Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography

被引:1
|
作者
Alikhani, Babak [1 ,3 ]
Werner, Martin [2 ]
Jamali, Leila [1 ]
Wacker, Frank [1 ]
Werncke, Thomas [1 ]
机构
[1] Hannover Med Sch, Inst Diagnost & Intervent Radiol, Hannover, Germany
[2] Hannover Med Sch, Inst Radiat Therapy & Special Oncol, Hannover, Germany
[3] DIAKOVERE gGmbH, Ctr Radiol & Nucl Med, Hannover, Germany
关键词
CT; Dual-source CT; Dose splitting; Iterative reconstruction; Image quality; RADIATION-DOSE REDUCTION; ITERATIVE RECONSTRUCTION; ENERGY CT; SPATIAL-RESOLUTION; ANGIOGRAPHY; PHANTOM;
D O I
10.1016/j.acra.2018.09.021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: The aim of this study is to analyze the image quality provided by a dual-source (DS) data set and a single source (SS) data set at the same radiation exposure, in order to evaluate a dose splitting method for dual-source scanning protocols. Materials and Methods: A 192-slice dual-source third generation CT (Somatom Force; Siemens Healthcare, Forchheim, Germany) was used to image a Catphan phantom (Catphan503; The Phantom Laboratory, Salem, New York) utilizing different X-ray tube voltages from 70 to 120 kVp with an organ-based tube-current modulation technique (X-CARE; Siemens Healthcare, Forchheim, Germany). In order to keep the radiation dose (given by volume computed tomography dose index) in a clinically relevant range, different X-ray tube time-current products ranging from 80 to 300 mAs were selected. The data sets by each X-ray tube voltage were collected using a single-source as well as a dual-source mode. The measurements in the dual-source mode were performed with five different tube currents of the X-ray tube A and B. Thereby, the tube current ratios were 50%vertical bar 50% (DS 0.5), 60%vertical bar 40% (DS 0.6), 70%vertical bar 30% (DS 0.7), 80%vertical bar 20% (DS 0.8), and 90%vertical bar 10% (DS 0.9). The images were reconstructed by the use of a filter-back projection (Br40) and an advanced mode led iterative reconstruction algorithms (advanced modeled iterative reconstruction algorithms [ADMIRE]; Siemens Healthcare, Forchheim, Germany) with a strength range of 1-5. The image quality was evaluated in terms of noise, contrast-to-noise ratio (CNR), low-contrast detectability expressed as the structural similarity index (SSIM) and spatial resolution quantified by the full width at half maximum of the line-spread function. Results: Image noise decreased by the use of the dual-source mode, which led to improvement of their CNR compared to the single source mode. SSIM showed an almost constant behavior by both modes. The spatial resolution indicated a lower trend by the dual-source mode in comparison to the single-source mode. However, the loss of the spatial resolution performance was lower than 5% for the dual-source modes. Conclusion: The presented phantom study demonstrated that SSIM and spatial resolution performance obtained by dual-source CT protocols showed a negligible variation to those by the single-source CT. However, the noise and CNR displayed an improvement for the dual-source CT. Therefore, the use of the dual-source CT enables to split the radiation dose between X-ray tubes and to compare the data sets with different radiation dose levels without loss in the image quality.
引用
收藏
页码:1095 / 1101
页数:7
相关论文
共 50 条
  • [1] Radiation dose and image quality of computed tomography of the supra-aortic arteries: A comparison between single-source and dual-source CT Scanners
    Saba, Luca
    di Martino, Michele
    Siotto, Paolo
    Anzidei, Michele
    Argiolas, Giovanni Maria
    Porcu, Michele
    Suri, Jasjit S.
    Wintermark, Max
    [J]. JOURNAL OF NEURORADIOLOGY, 2018, 45 (02) : 136 - 141
  • [2] Dual-Source Versus Single-Source Cardiac CT Angiography: Comparison of Diagnostic Image Quality
    Donnino, Robert
    Jacobs, Jill E.
    Doshi, Jay V.
    Hecht, Elizabeth M.
    Kim, Danny C.
    Babb, James S.
    Srichai, Monvadi B.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) : 1051 - 1056
  • [3] Dual-source cardiac computed tomography: image quality and dose considerations
    Achenbach, Stephan
    Anders, Katharina
    Kalender, Willi A.
    [J]. EUROPEAN RADIOLOGY, 2008, 18 (06) : 1188 - 1198
  • [4] Dual-source cardiac computed tomography: image quality and dose considerations
    Stephan Achenbach
    Katharina Anders
    Willi A. Kalender
    [J]. European Radiology, 2008, 18 : 1188 - 1198
  • [5] Virtual Single-Source Computed Tomography Using Dual-Source Acquisition A New Technique for the Dose-Neutral Intraindividual Comparison of Different Scan Protocols
    Werncke, Thomas
    Meyer, Bernhard Christian
    Wacker, Frank K.
    von Falck, Christian
    [J]. INVESTIGATIVE RADIOLOGY, 2014, 49 (11) : 742 - 748
  • [6] Temporal resolution and motion artifacts in single-source and dual-source cardiac CT
    Schoendube, Harald
    Allmendinger, Thomas
    Stierstorfer, Karl
    Bruder, Herbert
    Flohr, Thomas
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [7] Dual-source CT coronary angiography in patients with atrial fibrillation: Comparison with single-source CT
    Wang, Yining
    Zhang, Zhuhua
    Kong, Lingyan
    Song, Lan
    Merges, Reto D.
    Chen, Jiuhong
    Jin, Zhengyu
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2008, 68 (03) : 434 - 441
  • [8] Dual-source computed tomography in patients with acute chest pain: feasibility and image quality
    Thomas Schertler
    Hans Scheffel
    Thomas Frauenfelder
    Lotus Desbiolles
    Sebastian Leschka
    Paul Stolzmann
    Burkhardt Seifert
    Thomas G. Flohr
    Borut Marincek
    Hatem Alkadhi
    [J]. European Radiology, 2007, 17 : 3179 - 3188
  • [9] Diagnostic Accuracy and Image Quality of Cardiac Dual-Source Computed Tomography in Patients with Arrhythmia
    Drosch, Tanja
    Reimann, Anja
    Brodoefel, Harald
    Tsiflikas, Ilias
    Thomas, Christoph
    Heuschmid, Martin
    Schroeder, Stephen
    Burgstahler, Christof
    [J]. CIRCULATION, 2008, 118 (18) : S845 - S845
  • [10] Dual-source computed tomography in patients with acute chest pain: feasibility and image quality
    Schertler, Thomas
    Scheffel, Hans
    Frauenfelder, Thomas
    Desbiolles, Lotus
    Leschka, Sebastian
    Stolzmann, Paul
    Seifert, Burkhardt
    Flohr, Thomas G.
    Marincek, Borut
    Alkadhi, Hatem
    [J]. EUROPEAN RADIOLOGY, 2007, 17 (12) : 3179 - 3188