Positioning of a plane-parallel ionization chamber in clinical electron beams and the impact on perturbation factors

被引:24
|
作者
Zink, K. [1 ]
Wulff, J. [1 ]
机构
[1] Univ Appl Sci Giessen Friedberg, Inst Med Phys & Strahlenschutz IMPS, D-35390 Giessen, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 08期
关键词
MONTE-CARLO-SIMULATION; STOPPING-POWER RATIOS; QUALITY CORRECTION FACTORS; ION CHAMBERS; MEGAVOLTAGE PHOTON; DOSIMETRY; PLATE; TRANSPORT; PENELOPE; PROTOCOL;
D O I
10.1088/0031-9155/54/8/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Current dosimetry protocols recommend the use of plane-parallel ionization chambers for the dosimetry of clinical electron beams. The necessary perturbation corrections p(wall) and p(cav) are assumed to be unity, independent of the depth of measurement and the energy of the primary electrons. To verify these assumptions detailed Monte Carlo studies of a Roos chamber in clinical electron beams with energies in the range of 6-21 MeV are performed at different depths in water and analyzed in terms of Spencer-Attix cavity theory. Separate simulations for the perturbation corrections p(wall) and p(cav) indicate quite different properties of both correction factors with depth. Dose as well as fluence calculations show a nearly depth-independent wall correction factor for a shift of the Roos chamber Delta z = -0.017 cm toward the focus. This value is in good agreement with the positioning recommendation given in all dosimetry protocols. Regarding the fluence perturbation p(cav) the simulation of the electron fluence inside the air cavity in comparison to water unambiguously reveals an in-scattering of low energy electrons, despite the fact, that the cavity is 'well guarded'. For depths beyond the reference depth z(ref) this effect is superimposed by an increased loss of primary electrons from the beam resulting in p(cav) > 1. This effect is largest for low electron energies but present for all electron energies involved in this study. Based on the different depth dependences of p(wall) and p(cav) it is possible to choose a chamber shift Delta z in a way to minimize the depth dependence of the overall perturbation factor p. For the Roos chamber this shift is Delta z = -0.04 cm independent of electron energy.
引用
收藏
页码:2421 / 2435
页数:15
相关论文
共 50 条
  • [21] CORRECTION FACTORS APPLIED TO PLANE-PARALLEL IONIZATION CHAMBERS
    NYSTROM, H
    KARLSSON, M
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (02): : 311 - 322
  • [22] CALIBRATION AND USE OF PLANE-PARALLEL IONIZATION CHAMBERS FOR THE DETERMINATION OF ABSORBED DOSE IN ELECTRON-BEAMS
    MATTSSON, LO
    JOHANSSON, KA
    SVENSSON, H
    ACTA RADIOLOGICA ONCOLOGY, 1981, 20 (06): : 385 - 399
  • [23] Ion recombination and polarity correction factors for a plane-parallel ionization chamber in a proton scanning beam
    Liszka, Malgorzata
    Stolarczyk, Liliana
    Klodowska, Magdalena
    Kozera, Anna
    Krzempek, Dawid
    Mojzeszek, Natalia
    Pezdracka, Anna
    Waligorski, Michael Patrick Russell
    Olko, Pawel
    MEDICAL PHYSICS, 2018, 45 (01) : 391 - 401
  • [24] Ion recombination and polarity correction for a plane-parallel ionization chamber in hadrontherapy
    Maestri, D.
    Mirandola, A.
    Magro, G.
    Mairani, A.
    Mastella, E.
    Molinelli, S.
    Russo, S.
    Vai, A.
    Ciocca, M.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S948 - S948
  • [25] SPACE CHARGE DISTORTION OF ELECTRIC FIELD IN A PLANE-PARALLEL IONIZATION CHAMBER
    BOAG, JW
    PHYSICS IN MEDICINE AND BIOLOGY, 1963, 8 (1-4): : 461 - &
  • [26] Experimental determination of pCo perturbation factors for plane-parallel chambers
    Kapsch, R. P.
    Bruggmoser, G.
    Christ, G.
    Dohm, O. S.
    Hartmann, G. H.
    Schuele, E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (23): : 7167 - 7181
  • [27] Analysis of calibration coefficients of plane-parallel ionization chambers calibrated in Co-60 and electron beams
    Bulski, W.
    Ulkowski, P.
    Gwiazdowska, B.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S236 - S236
  • [28] SKIN EXIT DOSE IN MEGAVOLTAGE X-RAY-BEAMS DETERMINED BY MEANS OF A PLANE-PARALLEL IONIZATION-CHAMBER (ATTIX CHAMBER)
    KRON, T
    OSTWALD, P
    MEDICAL PHYSICS, 1995, 22 (05) : 577 - 578
  • [29] Numerical solutions of differential equations of an ionization chamber: plane-parallel and spherical geometry
    Novkovic, D
    Tomasevic, M
    Subotic, K
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (03): : 559 - 575
  • [30] Wall correction factors for calibration of plane-parallel ionization chambers with high-energy photon beams
    Araki, F
    Ikeda, R
    Shirakawa, Y
    Shimonobou, T
    Moribe, N
    Takada, T
    Takahashi, M
    Oura, H
    Matoba, M
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09): : 2509 - 2517