Properties of a multipole ion trap studied by evaporative ion losses

被引:9
|
作者
Lakhmanskaya, O. Y. [1 ]
Best, T. [1 ]
Kumar, S. S. [1 ]
Endres, E. S. [1 ]
Hauser, D. [1 ]
Otto, R. [2 ]
Eisenbach, S. [3 ]
von Zastrow, A. [4 ]
Wester, R. [1 ]
机构
[1] Univ Insbruck, Inst Ionenphys & Angew Phys, A-6020 Innsbruck, Austria
[2] Univ Calif San Diego, Dept Chem & Biochem, La Jolla, CA 92093 USA
[3] Univ Freiburg, Inst Phys, D-79104 Freiburg, Germany
[4] Radboud Univ Nijmegen, Inst Mol & Mat IMM, NL-6525 AJ Nijmegen, Netherlands
基金
欧洲研究理事会;
关键词
Ion trap; 22-Pole trap; Ion temperature; Buffer gas cooling; ACTION SPECTROSCOPY; TEMPERATURE; H-3(+);
D O I
10.1016/j.ijms.2014.03.001
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
We have studied a scheme for analysing the effective trapping potential for ions in a 22-pole ion trap, based on the strong dependence of the lifetime of the ions on the buffer gas temperature. We distinguish two regimes for ion losses, either over one of the two end cap electrodes (axial direction) or over the effective radiofrequency potential (radial direction), and examine these schemes under different trap settings. We have found that evaporation over one of the end caps yields an effective temperature of the ions that is much larger than the buffer gas temperature. This may be a hint for an enhancement of the high-energy tail of the ions' velocity distribution. We also observe and investigate a breakdown of thermalisation for trap temperatures lower than 25 K. Measurements for OH- and D- anions have shown that this feature is independent of both the buffer gas to ion mass ratio and the trapping parameters. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:281 / 286
页数:6
相关论文
共 50 条
  • [11] ION ENERGY SPECTRA AND LIFETIMES IN A MULTIPOLE MAGNETIC TRAP
    DEMIDENKO, II
    LOMINO, NS
    PADALKA, VG
    SOVIET PHYSICS TECHNICAL PHYSICS-USSR, 1971, 15 (11): : 1823 - +
  • [12] THE NONLINEAR ION-TRAP .3. MULTIPOLE COMPONENTS IN 3 TYPES OF PRACTICAL ION-TRAP
    WANG, Y
    FRANZEN, J
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1994, 132 (03): : 155 - 172
  • [13] Determination of multipole coefficients in toroidal ion trap mass analysers
    Kotana, Appala Naidu
    Mohanty, Atanu K.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2016, 408 : 62 - 76
  • [14] Experimental demonstration of a surface-electrode multipole ion trap
    Maurice, Mark
    Allen, Curtis
    Green, Dylan
    Farr, Andrew
    Burke, Timothy
    Hilleke, Russell
    Clark, Robert
    JOURNAL OF APPLIED PHYSICS, 2015, 118 (07)
  • [15] Design Study of a Multipole Ion Trap for Beam Physics Applications
    Fukushima, Kei
    Okamoto, Hiromi
    Plasma and Fusion Research, 2015, 10 : 1 - 9
  • [16] Ion ring in a linear multipole trap for optical frequency metrology
    Champenois, C.
    Marciante, M.
    Pedregosa-Gutierrez, J.
    Houssin, M.
    Knoop, M.
    Kajita, M.
    PHYSICAL REVIEW A, 2010, 81 (04):
  • [17] Planar Multipole Ion Trap/Time-of-Flight Mass Spectrometer
    Lorenz, Ulrich J.
    Rizzo, Thomas R.
    ANALYTICAL CHEMISTRY, 2011, 83 (20) : 7895 - 7901
  • [18] Numerical simulations of kinetic ion temperature in a cryogenic linear multipole trap
    Asvany, Oskar
    Schlemmer, Stephan
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2009, 279 (2-3) : 147 - 155
  • [19] Ion-ion collision-induced evaporative cooling of atomic and molecular ions by autoresonance in an electrostatic ion beam trap
    Heber, O.
    Gangwar, R. K.
    Saha, K.
    Rappaport, M. L.
    Zajfman, D.
    31ST INTERNATIONAL CONFERENCE ON PHOTONIC, ELECTRONIC AND ATOMIC COLLISIONS (ICPEAC XXXI), 2020, 1412
  • [20] ION LOSSES FROM END-STOPPERED MIRROR TRAP
    CHERNIN, DP
    ROSENBLUTH, MN
    NUCLEAR FUSION, 1978, 18 (01) : 47 - 62