Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

被引:2
|
作者
Fenoglietto, P. [1 ]
Khodri, M. [2 ]
Nguyen, D. [3 ]
Josserand-Pietri, F. [4 ]
Ailleres, N. [1 ]
机构
[1] Inst Reg Canc Montpellier ICM, Dept Oncol Radiotherapie, F-34298 Montpellier 5, France
[2] ICL, Dept Oncol Radiotherapie, St Etienne, France
[3] Grp ORLAM, Serv Radiotherapie, Macon, France
[4] Grp ORLAM, Serv Radiotherapie, Villeurbanne, France
关键词
VOLUMETRIC-MODULATED ARC; QUALITY-ASSURANCE; CLINICAL IMPLEMENTATION; PROSTATE-CANCER; DYNAMIC IMRT; RADIOTHERAPY; DELIVERY; THERAPY; RAPIDARC; VERIFICATION;
D O I
10.1186/s13014-015-0577-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc (TM) device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. Patients and Methods: Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc (TM) technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm. Results: The number of points passing the gamma criteria between the measured and computed doses was 94.79 +/- 2.57 % for linac A and 94.61 +/- 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 +/- 1.26 % and 98.59 +/- 1.20 % points passing the threshold were obtained for linacs A and B, respectively. The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 +/- 0.06 % of the points passing the criteria. Conclusion: For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Quality assurance of linear accelerator: a comprehensive system using electronic portal imaging device
    Niyas, P.
    Abdullah, K. K.
    Noufal, M. P.
    Vysakh, R.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (02) : 138 - 149
  • [42] Using An Electronic Portal Imaging Device (EPID) for Correlating Linac Photon Beam Energies
    Yaddanapudi, S.
    Cai, B.
    Sun, B.
    Noel, C.
    Goddu, S.
    Mutic, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3517 - 3517
  • [43] CT, laser and Linac axes alignment using an electronic portal imaging device (EPID)
    Samant, S
    Xia, J
    Patra, P
    Sobczak, D
    MEDICAL PHYSICS, 2004, 31 (06) : 1876 - 1876
  • [44] Modeling Electronic Portal Imaging Device for Dose Reconstruction using Monte Carlo Method
    Jung, J. W.
    Kim, J. O.
    Yeo, I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S724 - S724
  • [45] Multi-leaf-collimator quality assurance using the electronic portal imaging device
    Ali, I
    Benedict, S
    Li, W
    Lerma, F
    Dogan, N
    Siebers, J
    MEDICAL PHYSICS, 2005, 32 (06) : 2168 - 2168
  • [46] Evaluation of set up uncertainity in HNIMRT treatment using an electronic portal imaging device
    Vanoni, V.
    Delana, A.
    Menegotti, L.
    Ziglio, F.
    Bolner, A.
    Mussari, S.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S317 - S317
  • [47] Kilo-voltage cone-beam computed tomography setup measurements for lung cancer patients; First clinical results and comparison with electronic portal-imaging device
    Borst, Gerben R.
    Sonke, Jan-Jakob
    Betgen, Anja
    Remeijer, Peter
    Van Herk, Marcel
    Lebesque, Joos V.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (02): : 555 - 561
  • [48] Dosimetry Study On An Amorphous Silicon Electronic Portal Imaging Device(a-Si EPID)
    Wu, Q.
    Hu, Y.
    MEDICAL PHYSICS, 2009, 36 (06) : 2596 - +
  • [49] QUALITY ASSURANCE OF ENHANCED DYNAMIC WEDGE (EDW) USING ELECTRONIC PORTAL IMAGING DEVICE (EPID)
    Narayanan, S.
    Arasu, K.
    Narayanan, G.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S235 - S235
  • [50] A Multileaf Collimator and Jaw Symmetry Quality Assurance Technique Using An Electronic Portal Imaging Device
    Lebron, S.
    Lu, B.
    Yan, G.
    Kahler, D.
    Liu, C.
    MEDICAL PHYSICS, 2017, 44 (06) : 2933 - 2933