A comparison of subdermal wire electrodes with collodion-applied disk electrodes in long-term EEG recordings in ICU

被引:33
|
作者
Young, G. Bryan
Ives, John R.
Chapman, Martin G.
Mirsattari, Seyed M.
机构
[1] Univ Western Ontario, Univ Hosp, Dept Clin Neurol Sci, London, ON N6A 5A5, Canada
[2] Univ Toronto, Toronto, ON, Canada
关键词
EEG; electrodes; subdermal wire; disk; artifact; cEEG; LTM;
D O I
10.1016/j.clinph.2006.02.006
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To compare long-term electroencephalographic (EEG) recordings of standard collodion-applied scalp disk electrodes (SDEs) with newly developed subdermal wire electrodes (SWEs) in comatose intensive care unit (ICU) patients. Methods: Ten comatose ICU patients had simultaneous recordings from 8 active SDEs and 8 active SVV'E for > 24 h. The timing and number of 60 Hz and other electrode artifacts were compared for each set of electrodes by an EEGer who read the recordings in a blinded manner. Results: Sixty Hertz artifact was seen in 16 of 80 SDE and one of 80 SWEs within the first 6 h (P = 0.0002). Large, persistent artifacts occurred in 30/80 SDE and 8 of 80 SWE (P=0.0001). Motion artifact with chest physiotherapy was more common in SWEs. Conclusions: SWE are less susceptible to artifacts and are more suitable for the long-term EEG monitoring in ICU. Significance: This is the first controlled study that demonstrates the superiority of SWEs compared to SDEs in an ICU population. (c) 2006 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:1376 / 1379
页数:4
相关论文
共 27 条
  • [1] Subdermal wire electrodes vs. collodion-applied disk electrodes for long-term ICU monitoring
    Young, B
    Ives, J
    Chapman, M
    Mirsattari, S
    [J]. JOURNAL OF THE NEUROLOGICAL SCIENCES, 2005, 238 : S525 - S525
  • [2] Ring and peg electrodes for minimally-Invasive and long-term sub-scalp EEG recordings
    Benovitski, Y. B.
    Lai, A.
    McGowan, C. C.
    Burns, O.
    Maxim, V.
    Nayagam, D. A. X.
    Millard, R.
    Rathbone, G. D.
    le Chevoir, M. A.
    Williams, R. A.
    Grayden, D. B.
    May, C. N.
    Murphy, M.
    D'Souza, W. J.
    Cook, M. J.
    Williams, C. E.
    [J]. EPILEPSY RESEARCH, 2017, 135 : 29 - 37
  • [3] Novel Dry Polymer Foam Electrodes for Long-Term EEG Measurement
    Lin, Chin-Teng
    Liao, Lun-De
    Liu, Yu-Hang
    Wang, I-Jan
    Lin, Bor-Shyh
    Chang, Jyh-Yeong
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (05) : 1200 - 1207
  • [4] THE VALUE OF STEREOTAXIC RECORDINGS AT OPERATION AND OF LONG-TERM IMPLANTED ELECTRODES IN PLANING LEUCOTOMIES
    SHERWOOD, SL
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1959, 11 (04): : 840 - 840
  • [5] Embedded Ultrathin Cluster Electrodes for Long-Term Recordings in Deep Brain Centers
    Etemadi, Leila
    Mohammed, Mohsin
    Thorbergsson, Palmi Thor
    Ekstrand, Joakim
    Friberg, Annika
    Granmo, Marcus
    Pettersson, Lina M. E.
    Schouenborg, Jens
    [J]. PLOS ONE, 2016, 11 (05):
  • [6] Ionic Liquid Gel-Assisted Electrodes for Long-Term Cutaneous Recordings
    Leleux, Pierre
    Johnson, Camryn
    Strakosas, Xenofon
    Rivnay, Jonathan
    Herve, Thierry
    Owens, Roisin M.
    Malliaras, George G.
    [J]. ADVANCED HEALTHCARE MATERIALS, 2014, 3 (09) : 1377 - 1380
  • [7] New electrodes for long-term EEG monitoring compatible with 3 Tesla MRI and CT imaging; Application to ICU patients
    Vulliemoz, Serge
    Ives, J. R.
    Spinelli, L.
    Perrig, S.
    Pellise, D.
    Lazeyras, F.
    Seeck, M.
    [J]. EPILEPSIA, 2007, 48 : 183 - 184
  • [8] Long-term stability of intracortical recordings using perforated and arrayed Parylene sheath electrodes
    Hara, Seth A.
    Kim, Brian J.
    Kuo, Jonathan T. W.
    Lee, Curtis D.
    Meng, Ellis
    Pikov, Victor
    [J]. JOURNAL OF NEURAL ENGINEERING, 2016, 13 (06)
  • [9] Characterization of a new type of dry electrodes for long-term recordings of surface-electromyogram
    Garcia, Gonzalo A.
    Zaccone, Franco
    Ruff, Roman
    Micera, Silvestro
    Hoffmann, Klaus-Peter
    Dario, Paolo
    [J]. 2007 IEEE 10TH INTERNATIONAL CONFERENCE ON REHABILITATION ROBOTICS, VOLS 1 AND 2, 2007, : 849 - +
  • [10] Wire Electrodes Embedded in Artificial Conduit for Long-term Monitoring of the Peripheral Nerve Signal
    Jung, Woohyun
    Jung, Sunyoung
    Kim, Ockchul
    Park, HyungDal
    Choi, Wonsuk
    Son, Donghee
    Chung, Seok
    Kim, Jinseok
    [J]. MICROMACHINES, 2019, 10 (03)