Assessing and Improving Cobalt-60 Digital Tomosynthesis Image Quality

被引:0
|
作者
Marsh, Matthew B. [1 ]
Schreiner, L. John [1 ]
Kerr, Andrew T. [1 ]
机构
[1] Queens Univ, Canc Ctr Southeastern Ontario, Dept Med Phys, Kingston, ON, Canada
关键词
Image guidance; radiation therapy; cobalt-60; tomosynthesis; IGRT; patient positioning; BEAM COMPUTED-TOMOGRAPHY; GUIDED RADIATION-THERAPY; CBDT; RADIOGRAPHY;
D O I
10.1117/12.2043066
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Image guidance capability is an important feature of modern radiotherapy linacs, and future cobalt-60 units will be expected to have similar capabilities. Imaging with the treatment beam is an appealing option, for reasons of simplicity and cost, but the dose needed to produce cone beam CT (CBCT) images in a Co-60 treatment beam is too high for this modality to be clinically useful. Digital tomosynthesis (DT) offers a quasi-3D image, of sufficient quality to identify bony anatomy or fiducial markers, while delivering a much lower dose than CBCT. A series of experiments were conducted on a prototype Co-60 cone beam imaging system to quantify the resolution, selectivity, geometric accuracy and contrast sensitivity of Co-60 DT. Although the resolution is severely limited by the penumbra cast by the 2 cm diameter source, it is possible to identify high contrast objects on the order of 1 mm in width, and bony anatomy in anthropomorphic phantoms is clearly recognizable. Low contrast sensitivity down to electron density differences of 3% is obtained, for uniform features of similar thickness. The conventional shift-and-add reconstruction algorithm was compared to several variants of the Feldkamp-Davis-Kress filtered backprojection algorithm result. The Co-60 DT images were obtained with a total dose of 5 to 15 cGy each. We conclude that Co-60 radiotherapy units upgraded for modern conformal therapy could also incorporate imaging using filtered backprojection DT in the treatment beam. DT is a versatile and promising modality that would be well suited to image guidance requirements.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Cobalt-60 Source Based Image Guidance in Broad Beam Cobalt-60 IMRT
    Kerr, A.
    Rawluk, N.
    MacDonald, A.
    Marsh, M.
    Schreiner, J.
    Darko, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S701 - S701
  • [2] Digital Tomosynthesis Image Quality in a Co-60 Treatment Beam
    Marsh, M.
    Schreiner, L. J.
    Kerr, A.
    MEDICAL PHYSICS, 2012, 39 (07) : 4633 - 4633
  • [3] COBALT-60
    ROBERTS, JE
    BRITISH JOURNAL OF RADIOLOGY, 1952, 25 (294): : 281 - 281
  • [4] Multiscale bilateral filtering for improving image quality in digital breast tomosynthesis
    Lu, Yao
    Chan, Heang-Ping
    Wei, Jun
    Hadjiiski, Lubomir M.
    Samala, Ravi K.
    MEDICAL PHYSICS, 2015, 42 (01) : 182 - 195
  • [5] COBALT-60
    REDWAY, LD
    NEW YORK STATE JOURNAL OF MEDICINE, 1952, 52 (18) : 2229 - 2230
  • [6] Megavoltage Digital Tomosynthesis Using a Radioactive Cobalt-60 Gamma Ray Source for Radiation Therapy Treatment Verification
    MacDonald, A.
    Darko, J.
    Schreiner, L. J.
    MEDICAL PHYSICS, 2009, 36 (09) : 4314 - 4314
  • [7] Total quality management of cobalt-60 sources
    Malkoske, GR
    RADIATION PHYSICS AND CHEMISTRY, 1999, 54 (06) : 601 - 608
  • [8] PURIFICATION BY COBALT-60
    不详
    TECHNOLOGY REVIEW, 1974, 76 (05): : 65 - 65
  • [9] Research and application of digital image processing technology in cobalt-60 based container inspection system
    Bo Sun
    Yun Liu
    Jun He
    CHEMICAL, BIOLOGICAL, RADIOLOGICAL, NUCLEAR, AND EXPLOSIVES (CBRNE) SENSING IX, 2008, 6954
  • [10] COBALT-60 TELETHERAPY
    不详
    JOURNAL OF THE COLLEGE OF RADIOLOGISTS OF AUSTRALASIA, 1965, 9 (02): : 174 - &