IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry

被引:7
|
作者
Zhang, Yang [1 ]
Zheng, Lin [1 ]
Bian, Ailing [1 ]
Zhou, Qi [1 ]
机构
[1] Chinese Acad Med Sci, Dept Ophthalmol, Peking Union Med Coll Hosp, Peking Union Med Coll, Dongdan Shuaifuyuan 1, Beijing 100730, Peoples R China
关键词
Intraocular pressure; Silicone oil tamponade; Goldmann applanation tonometry; Corvis ST tonometer; Corneal biomechanical properties; INTRAOCULAR-PRESSURE; BIOMECHANICAL PROPERTIES; CORNEAL BIOMECHANICS; GLAUCOMA; DEFORMATION; PARAMETERS; THICKNESS; IMPACT;
D O I
10.1007/s10792-017-0519-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT). Methods Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland-Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device. Results Bland-Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be -0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST-NCT and GAT-NCT. Conclusions In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.
引用
收藏
页码:697 / 703
页数:7
相关论文
共 50 条
  • [31] Comparison among Ocular Response Analyzer,Corvis ST and Goldmann applanation tonometry in healthy children
    Ramin Salouti
    Ali agha Alishiri
    Reza Gharebaghi
    Mostafa Naderi
    Khosrow Jadidi
    Ahmad Shojaei-Baghini
    Mohammadreza Talebnejad
    Zahra Nasiri
    Seyedmorteza Hosseini
    Fatemeh Heidary
    International Journal of Ophthalmology, 2018, 11 (08) : 1330 - 1336
  • [32] Comparison among Ocular Response Analyzer, Corvis ST and Goldmann applanation tonometry in healthy children
    Salouti, Ramin
    Alishiri, Ali Agha
    Gharebaghi, Reza
    Naderi, Mostafa
    Jadidi, Khosmw
    Shojaei-Baghini, Ahmad
    Talebnejad, Mohammadreza
    Nasiri, Zahra
    Hosseini, Seyedmorteza
    Heidary, Fatemeh
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2018, 11 (08) : 1330 - 1336
  • [33] Influence of Corneal Thickness in Keratoconic Corneas on IOP Measurement with IOPen, iCare, Dynamic Contour Tonometry and Goldmann Applanation Tonometry
    Klamann, M. K. J.
    Maier, A. -K. B.
    Gonnermann, J.
    Ruokonen, P.
    Bertelmann, E.
    Torun, N.
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2013, 230 (07) : 697 - 700
  • [34] Influence of IOP measurement through the wrong eyepiece of the slit lamp on Goldmann applanation tonometry
    Kubota, T.
    Nagata, T.
    Tawara, A.
    Okada, H.
    Ishibashi, S.
    Yamashita, M.
    Harada, Y.
    Yanase, M.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2008, 18 (06) : 910 - 914
  • [35] Comparison of Goldmann Applanation Tonometry and Dynamic Contour Tonometry in the Measurement of Intraocular Pressure in Eyes with Different Corneal Thicknesses
    Roszkowska, A. M.
    De Grazia, L.
    Cirone, M.
    Ferreri, G.
    OPHTHALMOLOGICA, 2009, 223 (04) : 244 - 249
  • [36] Intraocular pressure and corneal thickness - A comparison between non-contact tonometry and applanation tonometry
    Domke, N.
    Hager, A.
    Wiegand, W.
    OPHTHALMOLOGE, 2006, 103 (07): : 583 - 587
  • [37] The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.
    Kyei, Samuel
    Assiamah, Frank
    Kwarteng, Michael Agyemang
    Gboglu, Cynthia Pakyennu
    ETHIOPIAN JOURNAL OF HEALTH SCIENCES, 2020, 30 (06) : 999 - 1004
  • [38] Effect of central corneal thickness on intraocular pressure and comparison of Topcon CT-80 non-contact tonometry with Goldmann applanation tonometry
    Mansoori, Tarannum
    Balakrishna, Nagalla
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2018, 101 (02) : 206 - 212
  • [39] Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty
    Salvetat, M. L.
    Zeppieri, M.
    Miani, F.
    Tosoni, C.
    Parisi, L.
    Brusini, P.
    EYE, 2011, 25 (05) : 642 - 650
  • [40] Comparison of iCare tonometer and Goldmann applanation tonometry in normal corneas and in eyes with automated lamellar and penetrating keratoplasty
    M L Salvetat
    M Zeppieri
    F Miani
    C Tosoni
    L Parisi
    P Brusini
    Eye, 2011, 25 : 642 - 650