Comparative Evaluation of Foveal Avascular Zone on Two Optical Coherence Tomography Angiography Devices

被引:10
|
作者
Dave, Prachi A. [1 ]
Dansingani, Kunal K. [2 ]
Jabeen, Ayesha [3 ]
Jabeen, Asiya [3 ]
Ali, M. Hasnat [3 ]
Vupparaboina, Kiran Kumar [3 ]
Peguda, Hari Kumar [3 ]
Pappurru, Rajeev Reddy [3 ]
Agrawal, Rupesh [4 ]
Chhablani, Jay [3 ]
机构
[1] Dr Shroffs Char Eye Hosp, New Delhi, India
[2] Univ Nebraska Med Ctr, Truhlsen Eye Inst, Omaha, NE USA
[3] Srimati Kanuri Santhamma Ctr Vitreo Retinal Dis, LV Prasad Eye Inst, KAR Campus, Hyderabad, Telangana, India
[4] Tan Tock Seng Hosp, Natl Healthcare Grp, Eye Inst, Singapore, Singapore
关键词
DIABETIC-RETINOPATHY; QUANTIFICATION; AREA;
D O I
10.1097/OPX.0000000000001238
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE: This study compares foveal avascular zone (FAZ) geometry in healthy eyes as imaged by two commercially available optical coherence tomography angiography (OCTA) devices. Foveal avascular zone measurements are repeatable and reproducible with each OCTA device, but interdevice agreement was poor. We provide conversion factors between devices. PURPOSE: The purpose of this study was to perform comparative evaluation of FAZ geometry in healthy eyes as imaged by two commercially available OCTA devices. METHODS: Ninety-six eyes of 48 healthy subjects were imaged prospectively on each of two OCTA devices (DRI-OCT [Topcon Corporation, Tokyo, Japan]; Cirrus 5000 [Carl Zeiss Meditec Inc., Dublin, CA]). The FAZ was evaluated in the superficial capillary plexus layer of 6 x 6-mm foveal scans by two masked observers. Intraobserver and interobserver agreement was determined using intraclass correlation by using linear mixed models and Bland-Altman plots. K-means clustering was used to provide conversion values between two devices. Foveal avascular zone acircularity was calculated using scans from each device and compared. RESULTS: The intraobserver repeatability for DRI-OCT was 0.95 (95% confidence interval [CI], 0.90 to 0.98) for observer A and 0.92 (95% CI, 0.83 to 0.96) for observer B. Intraobserver repeatability for Cirrus 5000 was 0.988 (95% CI, 0.972 to 0.995) for observer A and 0.993 (95% CI, 0.983 to 0.997) for observer B. The interobserver variability between observers A and B for DRI-OCT was 0.87 (0.73 to 0.94) and for Cirrus 5000 was 0.984 (95% CI, 0.964 to 0.993). Poor interdevice agreement (0.205 [95% CI, -0.202 to 0.554]) was noted, and conversion formulas were devised to convert FAZ area measurements from one device to another. No significant correlation was found when comparing FAZ acircularity indices between devices (P = .39). CONCLUSIONS: Repeatable and reproducible FAZ area measurements were obtained with each respective OCTA device, but interdevice agreement was poor, yet quantifiable and systematic with calculable conversion factors between devices.
引用
收藏
页码:602 / 607
页数:6
相关论文
共 50 条
  • [31] Predictability of Progression of Diabetic Retinopathy by the Assessment of Foveal Avascular Zone with Optical Coherence Tomography Angiography
    Suzuki, Katsuya
    Nozaki, Miho
    Takase, Noriaki
    Kato, Aki
    Morita, Hiroshi
    Ozeki, Hironori
    Yoshida, Munenori
    Ogura, Yuichiro
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [32] Longitudinal analysis of the foveal avascular zone in diabetic retinopathy using optical coherence tomography angiography
    Cai, Yijun
    Mohamed-Noriega, Jibran
    Keane, Pearse Andrew
    Egan, Catherine A.
    Tufail, Adnan
    Sim, Dawn A.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2016, 57 (12)
  • [33] FOVEAL AVASCULAR ZONE VOLUME A New Index Based on Optical Coherence Tomography Angiography Images
    Xu, Qiuzhuo
    Zhang, Weiwei
    Zhu, Hongjing
    Chen, Qiang
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2021, 41 (03): : 595 - 601
  • [34] Variability of Foveal Avascular Zone Metrics Derived From Optical Coherence Tomography Angiography Images
    Linderman, Rachel E.
    Muthiah, Manickam N.
    Omoba, Sarah B.
    Litts, Katie
    Tarima, Sergey
    Visotcky, Alexis
    Kim, Judy E.
    Carroll, Joseph
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2018, 7 (05):
  • [35] Quantitative optical coherence tomography angiography of macular vascular structure and foveal avascular zone in glaucoma
    Choi, Jaewan
    Kwon, Junki
    Shin, Joong Won
    Lee, Jiyun
    Lee, Saem
    Kook, Michael S.
    PLOS ONE, 2017, 12 (09):
  • [36] Investigation of factors that may affect the foveal avascular zone: An optical coherence tomography angiography study
    Scanlon, Grainne
    O'Shea, Susan
    Amarandei, George
    Butler, John S.
    O'Dwyer, Veronica
    OPTOMETRY AND VISION SCIENCE, 2024, 101 (05) : 276 - 283
  • [37] Bilateral Enlargement of the Foveal Avascular Zone on Optical Coherence Tomography Angiography Associated With Cocaine Use
    Fortenbach, Christopher R.
    Modjtahedi, Bobeck S.
    JAMA OPHTHALMOLOGY, 2020, 138 (06) : 708 - 710
  • [38] Reproducibility and repeatability of foveal avascular zone measurements in healthy subjects by optical coherence tomography angiography
    Carpineto, Paolo
    Mastropasqua, Rodolfo
    Marchini, Giorgio
    Toto, Lisa
    Di Nicola, Marta
    Di Antonio, Luca
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (05) : 671 - 676
  • [39] Evaluation of Automatically Quantified Foveal Avascular Zone Metrics for Diagnosis of Diabetic Retinopathy Using Optical Coherence Tomography Angiography
    Lu, Yansha
    Simonett, Joseph M.
    Wang, Jie
    Zhang, Miao
    Hwang, Thomas
    Hagag, Ahmed M.
    Huang, David
    Li, Dengwang
    Jia, Yali
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (06) : 2212 - 2221
  • [40] Optical coherence tomography angiography analysis of the foveal avascular zone: evaluation and measurement of reproducibility and reliability in different settings.
    Arrigo, Alessandro
    La Spina, Carlo
    Carnevali, Adriano
    Marchese, Alessandro
    Sacconi, Riccardo
    Rabiolo, Alessandro
    Querques, Lea
    Cicinelli, Maria Vittoria
    Querques, Giuseppe
    Bandello, Francesco
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)