IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry

被引:7
|
作者
Zhang, Yang [1 ]
Zheng, Lin [1 ]
Bian, Ailing [1 ]
Zhou, Qi [1 ]
机构
[1] Chinese Acad Med Sci, Dept Ophthalmol, Peking Union Med Coll Hosp, Peking Union Med Coll, Dongdan Shuaifuyuan 1, Beijing 100730, Peoples R China
关键词
Intraocular pressure; Silicone oil tamponade; Goldmann applanation tonometry; Corvis ST tonometer; Corneal biomechanical properties; INTRAOCULAR-PRESSURE; BIOMECHANICAL PROPERTIES; CORNEAL BIOMECHANICS; GLAUCOMA; DEFORMATION; PARAMETERS; THICKNESS; IMPACT;
D O I
10.1007/s10792-017-0519-1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the postoperative intraocular pressure (IOP) of eyes following pars plana vitrectomy (PPV) combined with intravitreal silicone oil (SO) tamponade by Corneal Visualization Scheimpflug Technology (CST), Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT). Methods Thirty-eight participants who had undergone PPV combined with SO tamponade to treat vitreoretinal diseases were enrolled. Postoperative IOP measurements were obtained using CST, NCT and GAT. Inter-device agreement was assessed by Bland-Altman analysis. The correlation coefficient was used to describe the potential postoperative factors affecting the postoperative IOP differences between each device. Results Bland-Altman analysis revealed the bias between CST and GAT, between CST and NCT, and between GAT and NCT to be -0.2, 2.1 and 2.4 mmHg, respectively. CST and GAT correlated well with each other. NCT values were lower than those of GAT and CST (all p < 0.05), whereas CST values did not differ from the GAT readings. Central corneal thickness, corneal biomechanical properties and age showed significant correlation with the differences of CST-NCT and GAT-NCT. Conclusions In SO tamponade eyes, NCT obtains lower IOP than other tonometry techniques, and CST is highly consistent with GAT. CST offers an optional non-contact method for measuring postoperative IOP in SO tamponade eyes.
引用
收藏
页码:697 / 703
页数:7
相关论文
共 50 条
  • [1] IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry
    Yang Zhang
    Lin Zheng
    Ailing Bian
    Qi Zhou
    International Ophthalmology, 2018, 38 : 697 - 703
  • [2] Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry
    Luebke, Jan
    Bryniok, L.
    Neuburger, M.
    Jordan, J. F.
    Boehringer, D.
    Reinhard, T.
    Wecker, T.
    Anton, A.
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (11) : 2517 - 2521
  • [3] Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry
    Jan Luebke
    L. Bryniok
    M. Neuburger
    J. F. Jordan
    D. Boehringer
    T. Reinhard
    T. Wecker
    A. Anton
    International Ophthalmology, 2019, 39 : 2517 - 2521
  • [4] IOP Measured by Dynamic Contour Tonometry Correlates With IOP Measured by Goldmann Applanation Tonometry and Non-contact Tonometry in Japanese Individuals
    Ito, Ken
    Tawara, Akihiko
    Kubota, Toshiaki
    Harada, Yukinori
    JOURNAL OF GLAUCOMA, 2012, 21 (01) : 35 - 40
  • [5] Comparison of IOP-measurements obtained by the non-contact tonometer AT555 and Goldmann applanation tonometry
    Mueller-Holz, M
    Schmidt, E
    Boehm, AG
    Pillunat, LE
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [6] Comparison of intraocular pressure using Goldmann applanation tonometry versus non-contact tonometry in eyes with high-viscosity silicone oil
    Pagoulatos, Dionysios D.
    Kapsala, Zoi G.
    Makri, Olga E.
    Georgalas, Ilias G.
    Georgakopoulos, Constantinos D.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2020, 30 (03) : 494 - 499
  • [7] Comparison of rebound tonometry and non-contact airpuff tonometry to Goldmann applanation tonometry
    Demirci, Goktug
    Erdur, Sevil Karaman
    Tanriverdi, Cafer
    Gulkitik, Gokhan
    Ozsutcu, Mustafa
    THERAPEUTIC ADVANCES IN OPHTHALMOLOGY, 2019, 11
  • [8] NON-CONTACT TONOMETRY (KEELER PULSAIR) VS GOLDMANN APPLANATION TONOMETRY
    CLIMENHAGA, HW
    PLUCINSKA, H
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 1988, 23 (02): : 86 - 86
  • [9] COMPARATIVE TONOMETRY WITH NON-CONTACT TONOMETER (NCT), PNEUMATONOGRAPH (PTG), AND GOLDMANN APPLANATION TONOMETER (GAT)
    SHIONO, T
    HARA, S
    KIMURA, R
    FOLIA OPHTHALMOLOGICA JAPONICA, 1978, 29 (10): : 1552 - 1555
  • [10] Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry
    Bhartiya S.
    Bali S.J.
    Sharma R.
    Chaturvedi N.
    Dada T.
    International Ophthalmology, 2011, 31 (4) : 297 - 302