Flow visualization in a centrifugal blood pump with an eccentric inlet port

被引:8
|
作者
Yamane, T
Kodama, T
Yamamoto, Y
Shinohara, T
Nosé, Y
机构
[1] Natl Inst Adv Ind Sci & Technol, Inst Human Sci & Biomed Engn, Tsukuba, Ibaraki 3058564, Japan
[2] Tokyo Univ Sci, Chiba, Japan
[3] Miwatec Co Ltd, Kanagawa, Japan
[4] Baylor Coll Med, Houston, TX 77030 USA
关键词
centrifugal pump; flow visualization; eccentric inlet port; shear rate;
D O I
10.1111/j.1525-1594.2004.07290.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Flow visualization analysis was applied to the Baylor/Miwatec centrifugal artificial heart to evaluate its fluid dynamic characteristics regarding antithrombogenicity. An eccentric vortex was found both in the upper and the lower gaps of the impeller, which is supposed to be caused by the eccentric inlet port. Therefore, one-way flow toward the outlet is formed and washes the pivot. The combination of an eccentric vortex and a pivot bearing that is washed is unique to the Baylor/Miwatec pump. For the male pivots exposed to periodic wash, the minimum shear rate around the bottom pivot was estimated to be 650/s, which is higher than the threshold for thrombus formation shown by other studies. The wall shear rate at the impeller bottom surface was found to be larger in the top contact mode than in the bottom contact mode.
引用
收藏
页码:564 / 570
页数:7
相关论文
共 50 条
  • [1] Eccentric inlet port of the pivot bearing supported Gyro centrifugal pump
    Takami, Y
    Andrade, A
    Nakazawa, T
    Makinouchi, K
    Glueck, J
    Benkowski, R
    Nose, Y
    [J]. ARTIFICIAL ORGANS, 1997, 21 (04) : 312 - 317
  • [2] Inlet port positioning for a miniaturized centrifugal blood pump
    Takano, T
    Schulte-Eistrup, S
    Kawahito, S
    Maeda, T
    Nonaka, K
    Linneweber, J
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    [J]. ARTIFICIAL ORGANS, 2002, 26 (01) : 45 - 48
  • [3] Flow visualization studies in a centrifugal blood pump
    Ng, B.T.H.
    Chan, W.K.
    Yu, S.C.M.
    Chua, L.P.
    [J]. American Society of Mechanical Engineers, Bioengineering Division (Publication) BED, 1999, 42 : 163 - 164
  • [4] DEVELOPMENT AND EVALUATION OF ANTITHROMBOGENIC CENTRIFUGAL PUMP - THE BAYLOR-C-GYRO-PUMP-ECCENTRIC-INLET-PORT-MODEL
    OHARA, Y
    MAKINOUCHI, K
    GLUECK, J
    SUTHERLAND, B
    SHIMONO, T
    NAITO, K
    TASAI, K
    ORIME, Y
    TAKATANI, S
    NOSE, Y
    [J]. ARTIFICIAL ORGANS, 1994, 18 (09) : 673 - 679
  • [5] Visualization of relative flow patterns in centrifugal blood pump
    Chan, WK
    Yu, SCM
    Chua, LP
    Wong, YW
    [J]. KSME INTERNATIONAL JOURNAL, 2001, 15 (12): : 1869 - 1875
  • [6] A FLOW VISUALIZATION STUDY OF THE NCVC CENTRIFUGAL BLOOD PUMP
    ARAKI, K
    TAENAKA, Y
    MASUZAWA, T
    TATSUMI, E
    WAKISAKA, Y
    WATARI, M
    NAKATANI, T
    AKAGI, H
    BABA, Y
    ANAI, H
    PARK, YH
    EYA, K
    TODA, K
    TAKANO, H
    [J]. ARTIFICIAL ORGANS, 1994, 18 (09) : 669 - 672
  • [7] Visualization of relative flow patterns in centrifugal blood pump
    Weng Kong Chan
    S. C. M Yu
    L. P. Chua
    Y. W. Wong
    [J]. KSME International Journal, 2001, 15 : 1869 - 1875
  • [8] Visualization of flow structure in a centrifugal blood pump impeller
    Tsukiya T.
    Taenaka Y.
    Tatsumi E.
    Takano H.
    [J]. Journal of Artificial Organs, 2002, 5 (1) : 13 - 17
  • [9] Visualization study of the transient flow in the centrifugal blood pump impeller
    Tsukiya, T
    Taenaka, Y
    Tatsumi, E
    Takano, H
    [J]. ASAIO JOURNAL, 2002, 48 (04) : 431 - 436
  • [10] Flow visualization study to improve hemocompatibility of a centrifugal blood pump
    Nishida, M
    Asztalos, B
    Yamane, T
    Masuzawa, T
    Tsukiya, T
    Endo, S
    Taenaka, Y
    Miyazoe, Y
    Ito, K
    Konishi, Y
    [J]. ARTIFICIAL ORGANS, 1999, 23 (08) : 697 - 703