IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry

被引:7
|
作者
Zhang, Yang [1 ]
Zheng, Lin [1 ]
Bian, Ailing [1 ]
Zhou, Qi [1 ]
机构
[1] Chinese Acad Med Sci, Dept Ophthalmol, Peking Union Med Coll Hosp, Peking Union Med Coll, Dongdan Shuaifuyuan 1, Beijing 100730, Peoples R China
关键词
Intraocular pressure; Silicone oil tamponade; Goldmann applanation tonometry; Corvis ST tonometer; Corneal biomechanical properties; INTRAOCULAR-PRESSURE; BIOMECHANICAL PROPERTIES; CORNEAL BIOMECHANICS; GLAUCOMA; DEFORMATION; PARAMETERS; THICKNESS; IMPACT;
D O I
10.1007/s10792-017-0519-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT). Methods Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland-Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device. Results Bland-Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be -0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST-NCT and GAT-NCT. Conclusions In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.
引用
收藏
页码:697 / 703
页数:7
相关论文
共 50 条
  • [41] The accuracy of Goldmann applanation tonometry over silicone hydrogel contact lenses in patients with glaucoma
    Gomes, Beatriz A. F.
    Turiel, Paulo
    Cavalcanti, Rachel S.
    Moraes, Haroldo V., Jr.
    Santhiago, Marcony R.
    CONTACT LENS & ANTERIOR EYE, 2014, 37 (04): : 297 - 299
  • [42] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Li, Yuehua
    Shi, Jingming
    Duan, Xuanchu
    Fan, Fang
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (12) : 1765 - 1770
  • [43] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Yuehua Li
    Jingming Shi
    Xuanchu Duan
    Fang Fan
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1765 - 1770
  • [44] Corneal Modulus and IOP Measurements in Canine Eyes Using Goldmann Applanation Tonometry and Tono-pen
    Tang, Junhua
    Pan, Xueliang
    Weber, Paul A.
    Liu, Jun
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (11) : 7866 - 7871
  • [45] Inter-optometrist variability of IOP measurement for modern tonometers and their agreement with Goldmann Applanation Tonometry
    Campbell, Peter
    Edgar, David F.
    Shah, Rakhee
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2021, 104 (05) : 602 - 610
  • [46] Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry
    Oncel, Banu
    Dinc, Umut Asli
    Orge, Faruk
    Yalvac, Belkis Ilgaz
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (06) : 936 - 941
  • [47] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Khalid Kamel
    Edward Dervan
    Kevin Falzon
    Colm O’Brien
    Irish Journal of Medical Science (1971 -), 2019, 188 : 321 - 325
  • [48] Difference in intraocular pressure measurements between non-contact tonometry and Goldmann applanation tonometry and the role of central corneal thickness in affecting glaucoma referrals
    Kamel, Khalid
    Dervan, Edward
    Falzon, Kevin
    O'Brien, Colm
    IRISH JOURNAL OF MEDICAL SCIENCE, 2019, 188 (01) : 321 - 325
  • [50] ON TONOMETRY .5. PRESSURE OF GLAUCOMATOUS EYES MEASURED IN SITTING AND LYING POSITIONS BY MEANS OF GOLDMANN APPLANATION TONOMETER
    HETLANDE.J
    ACTA OPHTHALMOLOGICA, 1966, 44 (04): : 515 - &