APMP comparison of activity measurements of I-131 (APMP.RI(II)-K2.I-131)

被引:0
|
作者
Yunoki, Akira
Bignell, Lindsey [1 ]
Mo, Li [1 ]
Joseph, Leena [2 ]
Yuan, Ming-Chen
Park, Tae-Soon [3 ]
Soodprasert, Thongchai
Wurdiyanto, Gatot
机构
[1] ANSTO, Lucas Heights, NSW, Australia
[2] BARC, Bombay, Maharashtra, India
[3] KRISS, Taejon, South Korea
关键词
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The international comparison of activity measurements of 131I (APMP.RI(II)-K2.I-131) was carried out within the framework of the Asia-Pacific Metrology Programme (APMP). Seven laboratories took part in the comparison, and five of them undertook absolute measurements. One ampoule containing the same radioactive solution as that used in the APMP comparison was sent to the International Reference System (SIR) for activity comparison at the Bureau International des Poids et Mesures (BIPM) in order to link the APMP comparison to the BIPM.RI(II)-K1.I-131 comparison and to evaluate degrees of equivalence with the key comparison reference value (KCRV).