL IONIZATION CHAMBERS
    ALMOND, PR
    DANESHUARNEZHAD, M
    WILSON, D
    BHANDARE, N
    MEDICAL PHYSICS, 1987, 14 (03) : 491 - 491
  • [35] COMPARISON OF THE CALIBRATION FACTOR NGAS FOR A PLANE-PARALLEL IONIZATION-CHAMBER DETERMINED IN CO-60 AND HIGH-ENERGY ELECTRON-BEAMS
    MURALI, V
    MEENAKSHI, S
    LAKSHMANAN, AV
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (10): : 1503 - 1507
  • [36] ON THE CALIBRATION OF PLANE-PARALLEL IONIZATION CHAMBERS FOR ELECTRON-BEAM DOSIMETRY
    ANDREO, P
    RODRIGUES, LN
    LINDBORG, L
    KRAEPELIEN, T
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (05): : 1147 - 1165
  • [37] Proton dosimetry in a magnetic field: Measurement and calculation of magnetic field correction factors for a plane-parallel ionization chamber
    Gebauer, Benjamin
    Baumann, Kilian-Simon
    Fuchs, Hermann
    Georg, Dietmar
    Oborn, Brad M.
    Looe, Hui-Khee
    Luehr, Armin
    MEDICAL PHYSICS, 2024, 51 (03) : 2293 - 2305
  • [38] Determination of ion recombination and polarity effect correction factors for a plane-parallel ionization Bragg peak chamber under proton and carbon ion pencil beams
    Mirandola, A.
    Magro, G.
    Maestri, D.
    Mairani, A.
    Mastella, E.
    Molinelli, S.
    Russo, S.
    Avai, A.
    Ciocca, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (09):
  • [39] POLARITY EFFECT OF PLANE-PARALLEL IONIZATION CHAMBERS IN ELECTRON-RADIATION
    HAVERCROFT, JM
    KLEVENHAGEN, SC
    PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02): : 299 - 304
  • [40] Feasibility of Using Large Plane-Parallel Ionization Chamber to Validate Small Field Dosimetry
    Sabouri, P.
    Mossahebi, S.
    Langner, U.
    Guerrero, M.
    Yi, B.
    D'Souza, W.
    Chen, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E589 - E590